Determination of alkyl methylphosphonates and methylphosphonic acid by capillary electrophoresis with a diode-array spectrophotometric detector

被引:2
|
作者
Aleksenko, S. S. [1 ]
Tret'yakova, S. V. [1 ]
Mandych, V. G. [1 ]
机构
[1] Saratov Mil Inst Biol & Chem Safety, Saratov 410037, Russia
关键词
alkyl methylphosphonates; methylphosphonic acid; capillary electrophoresis; diode-array; AGENT DEGRADATION-PRODUCTS; GAS-CHROMATOGRAPHIC DETERMINATION; CHEMICAL WARFARE AGENTS; MASS SPECTROMETRY; SEPARATION; IDENTIFICATION; TRACE;
D O I
10.1134/S1061934811040022
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Method of capillary electrophoresis with a diode-array detector was used to study the conditions of the separation of three alkyl methylphosphonates and methylphosphonic acid, hydrolysis products of nerve agents. The optimal conditions providing their separation and determination are reported (background electrolyte based on 100 mM boric acid with the addition of 10 mM of phenylphosphonic acid (pH 6.0); voltage +30 kV, detection wavelength 210 nm, electrokinetic introduction of the sample (10 kV, 10 sec), separation time below 10 min). Calibration curve for all the compounds was linear in the concentration range 0.05-1.0 mg/mL, the correlation coefficient was 0.999 in all cases. The procedure was tested for river and ground waters containing a mixture of isopropyl, isobutyl, pinacolyl methylphosphonates and methylphosphonic acid. Relative standard deviations of the migration times under the conditions of the repeatability of the results lie in the range 1.9-2.7% (n = 5), the relative standard deviation of the reduced peak areas vary from 1.2 to 7.0%.
引用
收藏
页码:421 / 424
页数:4
相关论文
共 50 条
  • [1] Determination of alkyl methylphosphonates and methylphosphonic acid by capillary electrophoresis with a diode-array spectrophotometric detector
    S. S. Aleksenko
    S. V. Tret’yakova
    V. G. Mandych
    Journal of Analytical Chemistry, 2011, 66
  • [2] Effective separation and simultaneous determination of seven fluoroquinolones by capillary electrophoresis with diode-array detector
    Sun, Han-wen
    He, Pan
    Lv, Yun-kai
    Liang, Shu-xuan
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 852 (1-2): : 145 - 151
  • [3] Noise normalisation in capillary electrophoresis using a diode array detector
    Erny, Guillaume L.
    Calisto, Vania
    Esteves, Valdemar I.
    JOURNAL OF SEPARATION SCIENCE, 2011, 34 (14) : 1703 - 1707
  • [4] Determination of histamine in fish products by capillary electrophoresis and ion-pair liquid chromatography with diode-array detection
    Gardana, C
    Pietta, P
    Ciappellano, S
    Testolin, G
    JOURNAL OF THE SCIENCE OF FOOD AND AGRICULTURE, 1999, 79 (01) : 91 - 94
  • [5] Simultaneous determination of products and intermediates of L-dopa oxidation using capillary electrophoresis with diode-array detection
    Robinson, GM
    Smyth, MR
    ANALYST, 1997, 122 (08) : 797 - 802
  • [6] DETERMINATION OF LUTEOLIN AND APIGENIN IN HERBS BY CAPILLARY ELECTROPHORESIS WITH DIODE ARRAY DETECTION
    Maher, Hadir M.
    Al-Zoman, Nourah Z.
    Al-Shehri, Mona M.
    Al-Showiman, Hessa
    Al-Taweel, Areej M.
    Fawzy, Ghada A.
    Perveen, Shagufta
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2015, 43 (06) : 611 - 625
  • [7] Plasma creatinine and creatine quantification by capillary electrophoresis diode array detector
    Zinellu, A
    Caria, MA
    Tavera, C
    Sotgia, S
    Chessa, R
    Delana, L
    Carru, C
    ANALYTICAL BIOCHEMISTRY, 2005, 342 (02) : 186 - 193
  • [8] White LEDs as broad spectrum light sources for spectrophotometry: Demonstration in the visible spectrum range in a diode-array spectrophotometric detector
    Piasecki, Tomasz
    Breadmore, Michael C.
    Macka, Mirek
    ELECTROPHORESIS, 2010, 31 (22) : 3737 - 3744
  • [9] Simultaneous determination of the main metabolites in rice leaves using capillary electrophoresis mass spectrometry and capillary electrophoresis diode array detection
    Sato, S
    Soga, T
    Nishioka, T
    Tomita, M
    PLANT JOURNAL, 2004, 40 (01) : 151 - 163
  • [10] Determination of pKa values of quinolones from mobility and spectroscopic data obtained by capillary electrophoresis and a diode array detector
    Jiménez-Lozano, E
    Marqués, I
    Barrón, D
    Beltrán, JL
    Barbosa, J