Evaluation of the VIDAS Anti-HCV Assay for Detection of Hepatitis C Virus Infection

被引:6
作者
Hyun, Jungwon [1 ]
Ko, Dae-Hyun [1 ]
Kang, Hee Jung [1 ]
Whang, Dong Hee [2 ]
Cha, Young Joo [3 ]
Kim, Hyun Soo [1 ]
机构
[1] Hallym Univ, Coll Med, Dept Lab Med, Hwaseong, South Korea
[2] Inje Univ, Coll Med, Dept Lab Med, Seoul, South Korea
[3] Chung Ang Univ, Coll Med, Dept Lab Med, 102 Heukseok Ro, Seoul 06973, South Korea
关键词
Hepatitis C; Hepatitis C antibodies; Immunoassay; Immunoenzyme techniques; Immunoblotting; Real-time polymerase chain reaction; PERFORMANCE EVALUATION; BLOOD-DONORS; INDETERMINATE; RNA;
D O I
10.3343/alm.2016.36.6.550
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Anti-hepatitis C virus antibody (anti-HCV) assays are recommended for screening HCV-infected persons. The VIDAS Anti-HCV Assay (bioMerieux, France), based on the enzyme-linked fluorescence test principle, was recently introduced in Korea. We evaluated the clinical performance of the VIDAS assay. Methods: One hundred HCV-positive and 1,002 HCV-negative blood samples confirmed by Architect anti-HCV (Abbott Laboratories, USA) and COBAS TaqMan HCV real-time PCR (Roche Diagnostics, USA) or the Procleix Ultrio Plus Assay (Gen-Probe Incorporated, USA) were obtained from the Human Serum Bank (HSB) and tested by VIDAS. In case of discrepant results, we conducted a recombinant immunoblot assay (RIBA). Results: The agreement rates for known HCV-positive and HCV-negative samples between the VIDAS assay and the HSB testing were 100% (95% confidence interval [CI]: 96.4-100%) and 99.5% (95% CI: 98.8-99.8%), respectively. One of the five discrepant samples was positive for Core 2+ and NS3-2 2+ reactivity, two samples were negative, and the other two were indeterminate regarding NS4 2+ reactivity in RIBA. We observed a significant but weak positive correlation between the titers of VIDAS and Architect assays (r=0.315, P<0.001). Conclusions: The VIDAS anti-HCV assay, developed on the VIDAS automated immunoassay platform based on the ready-to-use, single-sample test concept may be useful in small-to-medium-sized laboratories. It showed good agreement with Architect anti-HCV and COBAS PCR assays and is therefore useful for detection of HCV infection. Weakly test-positive (ambiguous) samples require additional testing by another anti-HCV, RIBA, or HCV RNA assay.
引用
收藏
页码:550 / 554
页数:5
相关论文
共 18 条
[1]   Multicenter evaluation of a fully automated third-generation anti-HCV antibody screening test with excellent sensitivity and specificity [J].
Alborino, F. ;
Burighel, A. ;
Tiller, F. -W. ;
van Helden, J. ;
Gabriel, C. ;
Raineri, A. ;
Catapano, R. ;
Stekel, H. .
MEDICAL MICROBIOLOGY AND IMMUNOLOGY, 2011, 200 (02) :77-83
[2]  
Alter M.J., 2003, MMWR RECOMM REP, V52, P1
[3]  
Alter Miriam J, 2003, MMWR Recomm Rep, V52, P1
[4]  
Alter MJ., 2003, MMWR RECOMM REP, V52, P1
[5]  
[Anonymous], HEP C FACT SHEET 164
[6]   Evaluation of the new ARCHITECT anti-HCV screening test under routine laboratory conditions [J].
Berger, Annemarie ;
Rabenau, Holger ;
Allwinn, Regina ;
Doerr, Hans Wilhelm .
JOURNAL OF CLINICAL VIROLOGY, 2008, 43 (02) :158-161
[7]  
Getchell JP, 2013, MMWR-MORBID MORTAL W, V62, P362
[8]   Laboratory Diagnostics for Hepatitis C Virus Infection [J].
Kamili, Saleem ;
Drobeniuc, Jan ;
Araujo, Aufra C. ;
Hayden, Tonya M. .
CLINICAL INFECTIOUS DISEASES, 2012, 55 :S43-S48
[9]   The significance of third-generation HCV RIBA-indeterminate, RNA-negative results in voluntary blood donors screened with sequential third-generation immunoassays [J].
Kiely, P ;
Kay, D ;
Parker, S ;
Piscitelli, L .
TRANSFUSION, 2004, 44 (03) :349-358
[10]   Clinical Performance Evaluation of Four Automated Chemiluminescence Immunoassays for Hepatitis C Virus Antibody Detection [J].
Kim, Sinyoung ;
Kim, Jeong-Ho ;
Yoon, Seoyoung ;
Park, Youn-Hee ;
Kim, Hyon-Suk .
JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (12) :3919-3923