Comparation of the new rebound tonometer IOPen and the Goldmann tonometer, and their relationship to corneal properties

被引:18
作者
Moreno-Montanes, J. [1 ]
Gosende, I. [1 ]
Caire, J. [1 ]
Garcia-Granero, M. [2 ]
机构
[1] Univ Navarra Clin, Dept Ophthalmol, Pamplona 31008, Spain
[2] Univ Navarra, Dept Genet, Unidad Estadist, E-31080 Pamplona, Spain
关键词
corneal properties; goldmann tonometer; ocular response analyzer; rebound tonometer; OCULAR RESPONSE ANALYZER; INTRAOCULAR-PRESSURE; BIOMECHANICAL PROPERTIES; APPLANATION TONOMETER; THICKNESS; TONOPEN;
D O I
10.1038/eye.2010.137
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To compare the intraocular pressures (IOPs) obtained with the IOPen rebound tonometer, Goldmann applanation tonometer (GAT) and the ocular response analyzer (ORA) and investigate the effects of corneal biomechanical properties on IOPen measurements. Methods A total of 198 normal eyes were included in this cross-sectional and randomized study. Three measurements were taken using IOPen. Agreement between tonometers was calculated using the Bland and Altman limits of agreement (LoA) analysis. Results The median IOPen IOP was 3 mm Hg below the GAT (P<0.001), 3 mm Hg below the ORA IOP similar to Goldmann (IOPg), and 3 mm Hg below the ORA IOP corrected using corneal parameters (IOPcc)(P<0.01). The LoA width between the IOPen and GAT IOPs varied between 13.92 (mean IOPen IOP) and 15.99 mm Hg (third IOPen measurement). The central corneal thickness (CCT) was unrelated to IOPen measurements (P>0.05). Corneal hysteresis (CH) and corneal rigidity factor (CRF) were correlated with IOPen and GAT. Conclusions IOPen underestimated the IOP compared with GAT and ORA. The effect of measurement quality or measurement order on IOPen was low. CCT did not affect the IOPen, but the CH and CRF did. The LoA width between the IOPen and GAT IOPs was higher than between the ORA IOPg or ORA IOPcc and GAT IOPs. Eye (2011) 25, 50-56; doi: 10.1038/eye.2010.137; published online 1 October 2010
引用
收藏
页码:50 / 56
页数:7
相关论文
共 17 条
  • [1] Comparison of the ICare® rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists
    Abraham, L. M.
    Epasinghe, N. C. R.
    Selva, D.
    Casson, R.
    [J]. EYE, 2008, 22 (04) : 503 - 506
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients
    Brusini, P
    Salvetat, ML
    Zeppieri, M
    Tosoni, C
    Parisi, L
    [J]. JOURNAL OF GLAUCOMA, 2006, 15 (03) : 213 - 217
  • [4] Assessment of corneal biomechanical properties and intraocular pressure with the Ocular Response Analyzer in childhood myopia
    Chang, Pei-Yao
    Chang, Shu-Wen
    Wang, Jiun-Yi
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2010, 94 (07) : 877 - 881
  • [5] Assessment of true intraocular pressure: The gap between theory and practical data
    Chihara, Etsuo
    [J]. SURVEY OF OPHTHALMOLOGY, 2008, 53 (03) : 203 - 218
  • [6] The influence of cornea properties on rebound tonometry
    Chui, Wan-sang
    Lam, Andrew
    Chen, Davie
    Chiu, Roger
    [J]. OPHTHALMOLOGY, 2008, 115 (01) : 80 - 84
  • [7] Clinical evaluation of rebound tonometer
    Davies, LN
    Bartlett, H
    Mallen, EAH
    Wolffsohn, JS
    [J]. ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02): : 206 - 209
  • [8] Effect of corneal drying on central corneal thickness
    Dayanir, V
    Sakarya, R
    Özcura, F
    Kir, E
    Aktunç, T
    Özkan, BS
    Okyay, P
    [J]. JOURNAL OF GLAUCOMA, 2004, 13 (01) : 6 - 8
  • [9] Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population
    Fernandes, P
    Díaz-Rey, JA
    Queirós, A
    Gonzalez-Meijome, JM
    Jorge, J
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) : 436 - 440
  • [10] Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer
    Goldblum, D
    Kontiola, AI
    Mittag, T
    Chen, B
    Danias, J
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2002, 240 (11) : 942 - 946