Effect of CT slice thickness on accuracy of implant positioning in navigated total hip arthroplasty

被引:10
作者
Hirasawa, N. [1 ,2 ]
Matsubara, M.
Ishii, K.
Hagio, S.
Okuda, N.
Sekiya, I. [2 ]
Muneta, T. [2 ]
机构
[1] Nissan Tamagawa Gen Hosp, Ctr Adult Hip Reconstruct, Setagaya Ku, Tokyo, Japan
[2] Tokyo Med & Dent Univ, Grad Sch Med, Sect Orthoped Surg, Tokyo, Japan
关键词
CT-based navigation; total hip arthroplasty; radiation exposure; REPLACEMENT;
D O I
10.3109/10929088.2010.525025
中图分类号
R61 [外科手术学];
学科分类号
摘要
CT-based navigation appears to be very accurate with regard to cup anteversion and leg length, but inaccurate with regard to stem anteversion. As for the varus/valgus angle of the stem, the navigated approach seems to be very accurate in terms of the numerical value, but this does not satisfy us: Stem anteversion is still inaccurate with this system, while cup inclination is sufficiently accurate with both navigation and manual methods. Use of 1-mm CT slices results in twice the radiation exposure associated with 3-mm CT slices, but there is little difference with respect to accuracy. It is therefore recommended to use a CT-based navigation system with 3-mm CT slices for accurate and safe total hip arthroplasty.
引用
收藏
页码:83 / 89
页数:7
相关论文
共 14 条
[1]  
Bose WJ, 2000, ORTHOPEDICS, V23, P433
[2]  
GRUTZNER PA, 2004, INJURY S1, V35, pSA90
[3]  
HUBE R, 2003, SURG TECHNOL INT, V11, P273
[4]   Imageless navigation for insertion of the acetabular component in total hip arthroplasty -: Is it as accurate as CT-based navigation? [J].
Kalteis, T ;
Handel, A ;
Bäthis, H ;
Perlick, L ;
Tingart, M ;
Grifka, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2006, 88B (02) :163-167
[5]   DISLOCATIONS AFTER TOTAL HIP-REPLACEMENT ARTHROPLASTIES [J].
LEWINNEK, GE ;
LEWIS, JL ;
TARR, R ;
COMPERE, CL ;
ZIMMERMAN, JR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (02) :217-220
[6]   THE DEFINITION AND MEASUREMENT OF ACETABULAR ORIENTATION [J].
MURRAY, DW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (02) :228-232
[7]  
Röttinger H, 2006, ORTHOPADE, V35, P708, DOI 10.1007/s00132-006-0962-6
[8]   L-shaped caliper for limb length measurement during total hip arthroplasty [J].
Shiramizu, K ;
Naito, M ;
Shitama, T ;
Nakamura, Y ;
Shitama, H .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (07) :966-969
[9]  
Shrimpton P C., 2005, NRPB-W67 doses from computed tomography (CT) examinations in the UK-2003 review
[10]  
Sugano N, 2001, Comput Aided Surg, V6, P195, DOI 10.1002/igs.10011