Mechanical degradation of biological heart valve tissue induced by low diameter crimping: An early assessment

被引:30
作者
Khoffi, Foued [1 ]
Heim, Frederic [1 ]
机构
[1] ENSISA, Lab Phys & Mecan Text EAC CNRS 7189, F-68093 Mulhouse, France
关键词
Heart valve; Transcatheter; Pericardium leaflets; Crimping; LONG-TERM OUTCOMES; AORTIC-STENOSIS; IMPLANTATION; DURABILITY;
D O I
10.1016/j.jmbbm.2015.01.005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Transcatheter aortic valve implantation (TAVI) has become today an increasingly attractive procedure to relieve patients from aortic valve disease. However, the procedure requires crimping biological tissue within a metallic stent for low diameter catheter insertion purpose. This step induces specific stress in the leaflets especially when the crimping diameter is small. One concern about crimping is the potential degradations undergone by the biological tissue, which may limit the durability of the valve once implanted. The purpose of the present work is to study the effect of low diameter crimping on the mechanical performances of pericardium valve prototypes. The prototypes were compressed to a diameter of 1 mm within braided stents for 20 min. SEM observations performed on crimped material show that crimped leaflets undergo degradations characterized by apparent surface defects. Moreover mechanical extension tests were performed on pericardium strips before and after crimping. The strips (15 mm long, 5 mm wide) were taken from both crimped and native leaflets considering 2 different valve diameters, 19 and 21 mm. In order to prevent the premature drying of the pericardium tissue during the procedure, the biological tissue was kept in contact with a formaldehyde solution. Results show that the ultimate strength value decreases nearly by up to 50%. The modifications observed in the material may jeopardize the long term durability of the device. However, further tests are necessary with a larger amount of samples to confirm these early results. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:71 / 75
页数:5
相关论文
共 11 条
[1]   The Effects of Transcatheter Valve Crimping on Pericardial Leaflets [J].
Alavi, S. Hamed ;
Groves, Elliott M. ;
Kheradvar, Arash .
ANNALS OF THORACIC SURGERY, 2014, 97 (04) :1260-1266
[2]   Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis - First human case description [J].
Cribier, A ;
Eltchaninoff, H ;
Bash, A ;
Borenstein, N ;
Tron, C ;
Bauer, F ;
Derumeaux, G ;
Anselme, F ;
Laborde, F ;
Leon, MB .
CIRCULATION, 2002, 106 (24) :3006-3008
[3]   Percutaneous therapies for valvular heart disease [J].
Davidson, Michael J. ;
White, Jennifer K. ;
Baim, Donald S. .
CARDIOVASCULAR PATHOLOGY, 2006, 15 (03) :123-129
[4]  
Khoffi F., 2014, J Biomed Mater Res B Appl Biomater
[5]   Crimping May Affect the Durability of Transcatheter Valves: An Experimental Analysis [J].
Kiefer, Philipp ;
Gruenwald, Felix ;
Kempfert, Joerg ;
Aupperle, Heike ;
Seeburger, Joerg ;
Mohr, Friedrich Wilhelm ;
Walther, Thomas .
ANNALS OF THORACIC SURGERY, 2011, 92 (01) :155-160
[6]   Percutaneous pulmonary valve replacement: autologous tissue-engineered valved stents [J].
Metzner, Anja ;
Stock, Ulrich A. ;
Iino, Kenji ;
Fischer, Gunther ;
Huemme, Tim ;
Boldt, Jessica ;
Braesen, Jan Hinrich ;
Bein, Berthold ;
Renner, Jochen ;
Cremer, Jochen ;
Lutter, Georg .
CARDIOVASCULAR RESEARCH, 2010, 88 (03) :453-461
[7]   Long-Term Outcomes After Transcatheter Aortic Valve Implantation in High-Risk Patients With Severe Aortic Stenosis The UK TAVI (United Kingdom Transcatheter Aortic Valve Implantation) Registry [J].
Moat, Neil E. ;
Ludman, Peter ;
de Belder, Mark A. ;
Bridgewater, Ben ;
Cunningham, Andrew D. ;
Young, Christopher P. ;
Thomas, Martyn ;
Kovac, Jan ;
Spyt, Tom ;
MacCarthy, Philip A. ;
Wendler, Olaf ;
Hildick-Smith, David ;
Davies, Simon W. ;
Trivedi, Uday ;
Blackman, Daniel J. ;
Levy, Richard D. ;
Brecker, Stephen J. D. ;
Baumbach, Andreas ;
Daniel, Tim ;
Gray, Huon ;
Mullen, Michael J. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 58 (20) :2130-2138
[8]   Long-Term Outcomes After Transcatheter Aortic Valve Implantation Insights on Prognostic Factors and Valve Durability From the Canadian Multicenter Experience [J].
Rodes-Cabau, Josep ;
Webb, John G. ;
Cheung, Anson ;
Ye, Jian ;
Dumont, Eric ;
Osten, Mark ;
Feindel, Christopher M. ;
Natarajan, Madhu K. ;
Velianou, James L. ;
Martucci, Giussepe ;
DeVarennes, Benoit ;
Chisholm, Robert ;
Peterson, Mark ;
Thompson, Christopher R. ;
Wood, David ;
Toggweiler, Stefan ;
Gurvitch, Ronen ;
Lichtenstein, Samuel V. ;
Doyle, Daniel ;
DeLarochelliere, Robert ;
Teoh, Kevin ;
Chu, Victor ;
Bainey, Kevin ;
Lachapelle, Kevin ;
Cheema, Asim ;
Latter, David ;
Dumesnil, Jean G. ;
Pibarot, Philippe ;
Horlick, Eric .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 60 (19) :1864-1875
[9]  
Scheuer M, 2012, J MAT SCI RES, V1, P3
[10]   5-Year Outcome After Transcatheter Aortic Valve Implantation [J].
Toggweiler, Stefan ;
Humphries, Karin H. ;
Lee, May ;
Binder, Ronald K. ;
Moss, Robert R. ;
Freeman, Melanie ;
Ye, Jian ;
Cheung, Anson ;
Wood, David A. ;
Webb, John G. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 61 (04) :413-419