Large scan field, high spatial resolution flat-panel detector based volumetric CT of the whole human skull base and for maxillofacial imaging

被引:23
作者
Bartling, S. H.
Majdani, O.
Gupta, R.
Rodt, T.
Dullin, C.
Fitzgerald, P. F.
Becker, H.
机构
[1] Hannover Med Sch, Dept Neuroradiol, D-30625 Hannover, Germany
[2] Hannover Med Sch, Dept Otorhinolaryngol, D-30625 Hannover, Germany
[3] Massachusetts Gen Hosp, Boston, MA 02114 USA
[4] Hannover Med Sch, Dept Neurosurg, D-30625 Hannover, Germany
[5] Goettingen Univ Hosp, Dept Diagnost Radiol, D-37075 Gottingen, Germany
[6] Syst & Applicat Lab, GE Global Res, Moscow 123098, Russia
关键词
computerized tomography; skull base; maxillofacial; flat-panel detector;
D O I
10.1259/dmfr/19164138
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To assess the feasibility of flat-panel detector based volumetric CT (fpVCT) scanning of the whole human skull base and maxillofacial region, which has thus far only been demonstrated on small, excised specimens. Flat-panel detectors offer more favourable imaging properties than image intensifiers. It is therefore likely that they will replace them in cone-beam CT scanners that are currently used to scan parts of the skull base and maxillofacial region. Furthermore, the resolution of current CT imaging limits diagnosis, surgical planning and intraoperative navigation within these regions. fpVCT might overcome these limitations because it offers higher resolution of high contrast structures than current CT. Methods: Three embalmed cadaver heads were scanned in two scanners: an experimental fpVCT that offers a scan field large enough for a whole human head, and in a current multislice CT (MSCT). 28 structures were compared. Results: Both scanners produced bone images of diagnostic quality. Small high contrast structures such as parts of the ossicular chain and thin bony laminas were better delineated in fpVCT than in MSCT. fpVCT of maxillofacial region and skull base was rated superior to MSCT (P = 0.002) as found in this limited, experimental study. Conclusions: High spatial resolution fpVCT scanning of both regions in a whole human head is feasible and might be slightly superior to MSCT. fpVCT could improve diagnostic accuracy in selected cases, as well as surgical planning and intraoperative navigation accuracy.
引用
收藏
页码:317 / 327
页数:11
相关论文
共 44 条
  • [1] Using a flat-panel detector in high resolution cone beam CT for dental imaging
    Baba, R
    Ueda, K
    Okabe, M
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2004, 33 (05) : 285 - 290
  • [2] Comparison of flat-panel detector and image-intensifier detector for cone-beam CT
    Baba, R
    Konno, Y
    Ueda, K
    Ikeda, S
    [J]. COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2002, 26 (03) : 153 - 158
  • [3] High-resolution flat-panel volume-CT of temporal bone - Part 1 - Axial preoperative anatomy
    Bartling, SH
    Shukla, V
    Becker, H
    Brady, TJ
    Hayman, A
    Gupta, R
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2005, 29 (03) : 420 - 423
  • [4] BASE OF SKULL - COMPARISON OF COMPUTED AND CONVENTIONAL TOMOGRAPHY
    BECKER, H
    GRAU, H
    HACKER, H
    PLODER, KW
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (01) : 113 - 118
  • [5] Benndorf G, 2005, AM J NEURORADIOL, V26, P1813
  • [6] Computer-assisted intraoperative navigation during skull base surgery
    Carrau, RL
    Snyderman, CH
    Curtin, HD
    Janecka, IP
    Stechison, M
    Weissman, JL
    [J]. AMERICAN JOURNAL OF OTOLARYNGOLOGY, 1996, 17 (02) : 95 - 101
  • [7] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [8] FITZGERALD P, 2004, P RSNA, pSSC17
  • [9] Fuchs T, 2003, PHYS MEDICA, V19, P153
  • [10] CT detection of facial canal dehiscence and semicircular canal fistula: Comparison with surgical findings
    Fuse, T
    Tada, Y
    Aoyagi, M
    Sugai, Y
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (02) : 221 - 224