Fit and seal characteristics of a new paediatric tracheal tube with high volume-low pressure polyurethane cuff

被引:82
作者
Dullenkopf, A [1 ]
Gerber, AC [1 ]
Weiss, M [1 ]
机构
[1] Univ Zurich, Childrens Hosp, Dept Anaesthesia, CH-8032 Zurich, Switzerland
关键词
children; paediatric; croup; sealing; tracheal; cuff; high volume-low pressure; morbidity; tracheal tube;
D O I
10.1111/j.1399-6576.2005.00599.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background: To evaluate a new paediatric tracheal tube (Microcuff, Weinheim, Germany) with an ultrathin high volume-low pressure polyurethane cuff. Methods: With approval of the Hospital Ethics Committee tracheas of children undergoing general anaesthesia were intubated using a Microcuff tube. Tube sizes were selected according to: internal diameter (mm) = age/4 + 3.5 in children aged greater than or equal to2 years. In newborns (greater than or equal to3 kg) less than or equal to1 year, ID 3.0-mm tubes, and in children from 1 to 2 years, internal diameter 3.5-mm tubes were used. Tubes were classified too large if no air leakage was obtained at an airway pressure of 20 cm H2O with the cuff not inflated. Sealing pressure was assessed by auscultation. Post-extubation croup requiring therapy was noted. Results: Five-hundred children were studied. In eight children the tubes were too large. Sealing pressure was 9.7 +/- 2.5 cm H2O (4-20). In two patients postextubation croup required singular short-term therapy. Conclusion: Microcuff paediatric tracheal tubes provided tracheal sealing with cuff pressures considerably lower than usually accepted. The rate of tube exchange was very low (1.6%), as was the rate of airway morbidity (croup requiring therapy; 0.4%).
引用
收藏
页码:232 / 237
页数:6
相关论文
共 31 条
[1]  
CORFIELD HMC, 1963, BRIT J ANAESTH, V35, P34
[2]   CUFFED ENDOTRACHEAL-TUBES IN PEDIATRIC INTENSIVE-CARE [J].
DEAKERS, TW ;
REYNOLDS, G ;
STRETTON, M ;
NEWTH, CJL .
JOURNAL OF PEDIATRICS, 1994, 125 (01) :57-62
[3]   Tracheal sealing characteristics of pediatric cuffed tracheal tubes [J].
Dullenkopf, A ;
Schmitz, A ;
Gerber, AC ;
Weiss, M .
PEDIATRIC ANESTHESIA, 2004, 14 (10) :825-830
[4]   Air leakage around endotracheal tube cuffs [J].
Dullenkopf, A ;
Schmitz, A ;
Frei, M ;
Gerber, AC ;
Weiss, M .
EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2004, 21 (06) :448-453
[5]   Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube [J].
Dullenkopf, A ;
Gerber, A ;
Weiss, M .
INTENSIVE CARE MEDICINE, 2003, 29 (10) :1849-1853
[6]   Prediction of tracheal tube size in children using multiple variables [J].
Eck, JB ;
Dear, GD ;
Phillips-Bute, BG ;
Ginsberg, B .
PAEDIATRIC ANAESTHESIA, 2002, 12 (06) :495-498
[7]  
Erb T, 2001, ANAESTHESIST, V50, P395, DOI 10.1007/s001010100158
[8]   Endotracheal tube cuff pressure is unpredictable in children [J].
Felten, MAL ;
Schmautz, E ;
Delaporte-Cerceau, S ;
Orliaguet, GA ;
Carli, PA .
ANESTHESIA AND ANALGESIA, 2003, 97 (06) :1612-1616
[9]  
Fine GF, 2000, ANESTHESIOLOGY, V93, pU227
[10]  
Fine GF, 2004, PEDIATR ANESTH, V14, P38, DOI 10.1046/j.1460-9592.2003.01203.x