Outcomes of midterm circulatory support by left ventricular assist device implantation with descending aortic anastomosis

被引:19
作者
Nawata, Kan [1 ]
Nishimura, Takashi [2 ]
Kyo, Shunei [2 ]
Hisagi, Motoyuki [2 ]
Kinoshita, Osamu [1 ]
Saito, Aya [1 ]
Motomura, Noboru [1 ]
Takamoto, Shinichi [3 ]
Ono, Minoru [1 ]
机构
[1] Tokyo Univ Hosp, Dept Cardiothorac Surg, Bunkyo Ku, Tokyo 113, Japan
[2] Univ Tokyo, Grad Sch Med, Dept Therapeut Strategy Heart Failure, Tokyo, Japan
[3] Mitsui Mem Hosp, Tokyo 101, Japan
关键词
Left ventricular assist device; Descending aortic anastomosis; Toyobo; Jarvik; 2000; Intermittent low-speed mode; FLOW; TRANSPLANTATION; CANNULATION; INFECTION; PULSATILE;
D O I
10.1007/s10047-010-0521-0
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For some patients undergoing left ventricular assist device (LVAD) implantation, the perfusion tube is anastomosed to the descending aorta instead of the currently more prevalently used ascending aorta. Purpose of this study was to assess retrospectively the outcomes of LVAD patients with descending aortic anastomosis. Between March 2007 and March 2010, six patients underwent LVAD implantation with descending aortic anastomosis with Toyobo or Jarvik 2000 LVAD at our institute. Their average circulatory support time was 434 (range 82-751) days. Both types of LVAD afforded adequate circulatory support, and inotrope treatment and mechanical ventilation were discontinued relatively early. Echocardiograms of the three patients with Jarvik 2000 LVAD revealed antegrade flow in the ascending aorta during the intermittent low-speed period. Among them, one patient developed infarction in the right brain hemisphere because of thromboembolism, whereas another patient developed pneumonia in the left lung followed by a lethal systemic infection. One patient on Toyobo LVAD support reached heart transplantation without morbidity. Another patient implanted with Toyobo LVAD, whose left ventricular function was too poor to generate forward flow through aortic valve, developed thrombus in the ascending aorta. No embolic events were observed in the organs below the diaphragm. In conclusion, descending aortic anastomosis of the perfusion tube can be used for LVAD implantation for some patients, but considerable risks of morbidities, including thromboembolic events and/or infection, should be recognized.
引用
收藏
页码:197 / 201
页数:5
相关论文
共 11 条
[1]   Retention of left ventricular assist device outflow grafts after transplantation [J].
Cohn, William E. ;
Fikfak, Vid ;
Gregoric, Igor D. ;
Frazier, O. H. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2008, 27 (08) :865-868
[2]   In vitro characterization of aortic retrograde and antegrade flow from pulsatile and non-pulsatile ventricular assist devices [J].
DiGiorgi, PL ;
Smith, DL ;
Naka, Y ;
Oz, MC .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2004, 23 (02) :186-192
[3]   Midterm experience with the Jarvik 2000 axial flow left ventricular assist device [J].
Haj-Yahia, Saleem ;
Birks, Emma J. ;
Rogers, Paula ;
Bowles, Christopher ;
Hipkins, Mandy ;
George, Robert ;
Amrani, Mohammed ;
Petrou, Mario ;
Pepper, John ;
Dreyfus, Gilles ;
Khaghani, Asghar .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2007, 134 (01) :199-203
[4]   Infection after implantation of pulsatile mechanical circulatory support devices [J].
Holman, William L. ;
Kirklin, James K. ;
Naftel, David C. ;
Kormos, Robert L. ;
Desvign-Nickens, Patricia ;
Camacho, Margarita T. ;
Ascheim, Deborah D. .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (06) :1632-U309
[5]   Femoral versus aortic cannulation for surgery of chronic ascending aortic aneurysm [J].
Lakew, F ;
Pasek, P ;
Zacher, M ;
Diegeler, A ;
Urbanski, PP .
ANNALS OF THORACIC SURGERY, 2005, 80 (01) :84-89
[6]   Effects of left ventricular assist device support and outflow graft location upon aortic blood flow [J].
Litwak, KN ;
Koenig, SC ;
Tsukui, H ;
Kihara, S ;
Wu, ZJ ;
Pantalos, GM .
ASAIO JOURNAL, 2004, 50 (05) :432-437
[7]   Pseudoaneurysm after heart transplantation with history of LVAD driveline infection [J].
Omoto, T ;
Minami, K ;
Muramatsu, T ;
Kyo, S ;
Körfer, R .
ANNALS OF THORACIC SURGERY, 2001, 72 (01) :263-264
[8]   Left ventricular assist device implantation via left thoracotomy: Alternative to repeat sternotomy [J].
Pierson, RN ;
Howser, R ;
Donaldson, T ;
Merrill, WH ;
Dignan, RJ ;
Drinkwater, DC ;
Christian, KG ;
Butler, J ;
Chomsky, D ;
Wilson, JR ;
Clark, R ;
Davis, SF .
ANNALS OF THORACIC SURGERY, 2002, 73 (03) :997-999
[9]   Does the arterial cannulation site for circulatory arrest influence stroke risk? [J].
Svensson, LG ;
Blackstone, EH ;
Rajeswaran, J ;
Sabik, JF ;
Lytle, BW ;
Gonzalez-Stawinski, G ;
Varvitsiotis, P ;
Banbury, MK ;
McCarthy, PM ;
Pettersson, GB ;
Cosgrove, DM .
ANNALS OF THORACIC SURGERY, 2004, 78 (04) :1274-1284
[10]   Ventricular assist device support in patients with mechanical heart valves [J].
Swartz, MT ;
Lowdermilk, GA ;
Moroney, DA ;
McBride, LR .
ANNALS OF THORACIC SURGERY, 1999, 68 (06) :2248-2251