True volumetric method for flow cytometric enumeration of CD34+ stem cells and its agreement with a standard bead-based single-platform protocol

被引:3
作者
Mortazavi, Samira [1 ]
Ardalan, Farid Azmoudeh [1 ]
Nodehi, Seyyed Reza Safaee [2 ]
Karder, Fahimeh Firouzjaie [3 ]
Miraliakbari, Nafisehsadat [3 ]
机构
[1] Univ Tehran Med Sci, Dept Pathol, Tehran 1419733141, Iran
[2] Univ Tehran Med Sci, Dept Hematol & Oncol, Vali E Asr Hosp, Tehran 1419733141, Iran
[3] Univ Tehran Med Sci, Flow Cytometry Div, Vali E Asr Clin Lab, Tehran 1419733141, Iran
关键词
CD34; antigen; enumeration; flow cytometry; stem cell; transplantation; HEMATOPOIETIC PROGENITOR CELLS; CORD BLOOD; CLINICAL TRANSPLANTATION; BD FACSCALIBUR; COUNTS; MARROW; ASSAY;
D O I
10.3109/14653249.2012.667875
中图分类号
Q813 [细胞工程];
学科分类号
摘要
Background aims. Stem cells are commonly enumerated with bead-based methods in blood and marrow progenitor cell transplantation centers. We compared the International Society of Hematotherapy and Graft Engineering (ISHAGE) bead-based method with a true volumetric one that obviates the use of fluorescent beads for enumeration. Methods. From 31 samples, including 15 peripheral blood samples and 16 leukapheresis products, CD34(+) cells were enumerated with the single-platform bead-based ISHAGE method and a true volumetric method. After exclusion of two outliers, one from the peripheral blood group and the other from the leukapheresis group, the results were compared. Results. In the peripheral blood category, no significant difference was observed. However, a proportional systematic error was seen in the leukapheresis group. The systematic error was corrected in the leukapheresis group using a regression line equation. The 95% confidence interval of differences was [-5.83, 2.18] for the peripheral blood and [-38.40, 38.77] for the leukapheresis group after correction of the systematic error. Conclusions. The true volumetric method is a simple and reliable approach that can be used instead of the more popular bead-based procedures.
引用
收藏
页码:621 / 629
页数:9
相关论文
共 41 条
  • [1] Barbosa IL, 1999, CYTOMETRY, V38, P274, DOI 10.1002/(SICI)1097-0320(19991215)38:6<274::AID-CYTO3>3.3.CO
  • [2] 2-9
  • [3] Barnett D, 1999, BRIT J HAEMATOL, V106, P1059
  • [4] Guideline for the flow cytometric enumeration of CD34+ haematopoietic stem cells
    Barnett, D
    Janossy, G
    Lubenko, A
    Matutes, E
    Newland, A
    Reilly, JT
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 1999, 21 (05): : 301 - 308
  • [5] Barnett D, 1998, BRIT J HAEMATOL, V102, P553
  • [6] Measurement error
    Bland, JM
    Altman, DG
    [J]. BRITISH MEDICAL JOURNAL, 1996, 312 (7047) : 1654 - 1654
  • [7] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [8] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [9] Brando B, 2001, CYTOMETRY, V43, P154, DOI 10.1002/1097-0320(20010201)43:2<154::AID-CYTO1031>3.0.CO
  • [10] 2-2