Evaluation of the Siemens ADVIA Centaur high-sensitivity cardiac troponin I assay in serum

被引:28
作者
Kavsak, Peter A. [1 ,4 ]
Worster, Andrew [2 ]
Hill, Stephen A. [1 ]
Jaffe, Allan S. [3 ]
机构
[1] McMaster Univ, Dept Pathol & Mol Med, Hamilton, ON, Canada
[2] McMaster Univ, Div Emergency Med, Hamilton, ON, Canada
[3] Hamilton Hlth Sci, Hamilton, ON, Canada
[4] Mayo Clin, Rochester, MN USA
关键词
High-sensitivity cardiac troponin; Analytical comparison; Storage; Stability; Agreement; Myocardial infarction; OUTCOMES; CUTOFFS;
D O I
10.1016/j.cca.2018.10.012
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: International recommendations on high-sensitivity cardiac troponin (hs-cTn) testing recommend that laboratories select only one sample type for testing. We evaluated the Siemens ADVIA Centaur hs-cTnI assay in serum and thereby provide information on imprecision, long-term storage stability, freeze-thaw stability, method comparison to other hs-cTnI assays, and clinical performance. Methods: Patients with chest pain onset < 6 h who already had Roche hs-cTnT, Beckman hs-cTnI and Abbott hs-cTnI results recorded and had non-thawed and frozen serum aliquots formed the study population (n = 134 patients with 305 serum aliquots obtained at either 0, 3 or 6 h stored below-70 degrees C since 2003) for measurement with the Siemens hs-cTnI assay in 2018. Receiver-operating characteristic curve analyses for myocardial infarction (MI) using the highest obtained hs-cTn concentration was performed. Additional comparison testing on serum samples stored frozen (at-70 degrees C for < 1 month in 2018) for the Siemens and Abbott hs-cTnI assays were performed, as well as precision testing in serum pools and freeze-thaw stability testing. Results: The Siemens hs-cTnI assay had an area under the curve (AUC) of 0.978 (95%CI: 0.937-0.996) for MI in the study cohort (Roche hs-cTnT AUC = 0.965 and Abbott AUC = 0.973). The Siemens hs-cTnI assay yielded higher cTnI concentrations than the other hs-cTn assays, with the same proportional bias (slope = 1.4) between the Siemens and Abbott hs-cTnI assays obtained from serum samples collected in 2003 and 2018. Over 3 months, a low serum pool of 3.5 ng/l achieved a CV of 20% (SD = 0.7, n = 42) and a high serum pool of 820 ng/l achieved a CV of 2.3% (SD = 20, n = 42). Three different serum pools recovered within 10% from baseline concentration after 5 freeze-thaw cycles for the Siemens hs-cTnI assay. Conclusions: In serum, the Siemens ADVIA Centaur hs-cTnI assay had excellent clinical performance for MI in an early chest pain onset population, acceptable precision at normal and highly elevated cTnI concentrations, long-term storage stability (15 y storage below-70 degrees C) and acceptable freeze-thaw stability, all of which supports serum as an acceptable sample type to use in clinical studies and in clinical practice.
引用
收藏
页码:216 / 221
页数:6
相关论文
共 16 条
[1]   Clinical Validation of a Novel High-Sensitivity Cardiac Troponin I Assay for Early Diagnosis of Acute Myocardial Infarction [J].
Boeddinghaus, Jasper ;
Twerenbold, Raphael ;
Nestelberger, Thomas ;
Badertscher, Patrick ;
Wildi, Karin ;
Puelacher, Christian ;
de Lavallaz, Jeanne du Fay ;
Keser, Elif ;
Gimenez, Maria Rubini ;
Wussler, Desiree ;
Kozhuharov, Nikola ;
Rentsch, Katharina ;
Miro, Oscar ;
Martin-Sanchez, F. Javier ;
Morawiec, Beata ;
Stefanelli, Sabrina ;
Geigy, Nicolas ;
Keller, Dagmar I. ;
Reichlin, Tobias ;
Mueller, Christian .
CLINICAL CHEMISTRY, 2018, 64 (09) :1347-1360
[2]  
Eggers KM., 2018, Clin Chem
[3]   Should detectable cardiac troponin concentrations in a healthy population be the only criterion for classifying high-sensitivity cardiac troponin assays? [J].
Kavsak, Peter A. .
CLINICAL BIOCHEMISTRY, 2018, 56 :1-3
[4]   Assessing matrix, interferences and comparability between the Abbott Diagnostics and the Beckman Coulter high-sensitivity cardiac troponin I assays [J].
Kavsak, Peter A. ;
Malinowski, Paul ;
Roy, Chantele ;
Clark, Lorna ;
Lamers, Shana .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (07) :1176-1181
[5]   Analytical comparison of three different versions of a high-sensitivity cardiac troponin I assay over 10 years [J].
Kavsak, Peter A. ;
Worster, Andrew ;
Hill, Stephen A. ;
MacRae, Andrew R. ;
Jaffe, Allan S. .
CLINICA CHIMICA ACTA, 2017, 475 :51-55
[6]   High-Sensitivity Cardiac Troponin Risk Cutoffs for Acute Cardiac Outcomes at Emergency Department Presentation [J].
Kavsak, Peter A. ;
Worster, Andrew ;
Ma, Jinhui ;
Shortt, Colleen ;
Clayton, Natasha ;
Sherbino, Jonathan ;
Hill, Stephen A. ;
McQueen, Matthew ;
Mehta, Shamir R. ;
Devereaux, Philip J. .
CANADIAN JOURNAL OF CARDIOLOGY, 2017, 33 (07) :898-903
[7]  
Kavsak Peter A, 2015, Pract Lab Med, V1, P28, DOI 10.1016/j.plabm.2015.02.001
[8]   Biomarkers for Predicting Serious Cardiac Outcomes at 72 Hours in Patients Presenting Early after Chest Pain Onset with Symptoms of Acute Coronary Syndromes [J].
Kavsak, Peter A. ;
Hill, Stephen A. ;
Supapol, Wendy Bhanich ;
Devereaux, Philip J. ;
Worster, Andrew .
CLINICAL CHEMISTRY, 2012, 58 (01) :298-302
[9]   Simulation Models of Misclassification Error for Single Thresholds of High-Sensitivity Cardiac Troponin I Due to Assay Bias and Imprecision [J].
Lyon, Andrew W. ;
Kavsak, Peter A. ;
Lyon, Oliver A. S. ;
Worster, Andrew ;
Lyon, Martha E. .
CLINICAL CHEMISTRY, 2017, 63 (02) :585-592
[10]   Multicenter comparison of imprecision at low concentrations of two regulatory approved high-sensitivity cardiac troponin I assays [J].
Raizman, Joshua E. ;
Fuezery, Anna ;
Tsui, Albert K. Y. ;
Higgins, Trefor ;
Goudreau, Bobbi Lynn ;
Clark, Lorna ;
Hill, Stephen A. ;
Kavsak, Peter A. .
CLINICA CHIMICA ACTA, 2018, 486 :219-220