Compression and flexural strength of bone cement mixed with blood

被引:21
|
作者
Tan, J. H. [1 ]
Koh, Bryan T. H. [2 ]
Ramruttun, A. K. [1 ]
Wang, Wilson [1 ]
机构
[1] Natl Univ Hlth Syst, Dept Orthoped Surg, Singapore, Singapore
[2] Natl Univ Singapore, Yong Loo Lin Sch Med, Singapore, Singapore
关键词
bone cements; compressive strength; polym ethyl methacrylate; TOTAL KNEE ARTHROPLASTY; TOTAL HIP-ARTHROPLASTY; FOLLOW-UP; PHYSICAL-PROPERTIES; FEMORAL COMPONENTS; SUCTION TECHNIQUE; TIBIAL COMPONENT; REPLACEMENT; INTERFACE; STEM;
D O I
10.1177/1602400223
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose. To assess the compression and flexural strength of bone cement mixed with 0 ml, 1 ml, or 2 ml of blood. Methods. High viscosity polymethyl methacrylate (PMMA) loaded with or without gentamicin was used. Blood was collected from total knee arthroplasty patients. In the same operating room, one pack of cement each was mixed with 0 ml (control), 1 ml, or 2 ml of blood for 1 minute during the dough phase. The dough was extruded into cylindrical and rectangular moulds for 20 minutes of setting, and then cured in phosphate buffered saline at 37 +/- 1 degrees C for 7 days. The samples were visually inspected for fractures and areas of weakness, and then scanned using micro computed tomography. 48 gentamicin-loaded and 59 non-gentamicin-loaded samples mixed with 0 ml (control), 1 ml, or 2 ml of blood were randomised for flexural and compression strength testing; each group had at least 6 samples. Results. In samples loaded with or without gentamicin, the flexural and compressive strength was highest in controls, followed by samples mixed with 1 ml or 2 ml of blood. In samples mixed with 2 ml of blood, the flexural strength fell below the standard of 50 MPa. In samples mixed with 2 ml of blood and all gentamicin-loaded samples, the compressive strength fell below the standard of 70 MPa. Micro computed tomography revealed areas of voids and pores indicating the presence of laminations and partitions within. Conclusion. The biomechanical strength of PMMA contaminated with blood may decrease. Precautions such as saline lavage, pack drying the bone, change of gloves, and prompt insertion of the implant should be taken to prevent blood from contaminating bone cement.
引用
收藏
页码:240 / 244
页数:5
相关论文
共 50 条
  • [1] Factors Affecting Flexural Strength in Cement Within Cement Revisions
    Dang, Kieran
    Pelletier, Matthew H.
    Walsh, William R.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (08) : 1540 - 1548
  • [2] Influence of load type on flexural strength of a bone cement based on PMMA
    Vallo, CI
    POLYMER TESTING, 2002, 21 (07) : 793 - 800
  • [3] The effect of isotonic saline on the strength of bone cement during the polymerisation period
    Pearce, Christopher J.
    Patel, Shelain
    Sexton, Shaun A.
    Reeves, Alexander
    Khairandish, Hamidreza
    Khaleel, Arshad
    HIP INTERNATIONAL, 2011, 21 (02) : 238 - 242
  • [4] Effects of Fly Ash Addition on Compressive Strength and Flexural Strength of Foamed Cement CompoSites
    Wahab, Roshasmawi Abdul
    Noor, Mazlee Mohd
    Jamaludin, Shamsul Baharin
    Ismail, Khairul Nizar
    2ND INTERNATIONAL CONFERENCE ON SUSTAINABLE MATERIALS (ICOSM 2013), 2013, 795 : 664 - +
  • [5] The effect of application time of two types of bone cement on the cement–bone interface strength
    Dahabreh Z.
    Phillips H.K.
    Stewart T.
    Stone M.
    European Journal of Orthopaedic Surgery & Traumatology, 2015, 25 (4) : 775 - 781
  • [6] Compression Strength and Porosity of Single-Antibiotic Cement Vacuum-Mixed With Vancomycin
    Brock, Hugh S.
    Moodie, Patrick G.
    Hendricks, Kelly J.
    McIff, Terence E.
    JOURNAL OF ARTHROPLASTY, 2010, 25 (06) : 990 - 997
  • [7] Strength and deformation properties of compacted cement-mixed gravel evaluated by triaxial compression tests
    Lohani, TN
    Kongsukprasert, L
    Watanabe, K
    Tatsuoka, F
    SOILS AND FOUNDATIONS, 2004, 44 (05) : 95 - 108
  • [8] Bone cement usage modalities at a multi-site university hospital centre
    Barral, Marine
    Lalande, Laure
    Viste, Anthony
    Besse, Jean-Luc
    Fessy, Michel-Henri
    Carre, Emmanuelle
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2019, 105 (02) : 383 - 388
  • [9] Application of expert systems in prediction of flexural strength of cement mortars
    Gulbandilar, Eyyup
    Kocak, Yilmaz
    COMPUTERS AND CONCRETE, 2016, 18 (01) : 1 - 16
  • [10] The cement-bone bond is weaker than cement-cement bond in cement-in-cement revision arthroplasty. A comparative biomechanical study
    Ceynowa, Marcin
    Zerdzicki, Krzysztof
    Klosowski, Pawel
    Zrodowski, Maciej
    Pankowski, Rafal
    Roclawski, Marek
    Mazurek, Tomasz
    PLOS ONE, 2021, 16 (02):