Calibration and Error Analysis of Surgical Instrument Based on Stereo Camera

被引:2
作者
Cao, Li [1 ]
Zhang, Yong [1 ]
Wei, Jun [1 ]
Hu, Yan [1 ]
Liu, Da [2 ]
机构
[1] Univ Jinan, Sch Control Sci & Engn, Jinan, Peoples R China
[2] Beijing Univ, Inst Robot, Beijing, Peoples R China
来源
2010 3RD INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2010), VOLS 1-7 | 2010年
关键词
instrument positioning system; calibration; stereo camera; least squares; error analysis;
D O I
10.1109/BMEI.2010.5639405
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper proposes a method to calibrate and position surgical instruments using stereo camera, discusses the calibration error and its propagation when the instrument is positioned. A planar marker composed by X corners is attached to the instrument, and a reference frame is established on the marker. The structural parameter of instrument is calibrated on clinical site by rotating the instrument around its tip on proposed calibration template; thereafter the instruments can be distinguished and positioned by stereo camera. The method is proved by applying it in a simple instrument positioning system. The theoretical analysis and experiments reveal that calibration error depends not only on the accuracy of marker localization, but also on the marker configuration and the tip position relative to the marker reference frame. Furthermore, we analyze the calibration error propagation and find out that the instruments positioning error do not exceed the sum of calibration error and markers localization error when marker is configured properly.
引用
收藏
页码:1435 / 1439
页数:5
相关论文
共 13 条
[1]   Systems for tracking minimally invasive surgical instruments [J].
Chmarra, M. K. ;
Grimbergen, C. A. ;
Dankelman, J. .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2007, 16 (06) :328-340
[2]  
Gordon RA, 2009, REAL ANAL EXCH, V35, P205
[3]   Binocular vision and motion-in-depth [J].
Harris, Julie M. ;
Nefs, Harold T. ;
Grafton, Catherine E. .
SPATIAL VISION, 2008, 21 (06) :531-547
[4]  
Hotraphinyo LF, 2001, P ANN INT IEEE EMBS, V23, P3454, DOI 10.1109/IEMBS.2001.1019573
[5]  
Lewis F.L., 1993, Control of robot manipulators
[6]  
MA SD, 1998, COMPUTER VISION THEO
[7]  
MAURER CR, 2002, ADV TECHNIQUES IMAGE, P10
[8]  
Schmerber S, 2001, Comput Aided Surg, V6, P1
[9]   Least-square fitting with spheres [J].
Spath, H .
JOURNAL OF OPTIMIZATION THEORY AND APPLICATIONS, 1998, 96 (01) :191-199
[10]  
Stoll J, 2005, LECT NOTES COMPUT SC, V3750, P41, DOI 10.1007/11566489_6