IMMERSION B-GUIDED VERSUS CONTACT A-MODE BIOMETRY FOR ACCURATE MEASUREMENT OF AXIAL LENGTH AND INTRAOCULAR LENS POWER CALCULATION IN SILICONIZED EYES

被引:14
作者
Abu El Einen, Khaled G. [1 ]
Shalaby, Mervat H. [2 ]
El Shiwy, Hoda T. [1 ]
机构
[1] Cairo Univ, Fac Med, Dept Ophthalmol, Cairo 12411, Egypt
[2] Cairo Univ, Fac Med, Ultrasonog Diagnost Ctr, Cairo 12411, Egypt
来源
RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES | 2011年 / 31卷 / 02期
关键词
A-mode biometry; axial length; B-guided biometry; IOL power calculation; recurrent retinal detachment; silicone cataract; silicone-retinal interface; OIL;
D O I
10.1097/IAE.0b013e3181e17f39
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To compare the accuracy of the immersion B-guided versus the contact A-mode biometry in measuring the axial length and intraocular lens power of the siliconized eyes. Methods: Sixty eyes with silicone cataract were divided into two groups. In Group A, the axial length of 30 eyes was measured by contact A-mode biometry. In Group B, the axial length of 30 eyes was measured by immersion B-guided biometry. The postoperative refraction was done after 3 months and was compared with the attempted values. Results: Twenty-five eyes of the B-guided group (83.3%) achieved a refraction deviation of +/- 1 D or less versus 14 (46.6%) in the A-mode group. The mean refractive deviation was less with the B-mode (0.15 +/- 0.94) versus contact A-mode (-0.95 +/- 1.157); the difference is statistically significant (P = 0.024, Fisher test). Conclusion: B-guided biometry was found to be more accurate than the contact method in measuring axial length and intraocular lens power calculation in siliconized eyes and spared the time and cost needed for a separate B-scan. RETINA 31: 262-265, 2011
引用
收藏
页码:262 / 265
页数:4
相关论文
共 11 条
  • [1] Combined phacoemulsification and transpupillary drainage of silicone oil: results and complications
    Assi, A
    Woodruff, S
    Gotzaridis, E
    Bunce, C
    Sullivan, P
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (08) : 942 - 945
  • [2] B-mode-guided vector-A-mode versus A-mode biometry to determine axial length and intraocular lens power
    Berges, O
    Puech, M
    Assouline, M
    Letenneur, L
    Gastellu-Etchegorry, M
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1998, 24 (04) : 529 - 535
  • [3] BUTCHER J M, 1991, Eye (London), V5, P708
  • [4] BYRNE SF, 1995, A SCAN AXIAL EYE LEN, P47
  • [5] BYRNE SF, 2002, AXIAL EYE LENGTH MEA
  • [6] Signal quality of biometry in silicone oil-filled eyes using partial coherence laser interferometry
    Dietlein, TS
    Roessler, G
    Lüke, C
    Dinslage, S
    Roters, S
    Jacobi, PC
    Walter, P
    Krieglstein, GK
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (05) : 1006 - 1010
  • [7] Refractive outcome of silicone oil removal and intraocular lens implantation using her interferometry
    Habibabadi, HF
    Hashemi, H
    Jalali, KH
    Amini, A
    Esfahani, MR
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2005, 25 (02): : 162 - 166
  • [8] Biometry and refractive outcome of eyes filled with silicone oil by standardized echography and partial coherence interferometry
    Nepp, J
    Krepler, K
    Jandrasits, K
    Hauff, W
    Hanselmayer, G
    Velikay-Parel, M
    Ossoinig, KC
    Wedrich, A
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2005, 243 (10) : 967 - 972
  • [9] SOURCES OF ERROR IN INTRAOCULAR-LENS POWER CALCULATION
    OLSEN, T
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1992, 18 (02) : 125 - 129
  • [10] PACKER M, 2002, J CATARACT REFR SURG, V28, P23942