Split Ring Resonator-Based Microwave Fluidic Sensors for Electrolyte Concentration Measurements

被引:183
作者
Velez, Paris [1 ]
Munoz-Enano, Jonathan [1 ]
Grenier, Katia [2 ]
Mata-Contreras, Javier [3 ]
Dubuc, David [2 ]
Martin, Ferran [1 ]
机构
[1] Univ Autonoma Barcelona, Dept Engn Electron, CIMITEC, Bellaterra 08193, Spain
[2] Natl Sci Res Ctr, Lab Anal & Architecture Syst, MH2F, F-31031 Toulouse, France
[3] Univ Malaga, Dept Ingn Comunicac, Malaga 29016, Spain
关键词
Microwave sensors; differential sensors; microfluidics; permittivity measurements; microstrip technology; split ring resonator (SRR); electrolyte concentration; ANGULAR DISPLACEMENT; ALIGNMENT; PAIRS; OCSRRS;
D O I
10.1109/JSEN.2018.2890089
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A differential microwave sensor, based on a pair of uncoupled microstrip lines each one loaded with a split ring resonator (SRR), is applied to the measurement of electrolyte concentration in deionized (DI) water. For that purpose, fluidic channels are added on top of the SRR gaps, the most sensitive parts of the structure. The operating principle is based on the measurement of the cross-mode insertion loss, highly sensitive to small asymmetries caused by differences between the reference liquid and the liquid under test (LUT). In this paper, the reference liquid is pure DI water (the solvent), whereas the solution, DI water with electrolyte content, is injected to the LUT channel. The proposed sensor is able to detect electrolyte concentrations as small as 0.25 g/L, with maximum sensitivity of 0.033 (g/L)(-1). The sensor is validated by measuring the concentration of three types of electrolytes, i.e., NaCl, KCl, and CaCl2. Finally, the sensor is applied to monitor variations of total electrolyte concentration in urine samples.
引用
收藏
页码:2562 / 2569
页数:8
相关论文
共 43 条
[1]  
[Anonymous], PEARSON REV RATIONAL
[2]   Real-Time Noninvasive Measurement of Glucose Concentration Using a Microwave Biosensor [J].
Bababjanyan, Arsen ;
Melikyan, Harutyun ;
Kim, Seungwan ;
Kim, Jongchel ;
Lee, Kiejin ;
Friedman, Barry .
JOURNAL OF SENSORS, 2010, 2010
[3]   Sodium chloride sensing by using a near-field microwave microprobe [J].
Babajanyan, Arsen ;
Kim, Jongchul ;
Kim, Songhui ;
Lee, Kiejin ;
Friedman, Barry .
APPLIED PHYSICS LETTERS, 2006, 89 (18)
[4]   Material Characterization Using Complementary Split-Ring Resonators [J].
Boybay, Muhammed Said ;
Ramahi, Omar M. .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2012, 61 (11) :3039-3046
[5]  
BUCK RP, 1976, ANAL CHEM, V48, pR23
[6]  
Daneshm, 2017, IEEE MTT S INT MICR, P1
[7]   High-Sensitivity Metamaterial-Inspired Sensor for Microfluidic Dielectric Characterization [J].
Ebrahimi, Amir ;
Withayachumnankul, Withawat ;
Al-Sarawi, Said ;
Abbott, Derek .
IEEE SENSORS JOURNAL, 2014, 14 (05) :1345-1351
[8]   Two-dimensional displacement and alignment sensor based on reflection coefficients of open microstrip lines loaded with split ring resonators [J].
Horestani, A. K. ;
Naqui, J. ;
Abbott, D. ;
Fumeaux, C. ;
Martin, F. .
ELECTRONICS LETTERS, 2014, 50 (08) :620-622
[9]   Two-dimensional alignment and displacement sensor based on movable broadside-coupled split ring resonators [J].
Horestani, Ali K. ;
Naqui, Jordi ;
Shaterian, Zahra ;
Abbott, Derek ;
Fumeaux, Christophe ;
Martin, Ferran .
SENSORS AND ACTUATORS A-PHYSICAL, 2014, 210 :18-24
[10]   Rotation Sensor Based on Horn-Shaped Split Ring Resonator [J].
Horestani, Ali Karami ;
Abbott, Derek ;
Fumeaux, Christophe .
IEEE SENSORS JOURNAL, 2013, 13 (08) :3014-3015