Renal volume assessment with 3D ultrasound
被引:0
作者:
Brancaforte, A.
[1
]
Serantoni, S.
[1
]
Barbosa, F. Silva
[1
]
Di Leo, G.
[2
]
Sardanelli, F.
[2
,3
]
Cornalba, G. P.
[1
]
机构:
[1] Univ Milan, Dipartimento Radiol Diagnost & Interventist, Osped San Paolo, I-20142 Milan, Italy
[2] IRCCS Policlin San Donato, Unita Radiol, Milan, Italy
[3] Univ Milan, Dipartimento Sci Med Chirurg, I-20142 Milan, Italy
来源:
RADIOLOGIA MEDICA
|
2011年
/
116卷
/
07期
关键词:
Ultrasound;
3DMSCT;
Renal volume;
665 ADULT VOLUNTEERS;
COMPUTED-TOMOGRAPHY;
KIDNEY VOLUME;
ACCURACY;
SIZE;
MRI;
AGE;
CT;
D O I:
10.1007/s11547-011-0691-8
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Purpose. This study was undertaken to determine the accuracy of 3D ultrasound (US) in assessing renal volume, with multislice computed tomography (MSCT) considered as the gold standard. Materials and methods. Forty-nine patients (30 men, 19 women; age range 30-82 years) underwent abdominal contrast-enhanced MSCT and 3D-US performed with a 3.5-MHz 3D/4D convex-array probe. The results of the two modalities were compared with the Wilcoxon test. Variability between the two measurements was determined with the Bland-Altman method and reported in terms of bias and coefficient of repeatability (CoR). Results. Mean values obtained were 210 ml with MSCT and 192 ml with 3D-US (p < 0.001). Analysis of variability per patient between MSCT and 3D-US showed a bias of 19 ml, a CoR of 47 ml and an accuracy of 78%, with an average 3D-US underestimation of 19 ml (9%). Analysis of variability per kidney showed a bias of 9 ml, a CoR of 34 ml and an accuracy of 80%. Conclusions. Three-dimensional US is a valuable technique for monitoring renal volume, whereas MSCT may be reserved for assessing renal anatomy and relationships with neighbouring organs.
引用
收藏
页码:1095 / 1104
页数:10
相关论文