Postoperative Electrocochleography from Hybrid Cochlear Implant users: An Alternative Analysis Procedure

被引:18
作者
Kim, Jeong-Seo [1 ,2 ]
Tejani, Viral D. [1 ,2 ]
Abbas, Paul J. [1 ,2 ]
Brown, Carolyn J. [1 ,2 ]
机构
[1] Univ Iowa, Dept Commun Sci & Disorders, 311 SHC,250 Hawkins Dr, Iowa City, IA 52242 USA
[2] Univ Iowa Hosp & Clin, Dept Otolaryngol Head & Neck Surg, Iowa City, IA 52242 USA
基金
美国国家卫生研究院;
关键词
Cochlear implant; Hybrid; Electrocochleography; Cochlear microphonic; Auditory nerve neurophonic; Neural response telemetry; Hearing preservation; ROUND WINDOW ELECTROCOCHLEOGRAPHY; MULTICENTER CLINICAL-TRIAL; AUDITORY-NERVE FIBERS; HEARING PRESERVATION; SPEECH-PERCEPTION; RESIDUAL HEARING; HAIR CELL; ELECTRODE; HISTOPATHOLOGY; RECOGNITION;
D O I
10.1016/j.heares.2018.10.016
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objective: Shorter electrode arrays and soft surgical techniques allow for preservation of acoustic hearing in many cochlear implant (CI) users. Recently, we developed a method of using the Neural Response Telemetry (NRT) system built in Custom Sound EP clinical software to record acoustically evoked electrocochleography (ECoG) responses from an intracochlear electrode in Nucleus Hybrid CI users (Abbas et al., 2017). We recorded responses dominated by the hair cells (cochlear microphonic, CM/DIF) and the auditory nerve (auditory nerve neurophonic, ANN/SUM). Unfortunately, the recording procedure was time consuming, limiting potential clinical applications. This report describes a modified method to record the ECoG response more efficiently. We refer to this modified technique as the "short window" method, while our previous technique (Abbas et al., 2017) is referred as the "long window" method. In this report, our goal was to 1) evaluate the feasibility of the short window method to record the CM/DIF and ANN/SUM responses, 2) characterize the reliability and sensitivity of the measures recorded using the short window method, and 3) evaluate the relationship between the CM/DIF and ANN/SUM measures recorded using the modified method and audiometric thresholds. Method: Thirty-four postlingually deafened adult Hybrid CI users participated in this study. Acoustic tone bursts were presented at four frequencies (250, 500, 750, and 1000 Hz) at various stimulation levels via an insert earphone in both condensation and rarefaction polarities. Acoustically evoked ECoG responses were recorded from the most apical electrode in the intracochlear array. These two responses were subtracted to emphasize the CM/DIF responses and added to emphasize the ANN/SUM responses. Response thresholds were determined based on visual inspection of time waveforms, and trough-to peak analysis technique was used to quantify response amplitudes. Within-subject comparison of responses measured using both short and long window methods were obtained from seven subjects. We also assessed the reliability and sensitivity of the short window method by comparing repeated measures from 19 subjects at different times. Correlations between CM/DIF and ANN/SUM measures using the short window recording method and audiometric thresholds were also assessed. Results: Regardless of the recording method, CM/DIF responses were larger than ANN/SUM responses. Responses obtained using the short window method were positively correlated to those obtained using the conventional long window method. Subjects who had stable acoustic hearing at two different time points had similar ECoG responses at those points, confirming high test-retest reliability of the short window method. Subjects who lost hearing between two different time points showed increases in ECoG thresholds, suggesting that physiologic ECoG responses are sensitive to audiometric changes. Correlations between CM/DIF and ANN/SUM thresholds and audiometric thresholds at all tested frequencies were significant. Conclusion: This study compares two different recording methods. Intracochlear ECoG measures recorded using the short window technique were efficient, reliable, and repeatable. We were able to collect more frequency specific data with the short window method, and observed similar results between the long window and short window methods. Correlations between physiological thresholds and audiometric thresholds were similar to those reported previously using the long window method (Abbas et al., 2017). This is an important finding because it demonstrates that clinically-available software can be used to measure frequency-specific ECoG responses with enhanced efficiency, increasing the odds that this technique might move from the laboratory into clinical practice. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:304 / 315
页数:12
相关论文
共 58 条
[1]   Using Neural Response Telemetry to Monitor Physiological Responses to Acoustic Stimulation in Hybrid Cochlear Implant Users [J].
Abbas, Paul J. ;
Tejani, Viral D. ;
Scheperle, Rachel A. ;
Brown, Carolyn J. .
EAR AND HEARING, 2017, 38 (04) :409-425
[2]   Round window electrocochleography before and after cochlear implant electrode insertion [J].
Adunka, Oliver F. ;
Giardina, Christopher K. ;
Formeister, Eric J. ;
Choudhury, Baishakhi ;
Buchman, Craig A. ;
Fitzpatrick, Douglas C. .
LARYNGOSCOPE, 2016, 126 (05) :1193-1200
[3]  
[Anonymous], 2011, National Health and Nutrition Examination Survey
[4]  
[Anonymous], 1965, Discharge Patterns of Single Fibers in the Cat's Auditory Nerve
[5]   Intraoperative Real-time Cochlear Response Telemetry Predicts Hearing Preservation in Cochlear Implantation [J].
Campbell, Luke ;
Kaicer, Arielle ;
Sly, David ;
Iseli, Claire ;
Wei, Benjamin ;
Briggs, Robert ;
O'Leary, Stephen .
OTOLOGY & NEUROTOLOGY, 2016, 37 (04) :332-338
[6]   Cochlear Response Telemetry: Intracochlear Electrocochleography via Cochlear Implant Neural Response Telemetry Pilot Study Results [J].
Campbell, Luke ;
Kaicer, Arielle ;
Briggs, Robert ;
O'Leary, Stephen .
OTOLOGY & NEUROTOLOGY, 2015, 36 (03) :399-405
[7]   Intraoperative Round Window Recordings to Acoustic Stimuli From Cochlear Implant Patients [J].
Choudhury, Baishakhi ;
Fitzpatrick, Douglas C. ;
Buchman, Craig A. ;
Wei, Benjamin P. ;
Dillon, Margaret T. ;
He, Shuman ;
Adunka, Oliver F. .
OTOLOGY & NEUROTOLOGY, 2012, 33 (09) :1507-1515
[8]   Assessment of Cochlear Function during Cochlear Implantation by Extra- and Intracochlear Electrocochleography [J].
Dalbert, Adrian ;
Pfiffner, Flurin ;
Hoesli, Marco ;
Koka, Kanthaiah ;
Veraguth, Dorothe ;
Roosli, Christof ;
Huber, Alexander .
FRONTIERS IN NEUROSCIENCE, 2018, 12
[9]   Contralateral suppression of otoacoustic emissions - Delay of effect? [J].
Durrant, JD .
JOURNAL OF COMMUNICATION DISORDERS, 1998, 31 (06) :485-488
[10]   THRESHOLD CHARACTERISTICS OF THE HUMAN AUDITORY BRAIN-STEM RESPONSE [J].
ELBERLING, C ;
DON, M .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1987, 81 (01) :115-121