Column radial homogeneity in high-performance liquid chromatography

被引:74
|
作者
Farkas, T
Sepaniak, MJ
Guiochon, G
机构
[1] UNIV TENNESSEE,DEPT CHEM,KNOXVILLE,TN 37996
[2] OAK RIDGE NATL LAB,DIV CHEM & ANALYT SCI,OAK RIDGE,TN 37831
关键词
flow velocity distribution; fluorescence detection; on-column; column efficiency;
D O I
10.1016/0021-9673(96)00107-0
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The radial distribution of analyte molecules within an elution band in HPLC was determined by local, on-column, fluorescence detection at the column outlet. Several optical fiber assemblies were implanted in the exit frit at different points over the column cross-section and the fluorescence of a laser-dye analyte was measured. The individual elements of a diode array were used as independent detectors. The distribution of the mobile phase velocity across the column was measured for a number of standard size analytical HPLC columns of different efficiencies, operated at different mobile phase linear velocities. The dependence of the column efficiency on these profiles is discussed.