Comparison of conventional and cadmium-zinc-telluride single-photon emission computed tomography for analysis of thallium-201 myocardial perfusion imaging: an exploratory study in normal databases for different ethnicities

被引:5
作者
Ishihara, Masaru [1 ,2 ]
Onoguchi, Masahisa [2 ]
Taniguchi, Yasuyo [3 ]
Shibutani, Takayuki [2 ]
机构
[1] Hyogo Canc Ctr, Dept Radiol, 13-70 Kitaoji, Akashi, Hyogo 6738558, Japan
[2] Kanazawa Univ, Grad Sch Med Sci, Dept Quantum Med Technol, 5-11-80 Kodatsuno, Kanazawa, Ishikawa 9200942, Japan
[3] Hyogo Brain & Heart Ctr, Dept Cardiol, 520 Saisho Ko, Himeji, Hyogo 6700981, Japan
关键词
Solid-state dedicated cardiac gamma camera; Conventional anger-type gamma camera; Single-photon emission computed tomography; Myocardial perfusion imaging; Thallium-201; LEFT-VENTRICULAR FUNCTION; GATED SPECT; HIGH-SPEED; GAMMA-CAMERA; SYSTEM; TECHNETIUM-99M-SESTAMIBI; MULTICENTER; PERFORMANCE; SESTAMIBI; DEFECTS;
D O I
10.1007/s10554-017-1205-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this study was to clarify the differences in thallium-201-chloride (thallium-201) myocardial perfusion imaging (MPI) scans evaluated by conventional anger-type single-photon emission computed tomography (conventional SPECT) versus cadmium-zinc-telluride SPECT (CZT SPECT) imaging in normal databases for different ethnic groups. MPI scans from 81 consecutive Japanese patients were examined using conventional SPECT and CZT SPECT and analyzed with the pre-installed quantitative perfusion SPECT (QPS) software. We compared the summed stress score (SSS), summed rest score (SRS), and summed difference score (SDS) for the two SPECT devices. For a normal MPI reference, we usually use Japanese databases for MPI created by the Japanese Society of Nuclear Medicine, which can be used with conventional SPECT but not with CZT SPECT. In this study, we used new Japanese normal databases constructed in our institution to compare conventional and CZT SPECT. Compared with conventional SPECT, CZT SPECT showed lower SSS (p < 0.001), SRS (p = 0.001), and SDS (p = 0.189) using the pre-installed SPECT database. In contrast, CZT SPECT showed no significant difference from conventional SPECT in QPS analysis using the normal databases from our institution. Myocardial perfusion analyses by CZT SPECT should be evaluated using normal databases based on the ethnic group being evaluated.
引用
收藏
页码:2057 / 2066
页数:10
相关论文
共 22 条
  • [1] Radiopharmaceutical agents for myocardial perfusion imaging
    Baggish, Aaron L.
    Boucher, Charles A.
    [J]. CIRCULATION, 2008, 118 (16) : 1668 - 1674
  • [2] Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera
    Ben-Haim, Simona
    Kacperski, Krzysztof
    Hain, Sharon
    Van Gramberg, Dean
    Hutton, Brian F.
    Erlandsson, Kjell
    Sharir, Tali
    Roth, Nathaniel
    Waddington, Wendy A.
    Berman, Daniel S.
    Ell, Peter J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 (09) : 1710 - 1721
  • [3] Stress Thallium-201/Rest Technetium-99m Sequential Dual Isotope High-Speed Myocardial Perfusion Imaging
    Berman, Daniel S.
    Kang, Xingping
    Tamarappoo, Balaji
    Wolak, Arik
    Hayes, Sean W.
    Nakazato, Ryo
    Thomson, Louise E. J.
    Kite, Faith
    Cohen, Ishac
    Slomka, Piotr J.
    Einstein, Andrew J.
    Friedman, John D.
    [J]. JACC-CARDIOVASCULAR IMAGING, 2009, 2 (03) : 273 - 282
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Left-ventricular dyssynchrony evaluated by Tl-201 gated SPECT myocardial perfusion imaging: a comparison with Tc-99m sestamibi
    Chen, Chien-Cheng
    Huang, Wen-Sheng
    Hung, Guang-Uei
    Chen, Wan-Chen
    Kao, Chia-Hung
    Chen, Ji
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2013, 34 (03) : 229 - 232
  • [6] Normal Myocardial Perfusion SPECT Database for the Spanish Population
    Cuberas-Borros, Gemma
    Aguade-Bruix, Santiago
    Boronat-de Ferrater, Maria
    Africa Muxi-Pradas, Maria
    Romero-Farina, Guillermo
    Castell-Conesa, Joan
    Aliaga, Veronica
    Garcia-Dorado, David
    Candell-Riera, Jaume
    [J]. REVISTA ESPANOLA DE CARDIOLOGIA, 2010, 63 (08): : 934 - 942
  • [7] DEPUEY E, 1995, J NUCL MED, V36, P952
  • [8] Comparison of Image Quality, Myocardial Perfusion, and Left Ventricular Function Between Standard Imaging and Single-Injection Ultra-Low-Dose Imaging Using a High-Efficiency SPECT Camera: The MILLISIEVERT Study
    Einstein, Andrew J.
    Blankstein, Ron
    Andrews, Howard
    Fish, Mathews
    Padgett, Richard
    Hayes, Sean W.
    Friedman, John D.
    Qureshi, Mehreen
    Rakotoarivelo, Harivony
    Slomka, Piotr
    Nakazato, Ryo
    Bokhari, Sabahat
    Di Carli, Marcello
    Berman, Daniel S.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (09) : 1430 - 1437
  • [9] Performance evaluation of D-SPECT: a novel SPECT system for nuclear cardiology
    Erlandsson, Kjell
    Kacperski, Krzysztof
    van Gramberg, Dean
    Hutton, Brian F.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (09) : 2635 - 2649
  • [10] Myocardial ischemic reduction evidenced by gated myocardial perfusion imaging after treatment results in good prognosis in patients with coronary artery disease
    Hon, Yusuke
    Yoda, Shunichi
    Nakanishi, Kanae
    Tano, Ayako
    Suzuki, Yasuyuki
    Matsumoto, Naoya
    Hirayama, Atsushi
    [J]. Journal of Cardiology, 2015, 65 (3-4) : 278 - 284