Fast interleaved echo-planar imaging with navigator: High resolution anatomic and functional images at 4 Tesla

被引:75
作者
Kim, SG
Hu, XP
Adriany, G
Ugurbil, K
机构
[1] Ctr. for Magnetic Resonance Research, Department of Radiology, Univ. of Minnesota Medical School, Minneapolis, MN
[2] CMRR, Univ. of Minnesota Medical School, Minneapolis, MN 55455
关键词
echo-planar imaging; navigator; high resolution imaging; functional MRI;
D O I
10.1002/mrm.1910350618
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Echo-planar imaging (EPI) is sensitive to magnetic field inhomogeneities, which lead to signal loss and geometric distortions of the image, Magnetic field inhomogeneities induced by susceptibility differences, as encountered in the human body, increase with the magnetic field strength, thus, complicating implementation of high resolution EPI techniques on high magnetic field systems. These problems were overcome by using a fast multishot high resolution EPI method that uses variable flip angles, center-out k-space sampling, and navigator echoes, This approach maximizes signal-to-noise ratio, reduces flow artifacts, and permits correction of intersegment amplitude and phase variations, providing high spatial and temporal resolution, This scheme can be implemented with a single magnetization preparation for contrast that precedes the segments, The utility of this ultrafast segmented EPI technique with navigator is demonstrated for anatomic and functional imaging studies on the human brain at 4 T.
引用
收藏
页码:895 / 902
页数:8
相关论文
共 30 条
  • [1] TIME COURSE EPI OF HUMAN BRAIN-FUNCTION DURING TASK ACTIVATION
    BANDETTINI, PA
    WONG, EC
    HINKS, RS
    TIKOFSKY, RS
    HYDE, JS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 25 (02) : 390 - 397
  • [2] PROCESSING STRATEGIES FOR TIME-COURSE DATA SETS IN FUNCTIONAL MRI OF THE HUMAN BRAIN
    BANDETTINI, PA
    JESMANOWICZ, A
    WONG, EC
    HYDE, JS
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 30 (02) : 161 - 173
  • [3] DYNAMIC MAPPING OF THE HUMAN VISUAL-CORTEX BY HIGH-SPEED MAGNETIC-RESONANCE-IMAGING
    BLAMIRE, AM
    OGAWA, S
    UGURBIL, K
    ROTHMAN, D
    MCCARTHY, G
    ELLERMANN, JM
    HYDER, F
    RATTNER, Z
    SHULMAN, RG
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1992, 89 (22) : 11069 - 11073
  • [4] BOWTELL R, 1994, P 2 ANN M ISMRM SAN, P411
  • [5] IMAGE-RECONSTRUCTION FOR ECHO PLANAR IMAGING WITH NONEQUIDISTANT KAPPA-SPACE SAMPLING
    BRUDER, H
    FISCHER, H
    REINFELDER, HE
    SCHMITT, F
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (02) : 311 - 323
  • [6] INTERLEAVED ECHO-PLANAR IMAGING ON A STANDARD MRI SYSTEM
    BUTTS, K
    RIEDERER, SJ
    EHMAN, RL
    THOMPSON, RM
    JACK, CR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (01) : 67 - 72
  • [7] ANALYSIS OF FLOW EFFECTS IN ECHO-PLANAR IMAGING
    BUTTS, K
    RIEDERER, SJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03): : 285 - 293
  • [8] Cho ZH, 1987, 6TH P SMRM ANN M, P912
  • [9] THEORETICAL ASPECTS OF MOTION SENSITIVITY AND COMPENSATION IN ECHO-PLANAR IMAGING
    DUERK, JL
    SIMONETTI, OP
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06): : 643 - 650
  • [10] ADAPTIVE TECHNIQUE FOR HIGH-DEFINITION MR IMAGING OF MOVING STRUCTURES
    EHMAN, RL
    FELMLEE, JP
    [J]. RADIOLOGY, 1989, 173 (01) : 255 - 263