Non-contact sensing of neonatal pulse rate using camera-based imaging: a clinical feasibility study

被引:99
|
作者
Paul, M. [1 ]
Karthik, S. [2 ]
Joseph, J. [2 ]
Sivaprakasam, M. [2 ]
Kumutha, J. [3 ]
Leonhardt, S. [1 ]
Antink, C. Hoog [1 ]
机构
[1] Rhein Westfal TH Aachen, Med Informat Technol MedIT, Helmholtz Inst Biomed Engn, Pauwelsstr 20, D-52074 Aachen, Germany
[2] Healthcare Technol Innovat Ctr HTIC, IIT Madras Res Pk, Chennai 600113, Tamil Nadu, India
[3] Saveetha Med Coll, Chennai 602105, Tamil Nadu, India
关键词
photoplethysmography; photoplethysmography imaging; pulse rate; heart rate; camera-based; remote; non-contact; neonate; infant;
D O I
10.1088/1361-6579/ab755c
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: Neonates and infants are patients who would benefit from less invasive vital sign sensing, especially from fewer cables and the avoidance of adhesive electrodes. Photoplethysmography imaging (PPGI) has been studied for medical applications in recent years: it is possible to assess various vital signs remotely, non-invasively, and without contact by using video cameras and light. However, studies on infants and especially on neonates in clinical settings are still rare. Hence, we conducted a single-center study to assess heart activity by estimating the pulse rate (PR) of 19 neonates. Approach: Time series were generated from tracked regions of interest (ROIs) and PR was estimated via a joint time-frequency analysis using a short-time Fourier transform. Artifacts, for example, induced by movement, were detected and flagged by applying a signal quality index in the frequency domain. Main results: The feasibility of PR estimation was demonstrated using visible light and near-infrared light at 850 nm and 940 nm, respectively: the estimated PR was as close as 3 heartbeats per minute in artifact-free time segments. Furthermore, an improvement could be shown when selecting the best performing ROI compared to the ROI containing the whole body. The main challenges are artifacts from motion, light sources, medical devices, and the detection and tracking of suitable regions for signal retrieval. Nonetheless, the PR extracted was found to be comparable to the contact-based photoplethysmography reference and is, therefore, a viable replacement if robust signal retrieval is ensured. Significance: Neonates are seldom measured by PPGI and studies reporting measurements on darker skin tones are rare. In this work, not only a single camera was used, but a synchronized camera setup using multiple wavelengths. Various ROIs were used for signal extraction to examine the capabilities of PPGI. In addition, qualitative observations regarding camera parameters and noise sources were reported and discussed.
引用
收藏
页数:22
相关论文
共 39 条
  • [21] Blood Volume Pulse Extraction for Non-Contact Heart Rate Measurement by Digital Camera Using Singular Value Decomposition and Burg Algorithm
    Tayibnapis, Iman Rahmansyah
    Yang, Yeon-Mo
    Lim, Ki Moo
    ENERGIES, 2018, 11 (05)
  • [22] Non-contact automatic vital signs monitoring of neonates in NICU using video camera imaging
    Khanam, Fatema-Tuz-Zohra
    Al-Naji, Ali
    Perera, Asanka G.
    Gibson, Kim
    Chahl, Javaan
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING-IMAGING AND VISUALIZATION, 2023, 11 (02) : 278 - 285
  • [23] Nanosecond Pulse Near-field Sensing based Non-contact Physiological Signals Measurement
    Lin, Hong-Dun
    Lee, Yen-Shien
    Su, Yu-Jen
    Chuang, Bor-Nian
    ADVANCED MEASUREMENT AND TEST, PTS 1-3, 2011, 301-303 : 1214 - 1219
  • [24] Camera-Based, Non-Contact, Vital-Signs Monitoring Technology May Provide a Way for the Early Prevention of SIDS in Infants
    Zhao, Fang
    Li, Meng
    Jiang, Zhongyi
    Tsien, Joe Z.
    Lu, Zhaohui
    FRONTIERS IN NEUROLOGY, 2016, 7
  • [25] Recovering Pulse Rate During Motion Artifact with a Multi-Imager Array for Non-Contact Imaging Photoplethysmography
    Estepp, Justin R.
    Blackford, Ethan B.
    Meier, Christopher M.
    2014 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN AND CYBERNETICS (SMC), 2014, : 1462 - 1469
  • [26] The Evolution of Heart Beat Rate Measurement Techniques From Contact Based Photoplethysmography to Non-Contact Based Photoplethysmography Imaging
    Khong, Wei Leong
    Mariappan, Muralindran
    2019 4TH IEEE INTERNATIONAL CIRCUITS AND SYSTEMS SYMPOSIUM (ICSYS), 2019,
  • [27] Non-Contact Respiration Rate Estimation using 24 GHz Pulse Radar Employing Envelope Detection
    Zainuddin, Suraya
    Abd Rashid, Nur Emileen
    Shariff, Khairul Khaizi Mohd
    Ali, Megat Syahirul Amin Megat
    Seman, Fauziahanim Che
    Pasya, Idnin
    2020 IEEE INTERNATIONAL RF AND MICROWAVE CONFERENCE (RFM), 2020,
  • [28] Non-Contact Remote Measurement of Heart Rate Variability using Near-Infrared Photoplethysmography Imaging
    Yu, Xinchi
    Paul, Michael
    Antink, Christoph Hoog
    Venema, Boudewijn
    Blazek, Vladimir
    Bollheimer, Cornelius
    Leonhardt, Steffen
    Teichmann, Daniel
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 846 - 849
  • [29] Non-contact heart rate measurement using low-cost RGB camera under complex light conditions
    Wang H.
    Zhang S.
    Multimedia Tools and Applications, 2025, 84 (3) : 1561 - 1575
  • [30] ROI area that can increase SNR of non-contact heart rate measurement based on face image using 2D camera
    Lee, Seongryeong
    Kim, Sungho
    2022 22ND INTERNATIONAL CONFERENCE ON CONTROL, AUTOMATION AND SYSTEMS (ICCAS 2022), 2022, : 1283 - 1287