Long-term competing risks for overall and cause-specific failure of rotating-hinge distal femoral arthroplasty for tumour reconstruction

被引:15
作者
Ogura, K. [1 ]
Fujiwara, T. [1 ]
Morris, C. D. [2 ]
Boland, P. J. [1 ]
Healey, J. H. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Orthopaed Serv, Dept Surg, 1275 York Ave, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, 1275 York Ave, New York, NY 10021 USA
关键词
TOTAL KNEE REPLACEMENT; ENDOPROSTHETIC REPLACEMENT; SURVIVAL; RESECTION; FEMUR; PROSTHESES; WALLDIUS;
D O I
10.1302/0301-620X.103B8.BJJ-2020-2323.R1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Aims Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. Methods We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotatinghinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the boneimplant fixation); or amputation. Results Multivariate Fine-Gray regression analyses revealed different hazards for each Henderson failure mode: percentage of femoral resection (p = 0.001) and extent of quadriceps muscle resection (p = 0.005) for overall prosthetic failure; extent of quadriceps muscle resection (p = 0.002) and fixation of femoral component (p = 0.011) for type 2 failure (aseptic loosening); age (p = 0.009) and percentage of femoral resection (p = 0.019) for type 3 failure (mechanical failure); and type of joint resection (p = 0.037) for type 4 (infection) were independent predictors. A bone stem ratio of > 2.5 reliably predicted aseptic loosening. Conclusion We identified independent risk factors for overall and cause-specific prosthetic failure after rotating-hinge knee distal femoral arthroplasty using a competing risk Fine-Gray model. A bone stem ratio > 2.5 reliably predicts aseptic loosening. An accurate knowledge of the risks of distal femoral arthroplasty after resection for tumour assists surgical planning and managing patient expectations.
引用
收藏
页码:1405 / 1413
页数:9
相关论文
共 30 条
[1]   Uncemented, custom-made, hydroxyapatite-coated collared distal femoral endoprostheses [J].
Batta, V. ;
Coathup, M. J. ;
Parratt, M. T. ;
Pollock, R. C. ;
Aston, W. J. ;
Cannon, S. R. ;
Skinner, J. A. ;
Briggs, T. W. ;
Blunn, G. W. .
BONE & JOINT JOURNAL, 2014, 96B (02) :263-269
[2]   Aseptic Loosening Rates in Distal Femoral Endoprostheses: Does Stem Size Matter? [J].
Bergin, Patrick F. ;
Noveau, Jenna B. ;
Jelinek, James S. ;
Henshaw, Robert M. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (03) :743-750
[3]   Survival of total knee replacement with a megaprosthesis after bone tumor resection [J].
Biau, D ;
Faure, F ;
Katsahian, S ;
Jeanrot, C ;
Tomeno, B ;
Anract, P .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2006, 88A (06) :1285-1293
[4]  
Bickels J, 2002, CLIN ORTHOP RELAT R, P225
[5]   What Are the Long-term Results of MUTARSA® Modular Endoprostheses for Reconstruction of Tumor Resection of the Distal Femur and Proximal Tibia? [J].
Bus, Michael P. A. ;
van de Sande, Michiel A. J. ;
Fiocco, Marta ;
Schaap, Gerard R. ;
Bramer, Jos A. M. ;
Dijkstra, P. D. Sander .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2017, 475 (03) :708-718
[6]   MODULAR UNCEMENTED PROSTHETIC RECONSTRUCTION AFTER RESECTION OF TUMORS OF THE DISTAL FEMUR [J].
CAPANNA, R ;
MORRIS, HG ;
CAMPANACCI, D ;
DELBEN, M ;
CAMPANACCI, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (02) :178-186
[7]   Long-Term Survival of Cemented Distal Femoral Endoprostheses with a Hydroxyapatite-Coated Collar A Histological Study and a Radiographic Follow-up [J].
Coathup, Melanie Jean ;
Batta, Vineet ;
Pollock, Robin C. ;
Aston, William J. ;
Cannon, Stephen R. ;
Skinner, John A. ;
Briggs, Timothy W. R. ;
Unwin, Paul S. ;
Blunn, Gordon William .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2013, 95A (17) :1569-1575
[8]  
Donato KA, 1998, ARCH INTERN MED, V158, P1855, DOI 10.1001/archinte.158.17.1855
[9]   Chemotherapy Curtails Bone Formation From Compliant Compression Fixation of Distal Femoral Endoprostheses [J].
Elalfy, Mohammad A. ;
Boland, Patrick J. ;
Healey, John H. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2019, 477 (01) :206-216
[10]   Early Equivalence of Uncemented Press-fit and CompressA® Femoral Fixation [J].
Farfalli, German L. ;
Boland, Patrick J. ;
Morris, Carol D. ;
Athanasian, Edward A. ;
Healey, John H. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (11) :2792-2799