Effects of refractive index mismatch in optical CT imaging of polymer gel dosimeters

被引:6
作者
Manjappa, Rakesh [1 ]
Makki, Sharath S. [1 ]
Kumar, Rajesh [2 ]
Kanhirodan, Rajan [1 ]
机构
[1] Indian Inst Sci, Dept Phys, Bangalore 560012, Karnataka, India
[2] BARC, Radiol Phys & Advisory Div, Mumbai 400094, Maharashtra, India
关键词
polymer gel dosimetry; optical tomography; image reconstruction; refraction correction; total variation denoising; COMPUTED-TOMOGRAPHY; APPROXIMATION; SCANNERS;
D O I
10.1118/1.4905043
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Proposing an image reconstruction technique, algebraic reconstruction technique-refraction correction (ART-rc). The proposed method takes care of refractive index mismatches present in gel dosimeter scanner at the boundary, and also corrects for the interior ray refraction. Polymer gel dosimeters with high dose regions have higher refractive index and optical density compared to the background medium, these changes in refractive index at high dose results in interior ray bending. Methods: The inclusion of the effects of refraction is an important step in reconstruction of optical density in gel dosimeters. The proposed ray tracing algorithm models the interior multiple refraction at the inhomogeneities. Jacob's ray tracing algorithm has been modified to calculate the pathlengths of the ray that traverses through the higher dose regions. The algorithm computes the length of the ray in each pixel along its path and is used as the weight matrix. Algebraic reconstruction technique and pixel based reconstruction algorithms are used for solving the reconstruction problem. The proposed method is tested with numerical phantoms for various noise levels. The experimental dosimetric results are also presented. Results: The results show that the proposed scheme ART-rc is able to reconstruct optical density inside the dosimeter better than the results obtained using filtered backprojection and conventional algebraic reconstruction approaches. The quantitative improvement using ART-rc is evaluated using gamma-index. The refraction errors due to regions of different refractive indices are discussed. The effects of modeling of interior refraction in the dose region are presented. Conclusions: The errors propagated due to multiple refraction effects have been modeled and the improvements in reconstruction using proposed model is presented. The refractive index of the dosimeter has a mismatch with the surrounding medium (for dry air or water scanning). The algorithm reconstructs the dose profiles by estimating refractive indices of multiple inhomogeneities having different refractive indices and optical densities embedded in the dosimeter. This is achieved by tracking the path of the ray that traverses through the dosimeter. Extensive simulation studies have been carried out and results are found to be matching that of experimental results. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:750 / 759
页数:10
相关论文
共 34 条
  • [1] [Anonymous], 2021, MATLAB version 7.10.0
  • [2] Polymer gel dosimetry
    Baldock, C.
    De Deene, Y.
    Doran, S.
    Ibbott, G.
    Jirasek, A.
    Lepage, M.
    McAuley, K. B.
    Oldham, M.
    Schreiner, L. J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (05) : R1 - R63
  • [3] Baldock C., 2009, J PHYS, V164, P1, DOI DOI 10.1088/1742-6596/164/1/012002
  • [4] Initial investigation of a novel light-scattering gel phantom for evaluation of optical CT scanners for radiotherapy gel dosimetry
    Bosi, Stephen
    Naseri, Pourandokht
    Puran, Alicia
    Davies, Justin
    Baldock, Clive
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (10) : 2893 - 2903
  • [5] Recent developments with a prototype fan-beam optical CT scanner
    Campbell, W. G.
    Jirasek, A.
    Wells, D.
    [J]. 7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [6] A prototype fan-beam optical CT scanner for 3D dosimetry
    Campbell, Warren G.
    Rudko, D. A.
    Braam, Nicolas A.
    Wells, Derek M.
    Jirasek, Andrew
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Use of Split Bregman denoising for iterative reconstruction in fluorescence diffuse optical tomography
    Chamorro-Servent, Judit
    Abascal, Juan F. P. J.
    Aguirre, Juan
    Arridge, Simon
    Correia, Teresa
    Ripoll, Jorge
    Desco, Manuel
    Vaquero, Juan J.
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2013, 18 (07)
  • [8] Best fit refractive index of matching liquid for 3D NIPAM gel dosimeters using optical CT
    Chen, Chin-Hsing
    Wu, Jay
    Hsieh, Bor-Tsung
    Chen, De-Shiou
    Wang, Tzu-Hwei
    Chien, Sou-Hsin
    Chang, Yuan-Jen
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 192 - 197
  • [9] The effects of internal refractive index variation in near-infrared optical tomography: a finite element modelling approach
    Dehghani, H
    Brooksby, B
    Vishwanath, K
    Pogue, BW
    Paulsen, KD
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (16) : 2713 - 2727
  • [10] Eliminating the need for refractive index matching in optical CT scanners for radiotherapy dosimetry: I. Concept and simulations
    Doran, Simon J.
    Yatigammana, Dylan N. B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (03) : 665 - 683