Experimental evaluation of validity of simplified Monte Carlo method in proton dose calculations

被引:34
作者
Kohno, R
Takada, Y
Sakae, T
Terunuma, T
Matsumoto, K
Nohtomi, A
Matsuda, H
机构
[1] Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 3058575, Japan
[2] Univ Tsukuba, Proton Med Res Ctr, Tsukuba, Ibaraki 3058575, Japan
[3] Univ Tsukuba, Inst Appl Phys, Tsukuba, Ibaraki 3058573, Japan
关键词
D O I
10.1088/0031-9155/48/10/303
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is important for proton therapy to calculate dose distributions accurately in treatment planning. Dose calculations in the body for treatment planning are converted to dose distributions in water, and the converted calculations are then generally evaluated by the dose measurements in water. In this paper, proton dose calculations were realized for a phantom simulating a clinical heterogeneity. Both dose calculations in the phantom calculated by two dose calculation methods, the range-modulated pencil beam algorithm (RMPBA) and the simplified Monte Carlo (SMC) method, and dose calculations converted to dose distributions in water by the same two methods were verified experimentally through comparison with measured distributions, respectively. For the RMPBA, though the converted calculations in water agreed moderately well with the measured ones, the calculated results in the actual phantom produced large errors. This meant that dose calculations in treatment planning should be evaluated by the dose measurements not in water but in the body with heterogeneity. On the other hand, the results calculated in the phantom, even by the less rigorous SMC method, reproduced the experimental ones well. This finding showed that actual dose distributions in the body should be predicted by the SMC method.
引用
收藏
页码:1277 / 1288
页数:12
相关论文
共 16 条
[1]   Monte Carlo and analytical calculation of proton pencil beams for computerized treatment plan optimization [J].
Carlsson, AK ;
Andreo, P ;
Brahme, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (06) :1033-1053
[2]  
*CERN APPL SOFTW G, 1994, GENAT VERS 3 21 DET
[3]   TREATMENT PLANNING FOR HEAVY-ION RADIOTHERAPY [J].
CHEN, GTY ;
SINGH, RP ;
CASTRO, JR ;
LYMAN, JT ;
QUIVEY, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (10) :1809-1819
[4]   MULTIPLE COULOMB SCATTERING OF 160 MEV PROTONS [J].
GOTTSCHALK, B ;
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM ;
WAGNER, MS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1993, 74 (04) :467-490
[5]   SOME PRACTICAL REMARKS ON MULTIPLE-SCATTERING [J].
HIGHLAND, VL .
NUCLEAR INSTRUMENTS & METHODS, 1975, 129 (02) :497-499
[6]   A pencil beam algorithm for proton dose calculations [J].
Hong, L ;
Goitein, M ;
Bucciolini, M ;
Comiskey, R ;
Gottschalk, B ;
Rosenthal, S ;
Serago, C ;
Urie, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08) :1305-1330
[7]   Simplified Monte Carlo dose calculation for therapeutic proton beams [J].
Kohno, R ;
Sakae, T ;
Takada, Y ;
Matsumoto, K ;
Matsuda, H ;
Nohtomi, A ;
Terunuma, T ;
Tsunashima, Y .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS, 2002, 41 (3A) :L294-L297
[8]   Range-modulated pencil beam algorithm for proton dose calculations [J].
Kohno, R ;
Takada, Y ;
Sakae, T ;
Nohtomi, A ;
Terunuma, T ;
Matsumoto, K ;
Matsuda, H .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2001, 40 (08) :5187-5193
[9]   Experimental evaluation of pencil beam algorithm by measurements of dose distributions of protons traversing an L-shaped phantom [J].
Kohno, R ;
Takada, Y ;
Sakae, T ;
Nohtomi, A ;
Terunuma, T ;
Yasuoka, K .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 2001, 40 (01) :441-445
[10]  
KOHNO R, 2003, IN PRESS J APPL PHYS