Analysis of partial volume correction on quantification and regional heterogeneity in cardiac PET

被引:2
|
作者
Turco, A. [1 ]
Nuyts, J. [1 ]
Duchenne, J. [2 ]
Gheysens, O. [1 ,3 ]
Voigt, J. U. [2 ,4 ]
Claus, P. [2 ]
Vunckx, K. [1 ]
机构
[1] Univ Leuven, Med Imaging Res Ctr,Dept Imaging,Pathology, Nucl Med,Mol imaging,MIRC,KU Leuven, B-3000 Leuven, Belgium
[2] Univ Leuven, Med Imaging Res Ctr,Dept Cardiovascular Sciences, Cardiology,MIRC,KU Leuven, B-3000 Leuven, Belgium
[3] Univ Hospitals Leuven, Dept Nucl Med, B-3000 Leuven, Belgium
[4] Univ Hospitals Leuven, Dept Cardiovascular Dis, B-3000 Leuven, Belgium
关键词
Cardiac PET; system resolution; partial volume correction; quantification; RECONSTRUCTION;
D O I
10.1007/s12350-016-0773-z
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The partial volume correction (PVC) of cardiac PET datasets using anatomical side information during reconstruction is appealing but not straightforward. Other techniques, which do not make use of additional anatomical information, could be equally effective in improving the reconstructed myocardial activity. Methods Resolution modeling in combination with different noise suppressing priors was evaluated as a means to perform PVC. Anatomical priors based on a high-resolution CT are compared to non-anatomical, edge-preserving priors (relative difference and total variation prior). The study is conducted on ex vivo datasets from ovine hearts. A simulation study additionally clarifies the relationship between prior effectiveness and myocardial wall thickness. Results Simple resolution modeling during data reconstruction resulted in over- and underestimation of activity, which hampers the absolute left ventricular quantification when compared to the ground truth. Both the edge-preserving and the anatomy-based PVC techniques improve the absolute quantification, with comparable results (Student t-test, P = .17). The relative tracer distribution was preserved with any reconstruction technique (repeated ANOVA, P = .98). Conclusions The use of edge-preserving priors emerged as optimal choice for quantification of tracer uptake in the left ventricular wall of the available datasets. Anatomical priors visually outperformed edge-preserving priors when the thinnest structures were of interest.
引用
收藏
页码:62 / 70
页数:9
相关论文
共 50 条
  • [1] Analysis of partial volume correction on quantification and regional heterogeneity in cardiac PET
    A. Turco
    J. Nuyts
    J. Duchenne
    O. Gheysens
    J. U. Voigt
    P. Claus
    K. Vunckx
    Journal of Nuclear Cardiology, 2020, 27 : 62 - 70
  • [2] PET quantification: Strategies for partial volume correction
    Bettinardi V.
    Castiglioni I.
    De Bernardi E.
    Gilardi M.C.
    Clinical and Translational Imaging, 2014, 2 (3) : 199 - 218
  • [3] Regional deconvolution method for partial volume correction in brain PET
    Rusinek, H
    Tsuia, WH
    de Leon, MJ
    MEDICAL IMAGING 2001: PHYSIOLOGY AND FUNCTION FROM MULTIDIMENSIONAL IMAGES, 2001, 4321 : 409 - 418
  • [4] Impact of Delineation and Partial Volume Effects Correction On PET Uptake Heterogeneity Quantification Through Textural Features Analysis for Therapy Response in Oncology
    Hatt, M.
    Tixier, F.
    Le Rest, C. Cheze
    Visvikis, D.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [5] Impact of VOI Size On PET Image Quantification Following Partial Volume Correction
    Chang, T.
    Chang, G.
    Pan, T.
    Clark, J.
    Mawlawi, O.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [6] The importance of appropriate partial volume correction for PET quantification in Alzheimer's disease
    Thomas, Benjamin A.
    Erlandsson, Kjell
    Modat, Marc
    Thurfjell, Lennart
    Vandenberghe, Rik
    Ourselin, Sebastien
    Hutton, Brian F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (06) : 1104 - 1119
  • [7] Absolute PET quantification with correction for partial volume effects within cerebral structures
    Labbe, C
    Koepp, M
    Ashburner, J
    Spinks, T
    Richardson, M
    Duncan, J
    Cunningham, V
    QUANTITATIVE FUNCTIONAL BRAIN IMAGING WITH POSITRON EMISSION TOMOGRAPHY, 1998, : 59 - 66
  • [8] The importance of appropriate partial volume correction for PET quantification in Alzheimer’s disease
    Benjamin A. Thomas
    Kjell Erlandsson
    Marc Modat
    Lennart Thurfjell
    Rik Vandenberghe
    Sebastien Ourselin
    Brian F. Hutton
    European Journal of Nuclear Medicine and Molecular Imaging, 2011, 38 : 1104 - 1119
  • [9] Partial Volume Correction Strategies in PET
    Rousset, Olivier
    Rahmim, Arman
    Alavi, Abass
    Zaidi, Habib
    PET CLINICS, 2007, 2 (02) : 235 - 249
  • [10] Partial Volume Correction of PET-Imaged Tumor Heterogeneity Using Expectation Maximization
    Barbee, D.
    Flynn, R.
    Holden, J.
    Jeraj, R.
    MEDICAL PHYSICS, 2008, 35 (06)