A rapid, automated VWF ristocetin cofactor activity assay improves reliability in the diagnosis of Von Willebrand disease

被引:13
|
作者
Bowyer, Annette E. [1 ,2 ]
Shepherd, Fiona [1 ]
Kitchen, Stephen [1 ]
Makris, Michael [1 ,2 ]
机构
[1] Royal Hallamshire Hosp, Sheffield Haemophilia & Thrombosis Ctr, Sheffield S10 2JF, S Yorkshire, England
[2] Univ Sheffield, Dept Cardiovasc Sci, Sheffield S10 2TN, S Yorkshire, England
关键词
VWF:RCo; Von Willebrand disease; Ristocetin; PREVALENCE;
D O I
10.1016/j.thromres.2010.11.029
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: The effective diagnosis and monitoring of Von Willebrand Disease (VWD) requires an accurate assessment of ristocetin co-factor activity (VWF:RCo). Current methodologies include automated platelet aggregometry and manual visual agglutination both of which are laborious to perform and notoriously subject to a high degree of inter and intra assay variation. Methods and Materials: We have evaluated an automated VWF: RCo assay (BC Von Willebrand Reagent, Siemens, Marberg, Germany) for use on the Sysmex CS2100i analyser (Milton Keynes, UK) and retrospectively compared the results with an in-house manual visual agglutination assay and VWF antigen (Siemens) in normal subjects and in 53 patients with various types of VWD and 23 patients following VWF therapeutic treatment. Results: The intra and interassay CV was improved with the automated assay (2.3% and 3.8% respectively) compared to 7% with the manual VWF: RCo assay. Good correlation was found between the two assays (r = 0.91) in 53 patients with VWD. The mean manual VWF: RCo was 0.25 IU/ml and mean automated VWF: RCo was 0.27 IU/ml. A comparable increase in VWF: RCo following treatment, mostly with Desmopressin, was found in 13 patients with type 1 VWD (mean 3.9 fold increase with manual VWF: RCo and 3.1 fold with the automated VWF: RCo). In 13 patients with type 2 or 3 VWD following treatment mostly with concentrate, a higher increase was found with the automated VWF: RCo assay than the manual assay (mean 11.9 fold manually and mean 20.3 automated). Conclusion: The automated VWF: RCo assay shows enhanced precision and analysis time in this difficult and time consuming laboratory test and its introduction should greatly improve the reliability of VWF testing. (C) 2010 Published by Elsevier Ltd.
引用
收藏
页码:341 / 344
页数:4
相关论文
共 50 条
  • [1] Comparison of a new automated von Willebrand factor activity assay with an aggregation von Willebrand ristocetin cofactor activity assay for the diagnosis of von Willebrand disease
    De Vleeschamer, An
    Devreese, Katrien
    BLOOD COAGULATION & FIBRINOLYSIS, 2006, 17 (05) : 353 - 358
  • [2] Evaluation of a rapid automated assay for analysis of von Willebrand ristocetin cofactor activity
    Strandberg, K
    Lethagen, S
    Andersson, K
    Carlson, A
    Hillarp, A
    CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2006, 12 (01) : 61 - 67
  • [3] Development of an ELISA method for testing VWF ristocetin cofactor activity with improved sensitivity and reliability in the diagnosis of von Willebrand disease
    Zhao, Yiming
    Gu, Ying
    Ji, Shundong
    Yang, Jianfeng
    Yu, Ziqian
    Ruan, Changgeng
    EUROPEAN JOURNAL OF HAEMATOLOGY, 2012, 88 (05) : 439 - 445
  • [4] A PLATE BASED PLASMA VON WILLEBRAND FACTOR (VWF):RISTOCETIN COFACTOR ACTIVITY ASSAY
    Chen, D.
    Pruthi, R. K.
    Nichols, W. L.
    Heit, J. A.
    Owen, W. G.
    AMERICAN JOURNAL OF HEMATOLOGY, 2008, 83 (12) : 942 - 942
  • [5] von Willebrand factor ristocetin cofactor (VWF:RCo) assay:: implementation on an automated coagulometer (ACL)
    Redaelli, R
    Corno, AR
    Borroni, L
    Mostarda, G
    Nichelatti, M
    Morra, E
    Baudo, F
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2005, 3 (12) : 2684 - 2688
  • [6] Validation of an automated assay to determine activity of ristocetin cofactor of von Willebrand factor
    Duboscq, Cristina
    Martinuzzo, Marta
    Manuel Ceresetto, Juan
    Carlos Otaso, Juan
    Barrera, Luis
    Maria Oyhamburu, Jose
    Stemmelin, German
    ACTA BIOQUIMICA CLINICA LATINOAMERICANA, 2016, 50 (02):
  • [7] Evaluation of an automated method for the measurement of Ristocetin cofactor activity of von Willebrand factor (VWF:RCo)
    Fortune, H.
    Rafferty, M.
    O'Connell, N.
    O'Donnell, J.
    Byrne, M.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 : 919 - 920
  • [8] A new plate based plasma von Willebrand factor (VWF): Ristocetin cofactor activity assay
    Chen, Dong
    Pruthi, Rijiv K.
    Nichols, William L.
    Heit, John A.
    Owen, Whyte G.
    BLOOD, 2007, 110 (11) : 638A - 638A
  • [9] Comparison of Von Willebrand factor (VWF) activity VWF:Ac with VWF ristocetin cofactor activity VWF:RCo
    Geisen, Ulrich
    Zieger, Barbara
    Nakamura, Lea
    Weis, Andreas
    Heinz, Juergen
    Michiels, Jan Jacques
    Heilmann, Claudia
    THROMBOSIS RESEARCH, 2014, 134 (02) : 246 - 250
  • [10] Improved performance characteristics of the von Willebrand factor ristocetin cofactor activity assay (VWF:RCo) using a novel automated assay protocol
    Hillarp, A.
    Stadler, M.
    Haderer, C.
    Weinberger, J.
    Kessler, C.
    Roemisch, J.
    Schwartz, B.
    HAEMOPHILIA, 2011, 17 (03) : 560 - 560