Novel Bifurcation Phantoms for Validation of Quantitative Coronary Angiography Algorithms

被引:12
作者
Girasis, Chrysafios [1 ]
Schuurbiers, Johan C. H. [1 ]
Onuma, Yoshinobu [1 ]
Serruys, Patrick W. [1 ]
Wentzel, Jolanda J. [1 ]
机构
[1] Erasmus MC, Biomech Lab, Dept Cardiol, NL-3000 CA Rotterdam, Netherlands
关键词
coronary angiography; quantitative coronary angiography; phantom; software validation; in vitro; SIROLIMUS-ELUTING STENTS; INTRAVASCULAR ULTRASOUND; COMPUTED-TOMOGRAPHY; CRUSH TECHNIQUE; MEDIUM-TERM; OFF-LINE; LESIONS; SYSTEM; CLASSIFICATION; ARTERIOGRAPHY;
D O I
10.1002/ccd.22704
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Validation is lacking for two- and three-dimensional (2D and 3D) bifurcation quantitative coronary angiography (QCA) algorithms. Methods: Six plexiglas phantoms were designed, each of them mimicking a coronary vessel with three successive bifurcations lesions, wherein at least one vessel segment had a percent diameter stenosis (DS) of >= 60%. The five most frequently occurring Medina classes (1,1,1), (1,1,0), (0,1,0), (0,1,1), and (1,0,0) were used in the design. Diameters of the daughter vessels in every bifurcation were dictated by the scaling law of Finet. Lesions were cosinus-shaped in longitudinal view and circular-shaped in cross-sectional view. At the level of the carina, lesions were becoming eccentric, favoring "plaque" at the outer bifurcation walls. Adjacent bifurcation lesions were kept distant by nontapering, stenosis-free segments of >= 10 mm length. The direction of the side branch relative to the main vessel was based on relevant literature. The phantoms were precision manufactured using computer-aided design and machining techniques. Because of the high drilling accuracy (within 10 mu m), the 3D luminal surface description of the phantom could be used to determine the true lumen dimensions and bifurcation angle (BA) values of the final geometry. Results: Our series of bifurcation phantoms comprised 33 narrowed and 21 stenosis-free vessel segments with a mean true minimal lumen diameter (MLD) value of 0.98 +/- 0.40 mm (range, 0.53-1.96 mm) and 2.29 +/- 0.74 mm (range, 1.40-4.00 mm), respectively. Overall, the mean true values for MLD, reference diameter, and DS were 1.49 +/- 0.85 mm, 2.70 +/- 0.71 mm, and 40.9% +/- 34.2%. The mean true values for the proximal and the distal BA were 123.6 degrees +/- 19.0 degrees and 69.6 degrees +/- 19.9 degrees, respectively. Conclusions: Six plexiglas phantoms containing a total of 18 bifurcations lesions with variable anatomy and Medina class were designed and precision manufactured to facilitate the validation of bifurcation QCA algorithms. (C) 2010 Wiley-Liss, Inc.
引用
收藏
页码:790 / 797
页数:8
相关论文
共 37 条
[1]   Final kissing balloon inflation by classic crush stenting did not improve the clinical outcomes for the treatment of unprotected left main bifurcation lesions: The importance of double-kissing crush technique [J].
Chen, Shaoliang ;
Zhang, Junjie ;
Ye, Fei ;
Chen, Yundai ;
Fang, Weiyi ;
Wei, Meng ;
He, Ben ;
Sun, Xuewen ;
Yang, Song ;
Kwan, Tak W. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2008, 71 (02) :166-172
[2]   Long-term outcomes after percutaneous coronary intervention of bifurcation narrowings [J].
Collins, Nicholas ;
Seidelin, Peter H. ;
Daly, Paul ;
Ivanov, Joan ;
Barolet, Alan ;
Mackie, Karen ;
Bui, Sanh ;
Schwartz, Leonard ;
Dzavik, Vladimir .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 102 (04) :404-410
[3]   Randomized study to evaluate sirolimus-eluting stents implanted at coronary bifurcation lesions [J].
Colombo, A ;
Moses, JW ;
Morice, MC ;
Ludwig, J ;
Holmes, DR ;
Spanos, V ;
Louvard, Y ;
Desmedt, B ;
Di Mario, C ;
Leon, MB .
CIRCULATION, 2004, 109 (10) :1244-1249
[4]   Randomized Study of the Crush Technique Versus Provisional Side-Branch Stenting in True Coronary Bifurcations The CACTUS (Coronary Bifurcations: Application of the Crushing Technique Using Sirolimus-Eluting Stents) Study [J].
Colombo, Antonio ;
Bramucci, Ezio ;
Sacca, Salvatore ;
Violini, Roberto ;
Lettieri, Corrado ;
Zanini, Roberto ;
Sheiban, Imad ;
Paloscia, Leonardo ;
Grube, Eberhard ;
Schofer, Joachim ;
Bolognese, Leonardo ;
Orlandi, Mario ;
Niccoli, Giampaolo ;
Latib, Azeem ;
Airoldi, Flavio .
CIRCULATION, 2009, 119 (01) :71-U118
[5]   QUANTITATIVE CORONARY ANGIOGRAPHY TO MEASURE PROGRESSION AND REGRESSION OF CORONARY ATHEROSCLEROSIS - VALUE, LIMITATIONS, AND IMPLICATIONS FOR CLINICAL-TRIALS [J].
DEFEYTER, PJ ;
SERRUYS, PW ;
DAVIES, MJ ;
RICHARDSON, P ;
LUBSEN, J ;
OLIVER, MF .
CIRCULATION, 1991, 84 (01) :412-423
[6]   Predictors of restenosis after treatment of bifurcational lesions with paclitaxel eluting stents: A multicenter prospective registry of 150 consecutive patients [J].
Di Mario, Carlo ;
Morici, Nuccia ;
Godino, Cosmo ;
Goktekin, Omer ;
Tamburino, Corrado ;
Barbagallo, Rossella ;
Antoniucci, David ;
Grube, Eberhard ;
Airoldi, Flavio ;
Zoccai, Giuseppe Biondi ;
Colombo, Antonio ;
Sangiorgi, Giuseppe M. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) :416-424
[7]   Predictors of long-term outcome after crush stenting of coronary bifurcation lesions: Importance of the bifurcation angle [J].
Dzavik, Vladimir ;
Kharbanda, Rajesh ;
Ivanov, Joan ;
Ing, Douglas J. ;
Bui, Sanh ;
Mackie, Karen ;
Ramsamujh, Rachael ;
Barolet, Alan ;
Schwartz, Leonard ;
Seidelin, Peter H. .
AMERICAN HEART JOURNAL, 2006, 152 (04) :762-769
[8]   Coronary artery plaque formation at coronary CT angiography: morphological analysis and relationship to hemodynamics [J].
Enrico, Benedetta ;
Suranyi, Pal ;
Thilo, Christian ;
Bonomo, Lorenzo ;
Costello, Philip ;
Schoepf, U. Joseph .
EUROPEAN RADIOLOGY, 2009, 19 (04) :837-844
[9]  
Finet Gerard, 2008, EuroIntervention, V3, P490, DOI 10.4244/EIJV3I4A87
[10]   PATTERNS IN VISUAL INTERPRETATION OF CORONARY ARTERIOGRAMS AS DETECTED BY QUANTITATIVE CORONARY ARTERIOGRAPHY [J].
FLEMING, RM ;
KIRKEEIDE, RL ;
SMALLING, RW ;
GOULD, KL ;
STUART, Y .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (04) :945-951