PFO closure complications from the AGA registry

被引:51
作者
Amin, Zahid [1 ]
Hijazi, Ziyad M. [1 ]
Bass, John L. [2 ]
Cheatham, John P. [3 ]
Hellenbrand, William [4 ]
Kleinman, Charles S. [4 ]
机构
[1] Rush Univ, Med Ctr, Chicago, IL 60612 USA
[2] Univ Minnesota, Childrens Hosp, Minneapolis, MN USA
[3] Nationwide Childrens Hosp, Columbus, OH USA
[4] Columbia Univ, Coll Phys & Surg, New York, NY USA
关键词
congenital heart disease in adults (CHDA); complications adult cath/intervention (COMP); patent foramen ovale/atrial septal defect (PFO);
D O I
10.1002/ccd.21582
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To evaluate all complications that occurred during or after cardiac catheterizations for Amplatzer PFO device closure of patent foramen ovale (PFO), determine the cause of the complications and recommend techniques to minimize complications in the future. Background: Rare complications were reported to the manufacturer of the Amplatzer PFO occluder since the introduction of the device. Methods: A panel of independent physicians reviewed all complications reported to the manufacturer to determine whether the complication was related to the device or related to the cardiac catheterization procedure. Demographic data, echocardiograms, operative reports, and time to occurrence of complications were reviewed. Results: A total of 11 events were reported. Only two patients had device related complications (erosion), an incidence of 0.018%. Two patients were found to have additional atrial septal defect after PFO closure. Two patients were thought to have an inflammatory reaction without any serious sequelae. Five complications were related to the cardiac catheterization procedure (atrial appendage perforation). Conclusions: Device related complications after Amplatzer PFO occluder placement are extremely rare. Cardiac catheterization related complications appear to be the most common cause of the hemodynamic compromise. Careful manipulation of catheters and wires, recognition of the location of the catheter by fluoroscopy and echocardiography will decrease the risk of such complications. (C) 2008 Wiley-Liss, Inc.
引用
收藏
页码:74 / 79
页数:6
相关论文
共 14 条
  • [1] Transcatheter patent foramen ovale closure for secondary prevention of paradoxical embolic events: Acute results from the FORECAST registry
    Alameddine, F
    Block, PC
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (04) : 512 - 516
  • [2] Erosion of amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk
    Amin, Z
    Hijazi, ZM
    Bass, JL
    Cheatham, JP
    Hellenbrand, WE
    Kleinman, CS
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) : 496 - 502
  • [3] Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder
    Divekar, A
    Gaamangwe, T
    Shaikh, N
    Raabe, M
    Ducas, J
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (08) : 1213 - 1218
  • [4] INCIDENCE AND SIZE OF PATENT FORAMEN OVALE DURING THE 1ST 10 DECADES OF LIFE - AN AUTOPSY STUDY OF 965 NORMAL HEARTS
    HAGEN, PT
    SCHOLZ, DG
    EDWARDS, WD
    [J]. MAYO CLINIC PROCEEDINGS, 1984, 59 (01) : 17 - 20
  • [5] Patent foramen ovale: Current pathology, pathophysiology, and clinical status
    Hara, H
    Virmani, R
    Ladich, E
    Mackey-Bojack, S
    Titus, J
    Reisman, M
    Gray, W
    Nakamura, M
    Mooney, M
    Poulose, A
    Schwartz, RS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (09) : 1768 - 1776
  • [6] Closure of patent foramen ovale for paradoxical emboli: Intermediate-term risk of recurrent neurological events following transcatheter device placement
    Hung, J
    Landzberg, MJ
    Jenkins, KJ
    King, MEE
    Lock, JE
    Palacios, IF
    Lang, P
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (05) : 1311 - 1316
  • [7] Indications for the closure of patent foramen ovale
    Landzberg, MJ
    Khairy, P
    [J]. HEART, 2004, 90 (02) : 219 - 224
  • [8] PREVALENCE OF PATENT FORAMEN OVALE IN PATIENTS WITH STROKE
    LECHAT, P
    MAS, JL
    LASCAULT, G
    LORON, P
    THEARD, M
    KLIMCZAC, M
    DROBINSKI, G
    THOMAS, D
    GROSGOGEAT, Y
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1988, 318 (18) : 1148 - 1152
  • [9] Iatrogenic atrial septal defect, erosion of the septum primum after device closure of a patent foramen ovale as a new medical entity
    Meier, B
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (01) : 165 - 168
  • [10] Patent foramen ovale and migraine
    Post, M. C.
    Luermans, J. G. L. M.
    Plokker, H. W. M.
    Budts, W.
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (01) : 9 - 14