Minidosimeters of L-alanine and 2-methylalanine (2MA) were prepared and tested as potential candidates for small radiation field dosimetry. To quantify the free radicals created by radiation a K-Band (24 GHz) EPR spectrometer was used. X-rays provided by a 6 MV clinical linear accelerator were used to irradiate the minidosimeters in the dose range of 0.5-30 Gy. The dose-response curves for both radiation sensitive materials displayed a good linear behavior in the dose range indicated with 2MA being more radiation sensitive than L-alanine. Moreover, 2MA showed a smaller LLD (lower limit detection) value. The proposed system minidosimeter/K-Band spectrometer was able to detect 10 Gy EPR spectra with good signal-to-noise ratio (S/N). The overall uncertainty indicates that this system shows a good performance for the detection of dose values of 20 Gy and above, which are dose values typically used in radiosurgery treatments. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:277 / 281
页数:5
相关论文
共 50 条
[1]   SOME OBSERVATIONS OF VIBRATIONAL-SPECTRUM OF SINGLE CRYSTALLINE L-ALANINE [J].
ADAMOWICZ, R ;
FISHMAN, E .
SPECTROCHIMICA ACTA PART A-MOLECULAR SPECTROSCOPY, 1972, A 28 (05) :889-+
[2]   SOURCES OF UNCERTAINTY IN THERAPY LEVEL ALANINE DOSIMETRY [J].
BARTOLOTTA, A ;
FATTIBENE, P ;
ONORI, S ;
PANTALONI, M ;
PETETTI, E .
APPLIED RADIATION AND ISOTOPES, 1993, 44 (1-2) :13-17
[3]   New developments in high field electron paramagnetic resonance with applications in structural biology [J].
Bennati, M ;
Prisner, TF .
REPORTS ON PROGRESS IN PHYSICS, 2005, 68 (02) :411-448
[4]   USE OF ALANINE AS A SOLID DOSIMETER [J].
BRADSHAW, WW ;
SPETZLER, HA ;
CADENA, DG ;
CRAWFORD, GW .
RADIATION RESEARCH, 1962, 17 (01) :11-+
[5]   Preliminary evaluation of second harmonic direct detection scheme for low-dose range in alanine/EPR dosimetry [J].
Chen, F ;
Graeff, CFO ;
Baffa, O .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :1357-1367
[6]   ESR/alanine dosimetry of high-energy electrons in radiotherapy [J].
Chu, S. ;
Wieser, A. ;
Feist, H. ;
Regulla, D.F. .
International journal of radiation applications and instrumentation. Part A, Applied radiation and isotopes, 1989, 40 (10-12) :993-996
[7]  
De Angelis C, 1999, PHYS MED BIOL, V44, P1181, DOI 10.1088/0031-9155/44/5/007
[8]  
Duggan D M, 1998, Med Dosim, V23, P153, DOI 10.1016/S0958-3947(98)00013-2
[9]   REFINEMENT OF L-ALANINE CRYSTAL STRUCTURE [J].
DUNITZ, JD ;
RYAN, RR .
ACTA CRYSTALLOGRAPHICA, 1966, 21 :617-&