The performance of gamma- and EtO-sterilised UHMWPE acetabular cups tested under severe simulator conditions. Part 2: wear particle characteristics with isolation protocols

被引:45
作者
Affatato, S
Bersaglia, G
Emiliani, D
Foltran, I
Taddei, P
Reggiani, M
Ferrieri, P
Toni, A
机构
[1] Ist Ortoped Rizzoli, Lab Tecnol Med, I-40136 Bologna, Italy
[2] Univ Bologna, Dipartimento Biochim G Moruzzi, Ctr Studio Spettroscopia Ramon, Sezione Chim & Propedeut Biochim, Bologna, Italy
[3] Univ Bologna, Dipartimento Sci Terra & Geol Ambientali, Bologna, Italy
[4] Div Ortoped & Traumatol 1, Bologna, Italy
关键词
polyethylene debris; sterilisation method; SEM analysis; micro-Raman spectroscopy;
D O I
10.1016/S0142-9612(03)00264-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultra-high molecular-weight polyethylene (UHMWPE) has been used in total joint replacement for the last three decades. Despite the recent advancements in prosthesis design, the wear of UHMWPE remains a serious clinical problem; the release of wear debris may induce osteolysis and implant loosening. Controlling the quality of the polyethylene is essential to improve its wear resistance and any potential adverse effect caused by processing, manufacturing or sterilisation should be avoided. To evaluate the influence of the sterilisation method (gamma-irradiation and ethylene oxide (EtO)-treatment) and third-body particles. gamma- and EtO-sterilised UHMWPE acetabular cups were tested against CoCrMo femoral heads in a hip joint simulator run for 2.5 million cycles in bovine calf serum in the presence of third-body polymethyltnethacrylate (PMMA) particles. A method not requiring ultra-centrifugation has been proposed for the isolation of polyethylene wear debris from the serum lubricant. SEM analysis allowed debris shape and morphology to be determined, and the wear mechanism operating in this study to be hypothesised, The morphological features of the wear debris were in agreement with clinical findings, enabling the hip simulator function to be validated. Micro-Raman spectroscopy coupled to PLS analysis showed that the mechanical friction during in vitro tests induced significant crystallinity changes in all the cups. The most significant changes were observed for the EtO-sterilised cups, which showed the highest weight loss. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:4045 / 4055
页数:11
相关论文
共 86 条
[1]   Mixed oxides prosthetic ceramic ball heads.: Part 1:: effect of the ZrO2 fraction on the wear of ceramic on polyethylene joints [J].
Affatato, S ;
Testoni, M ;
Cacciari, GL ;
Toni, A .
BIOMATERIALS, 1999, 20 (10) :971-975
[2]   Surface characterization and debris analysis of ceramic pairings after ten million cycles on a hip joint simulator [J].
Affatato, S ;
Ferrari, G ;
Chevalier, J ;
Ruggeri, O ;
Toni, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2002, 216 (H6) :419-424
[3]   The performance of gamma- and EtO-sterilised UHWMPE acetabular cups tested under severe simulator conditions. Part 1: role of the third-body wear process [J].
Affatato, S ;
Bersaglia, G ;
Foltran, I ;
Taddei, P ;
Fini, G ;
Toni, A .
BIOMATERIALS, 2002, 23 (24) :4839-4846
[4]   Effects of the sterilisation method on the wear of UHMWPE acetabular cups tested in a hip joint simulator [J].
Affatato, S ;
Bordini, B ;
Fagnano, C ;
Taddei, P ;
Tinti, A ;
Toni, A .
BIOMATERIALS, 2002, 23 (06) :1439-1446
[5]   Mixed-oxides prosthetic ceramic ball heads.: Part 2:: Effect of the ZrO2 fraction on the wear of ceramic on ceramic joints [J].
Affatato, S ;
Testoni, M ;
Cacciari, GL ;
Toni, A .
BIOMATERIALS, 1999, 20 (20) :1925-1929
[6]   Mixed oxides prosthetic ceramic ball heads.: Part 3:: effect of the ZrO2 fraction on the wear of ceramic on ceramic hip joint prostheses.: A long-term in vitro wear study [J].
Affatato, S ;
Goldoni, M ;
Testoni, M ;
Toni, A .
BIOMATERIALS, 2001, 22 (07) :717-723
[7]   Isolation and morphological characterisation of UHMWPE wear debris generated in vitro [J].
Affatato, S ;
Fernandes, B ;
Tucci, A ;
Esposito, L ;
Toni, A .
BIOMATERIALS, 2001, 22 (17) :2325-2331
[8]  
AFFATATO S, 2000, 12 C EUR SOC BIOM RO, P453
[9]   THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[10]   THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199