Energy spectra in water for the 6 MV X-ray therapeutic beam generated by Clinac-2300 linac

被引:29
|
作者
Konefal, Adam [1 ]
Bakoniak, Marzena [1 ]
Orlef, Andrzej [2 ]
Maniakowski, Zbigniew [2 ]
Szewczuk, Marek [2 ]
机构
[1] Silesian Univ, Inst Phys, Dept Nucl Phys & its Applicat, PL-40007 Katowice, Poland
[2] Ctr Oncol, Dept Med Phys, Gliwice Branch, PL-44100 Gliwice, Poland
关键词
Energy spectra in water; 6 MV therapeutic X-rays; MEASURED TRANSMISSION DATA; MONTE-CARLO CALCULATION; PHOTON-BEAM; BREMSSTRAHLUNG SPECTRA; DOSE CALCULATION; STOPPING-POWER; ELECTRON-BEAM; DOSIMETRY; RECONSTRUCTION; DISTRIBUTIONS;
D O I
10.1016/j.radmeas.2014.11.008
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
There is a lack of extensive data comprising energy spectra of therapeutic beams used in tele-radiotherapy, generated by medical linear accelerators. In particular there is a lack of the data for energy spectra in water. However, the spectra in water differ from those in air significantly because of strong photon scattering processes. The aim of this paper was the presentation of the wide range of detailed data for the 6 MV X-ray therapeutic beams from a medical linear accelerator Clinac-2300 by Varian. The presented data were derived by the use of Monte Carlo computer simulations (GEANT4 code). The performed investigations indicate that shapes of the spectra, total number of photon registered in a bin as well as mean energy of the considered beam depend on a depth in water, on a distance from the central-axis of the beam and on a radiation field size. However, shapes of the spectra as well as the beam mean energy does not depend on a bin size, independently of a depth in water and a radiation field whereas total number of photon registered in a bin is related to a bin size. Majority of the presented results were obtained for a cylindrical 1.41 cm(3) bin with the radius of 1.5 cm. The obtained data are very useful for the accurate absorbed depth-dose determination particularly outside the reference conditions, for advanced treatment planning systems, for constructors of medical accelerators etc. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:12 / 22
页数:11
相关论文
共 50 条
  • [21] Energy dependence of different materials in ESR dosimetry for clinical X-ray 10 MV beam
    Borgonove, A. F.
    Kinoshita, A.
    Chen, F.
    Nicolucci, P.
    Baffa, O.
    RADIATION MEASUREMENTS, 2007, 42 (6-7) : 1227 - 1232
  • [22] Effect of electron contamination of a 6 MV x-ray beam on near surface diode dosimetry
    Edwards, C. R.
    Mountford, P. J.
    Moloney, A. J.
    PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (24): : 6471 - 6482
  • [23] Broadband Detection of Dynamic Acoustic Emission Process Induced by 6 MV Therapeutic X-ray Beam from a Clinical Linear Accelerator
    Diao, Xianfen
    Zhu, Jing
    Li, Weihao
    Deng, Nan
    Chin, Chien Ting
    Zheng, Xiyuan
    Zhang, Xinyu
    Chen, Xin
    Li, Xianming
    Kuang, Yu
    2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [24] AN EXPERIMENTAL DETERMINATION OF ENERGY ABSORBED IN WATER FROM AN X-RAY BEAM
    ADAMS, GD
    RADIOLOGY, 1957, 69 (06) : 867 - 880
  • [25] Monte Carlo dosimetry of a cell culture irradiation model using a 6 MV X-ray beam
    Hernandez-Bojorquez, Mariana
    Trejo-Solis, Cristina
    Manuel Larraga-Gutierrez, Jose
    Martinez-Davalos, Arnulfo
    RADIATION PHYSICS AND CHEMISTRY, 2021, 180 (180)
  • [26] Peripheral dose measurements for X-ray beams of 6 and 18 MV: What is the effect of beam modifiers?
    Mazonakis, M.
    Zacharopoulou, F.
    Damilakis, J.
    Varveris, H.
    Gourtsoyiannis, N.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S141 - S142
  • [27] Dose Average LET of Electrons Generated Outside of 6 MV X-Ray Small Radiotherapy Field Edges
    Huerta-juan, Y.
    Xicohtencatl-hernandez, N.
    Massillon-JL, G.
    MEDICAL PHYSICS, 2022, 49 (06) : E494 - E494
  • [28] The Influence of a new Transmission Detector for "Online" Dose Verification on the Beam Characteristics of a 6MV X-Ray Therapy Beam
    Thoelking, J.
    Sekar, Y.
    Fleckenstein, J.
    Lohr, F.
    Wenz, F.
    Wertz, H.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2014, 190 : 60 - 60
  • [29] THE ENERGY CALIBRATION OF X-RAY ABSORPTION-SPECTRA USING MULTIPLE-BEAM DIFFRACTION
    HAGELSTEIN, M
    CUNIS, S
    FRAHM, R
    RABE, P
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 911 - 913
  • [30] A Study On the Influence of Photonuclear Reactions On the Biological Effectiveness of Therapeutic High Energy X-Ray Beam
    Wakita, A.
    Matsufuji, N.
    Kohno, T.
    Kodaira, S.
    Yokoyama, K.
    Suzuki, Y.
    Itami, J.
    MEDICAL PHYSICS, 2014, 41 (06) : 227 - 227