Separation and simultaneous determination of uric acid and ascorbic acid on a dynamically modified poly(dimethylsiloxane) microchip

被引:32
作者
Qiu, Jianding [1 ]
Hu, Pengfei [1 ]
Liang, Ruping [1 ]
机构
[1] Nanchang Univ, Dept Chem, Nanchang 330031, Peoples R China
关键词
D O I
10.2116/analsci.23.1409
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Poly(dimethylsiloxane) microfluidic channels alternately modified by poly(diallyldimethylammonium chloride) and poly(sodium 4-styrenesulfonate) were successfully used to separate uric acid and ascorbic acid. Results show that uric acid and ascorbicacid can be well separated and detected simultaneously in modified microchips coupled with in-channel electrochemical detection. Under the optimal conditions, the linear ranges of uric acid and ascorbic acid were both from 25 to 600 mu M, with the correlation coefficients of 0.997 and 0.996, respectively. The detection limits were 8 mu M for uric acid and 5 mu M for ascorbic acid. Factors influencing separation and detection, including buffer solution, detection potential and separation voltage, were investigated and optimized. In addition, the dependences of the current response on sensitivity and reproducibility were studied, and the stability of the device was also evaluated in detail. This method was successfully used to determine uric acid and ascorbic acid in human urine.
引用
收藏
页码:1409 / 1414
页数:6
相关论文
共 42 条
[1]   Ascorbic acid: much more than just an antioxidant [J].
Arrigoni, O ;
De Tullio, MC .
BIOCHIMICA ET BIOPHYSICA ACTA-GENERAL SUBJECTS, 2002, 1569 (1-3) :1-9
[2]  
Badal MY, 2002, J CHROMATOGR A, V947, P277
[3]   Fast and simultaneous detection of prominent natural antioxidants using analytical microsystems for capillary electrophoresis with a glassy carbon electrode:: A new gateway to food environments [J].
Blasco, AJ ;
Barrigas, I ;
González, MC ;
Escarpa, A .
ELECTROPHORESIS, 2005, 26 (24) :4664-4673
[4]  
CARLOS DG, 2005, ELECTROPHORESIS, V26, P703
[5]   Simultaneous detection of vitamin C and uric acid by capillary electrophoresis in plasma of diabetes and in aqueous humor in acute anterior uveitis [J].
Cheng, ML ;
Liu, TZ ;
Lu, FJ ;
Chiu, DTY .
CLINICAL BIOCHEMISTRY, 1999, 32 (06) :473-476
[6]   RESPONSE OF MICROVOLTAMMETRIC ELECTRODES TO HOMOGENEOUS CATALYTIC AND SLOW HETEROGENEOUS CHARGE-TRANSFER REACTIONS [J].
DAYTON, MA ;
EWING, AG ;
WIGHTMAN, RM .
ANALYTICAL CHEMISTRY, 1980, 52 (14) :2392-2396
[7]   Detection of uric acid with a new type of conducting polymer-based enzymatic sensor by bipotentiostatic technique [J].
Dobay, R ;
Harsányi, G ;
Visy, C .
ANALYTICA CHIMICA ACTA, 1999, 385 (1-3) :187-194
[8]   Separation of proteins on surface-modified poly(dimethylsiloxane) microfluidic devices [J].
Dou, YH ;
Bao, N ;
Xu, JJ ;
Meng, F ;
Chen, HY .
ELECTROPHORESIS, 2004, 25 (17) :3024-3031
[9]  
Dou YH, 2002, ELECTROPHORESIS, V23, P3558, DOI 10.1002/1522-2683(200210)23:20<3558::AID-ELPS3558>3.0.CO
[10]  
2-#