Measured Dose versus Organ Dose Performance in Digital Mammography Systems

被引:0
|
作者
McCullagh, James B. [1 ]
Baldelli, Paola [2 ]
Phelan, Niall [2 ]
机构
[1] BreastCheck, Natl Canc Screening Serv, Infirm Rd, Cork, Ireland
[2] Natl Canc Screening Serv, Dublin 7, Ireland
来源
DIGITAL MAMMOGRAPHY | 2010年 / 6136卷
关键词
Digital Mammography; Radiation Dose; Mean Glandular Dose; Quality Assurance;
D O I
暂无
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
The results of a dose survey of digital x-ray imaging equipment in a breast screening programme have been used to compare the measured dose values to the dose values displayed by the mammography systems. A total of 12,104 images were used in the survey, acquired on three types of digital mammography system: GE Essential, Ho logic Selenia and Sectra MDM L30. The results reveal that estimates of the MGD were largely similar between measured and displayed dose for all models in the survey. It was found that large discrepancies could occur on individual systems which may be attributable to various factors including inaccurate QA measurement data and poorly calibrated mammography systems.
引用
收藏
页码:86 / +
页数:2
相关论文
共 50 条
  • [31] Clinical performance of low dose positron emission mammography
    Mercier, Gustavo
    Vassilakis, Nectaria
    Brennan, Jennifer
    DeVita, Frank
    Kachnic, Lisa
    JOURNAL OF NUCLEAR MEDICINE, 2012, 53
  • [32] Mammography and fetal dose
    Behrman, Richard H.
    Homer, Marc J.
    RADIOLOGY, 2007, 243 (02) : 605 - 605
  • [33] DOSE REDUCTION IN MAMMOGRAPHY
    MINTZER, RA
    MATTHIES, HJ
    LIN, PP
    NEIMAN, HL
    ROGERS, LF
    SURGERY GYNECOLOGY & OBSTETRICS, 1978, 146 (01): : 79 - 80
  • [34] Dose and diagnostic performance comparison between phase-contrast mammography with synchrotron radiation and digital mammography: a clinical study report
    Fedon, Christian
    Rigon, Luigi
    Arfelli, Fulvia
    Dreossi, Diego
    Quai, Elisa
    Tonutti, Maura
    Tromba, Giuliana
    Cova, Maria Assunta
    Longo, Renata
    JOURNAL OF MEDICAL IMAGING, 2018, 5 (01)
  • [35] DOSE REDUCTION IN MAMMOGRAPHY
    MINTZER, RA
    MATTHIES, HJ
    LIN, PJP
    NEIMAN, HL
    ROGERS, LF
    INVESTIGATIVE RADIOLOGY, 1977, 12 (05) : 465 - 466
  • [36] Performance evaluation of six digital mammography systems
    Perez, Alessandra Maia Marques Martinez
    Lopes, Laura Alcantara Silva
    Caron, Renato Franca
    Oliveira, Bruno Beraldo
    Poletti, Martin Eduardo
    RADIATION PHYSICS AND CHEMISTRY, 2024, 218
  • [37] Performance of full-field digital mammography versus digital breast
    Wang, Mengru
    Zhuang, Shan
    Sheng, Liuli
    Zhao, Yu Nian
    Shen, Wenrong
    PRECISION MEDICAL SCIENCES, 2022, 11 (02): : 56 - 61
  • [38] Digital breast tomosynthesis versus digital mammography: a clinical performance study
    Gisella Gennaro
    Alicia Toledano
    Cosimo di Maggio
    Enrica Baldan
    Elisabetta Bezzon
    Manuela La Grassa
    Luigi Pescarini
    Ilaria Polico
    Alessandro Proietti
    Aida Toffoli
    Pier Carlo Muzzio
    European Radiology, 2010, 20 : 1545 - 1553
  • [39] Digital breast tomosynthesis versus digital mammography: a clinical performance study
    Gennaro, Gisella
    Toledano, Alicia
    di Maggio, Cosimo
    Baldan, Enrica
    Bezzon, Elisabetta
    La Grassa, Manuela
    Pescarini, Luigi
    Polico, Ilaria
    Proietti, Alessandro
    Toffoli, Aida
    Muzzio, Pier Carlo
    EUROPEAN RADIOLOGY, 2010, 20 (07) : 1545 - 1553
  • [40] Exposure Parameters and Average Glandular Dose in a Digital Mammography System in Japan
    Nishide, Hiroko
    Ohta, Kouji
    Murata, Kaori
    Kodera, Yoshie
    Shimamoto, Kazuhiro
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2018, 8 (05) : 1030 - 1034