Integrated circuit-based instrumentation for microchip capillary electrophoresis

被引:9
作者
Behnam, M. [1 ]
Kaigala, G. V. [1 ]
Khorasani, M. [1 ]
Martel, S. [2 ]
Elliott, D. G. [1 ]
Backhouse, C. J. [1 ]
机构
[1] Univ Alberta, Dept Elect & Comp Engn, ECERF, Edmonton, AB T6G 2V4, Canada
[2] DALSA Semicond, Bromont, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
ON-A-CHIP; MICROFLUIDIC APPLICATIONS; FLUORESCENCE DETECTION; BIOLOGICAL-FLUIDS; SYSTEMS; MICROSYSTEM; DNA; SEPARATIONS; PLATFORM; DEVICE;
D O I
10.1049/iet-nbt.2009.0018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Although electrophoresis with laser-induced fluorescence (LIF) detection has tremendous potential in lab on chip-based point-of-care disease diagnostics, the wider use of microchip electrophoresis has been limited by the size and cost of the instrumentation. To address this challenge, the authors designed an integrated circuit (IC, i.e. a microelectronic chip, with total silicon area of <0.25 cm(2), less than 5 mm x 5 mm, and power consumption of 28 mW), which, with a minimal additional infrastructure, can perform microchip electrophoresis with LIF detection. The present work enables extremely compact and inexpensive portable systems consisting of one or more complementary metal-oxide-semiconductor (CMOS) chips and several other low-cost components. There are, to the authors' knowledge, no other reports of a CMOS-based LIF capillary electrophoresis instrument (i.e. high voltage generation, switching, control and interface circuit combined with LIF detection). This instrument is powered and controlled using a universal serial bus (USB) interface to a laptop computer. The authors demonstrate this IC in various configurations and can readily analyse the DNA produced by a standard medical diagnostic protocol (end-labelled polymerase chain reaction (PCR) product) with a limit of detection of similar to 1 ng/mu l (similar to 1 ng of total DNA). The authors believe that this approach may ultimately enable lab-on-a-chip-based electrophoretic instruments that cost on the order of several dollars.
引用
收藏
页码:91 / 101
页数:11
相关论文
共 42 条
[1]  
Allen PhillipE., 2002, CMOS ANALOG CIRCUIT, V2nd
[2]  
ASOGAWA M, 2008, P 12 INT C MIN SYST, P1072
[3]   Lab-on-a-chip with integrated optical transducers [J].
Balslev, S ;
Jorgensen, AM ;
Bilenberg, B ;
Mogensen, KB ;
Snakenborg, D ;
Geschke, O ;
Kutter, JP ;
Kristensen, A .
LAB ON A CHIP, 2006, 6 (02) :213-217
[4]   An integrated CMOS high voltage supply for lab-on-a-chip systems [J].
Behnam, M. ;
Kaigala, G. V. ;
Khorasani, M. ;
Marshall, P. ;
Backhouse, C. J. ;
Elliott, D. G. .
LAB ON A CHIP, 2008, 8 (09) :1524-1529
[5]   Microfabricated bioprocessor for integrated nanoliter-scale Sanger DNA sequencing [J].
Blazej, Robert G. ;
Kumaresan, Palani ;
Mathies, Richard A. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2006, 103 (19) :7240-7245
[6]   An integrated nanoliter DNA analysis device [J].
Burns, MA ;
Johnson, BN ;
Brahmasandra, SN ;
Handique, K ;
Webster, JR ;
Krishnan, M ;
Sammarco, TS ;
Man, PM ;
Jones, D ;
Heldsinger, D ;
Mastrangelo, CH ;
Burke, DT .
SCIENCE, 1998, 282 (5388) :484-487
[7]   Sample stacking revisited: A personal perspective [J].
Chien, RL .
ELECTROPHORESIS, 2003, 24 (03) :486-497
[8]   Compact, high voltage power supply for the lab-on-a-chip [J].
Collins, GE ;
Wu, P ;
Lu, Q ;
Ramsey, JD ;
Bromund, RH .
LAB ON A CHIP, 2004, 4 (04) :408-411
[9]   A fully integrated microfluidic genetic analysis system with sample-in-answer-out capability [J].
Easley, Christopher J. ;
Karlinsey, James M. ;
Bienvenue, Joan M. ;
Legendre, Lindsay A. ;
Roper, Michael G. ;
Feldman, Sanford H. ;
Hughes, Molly A. ;
Hewlett, Erik L. ;
Merkel, Tod J. ;
Ferrance, Jerome P. ;
Landers, James P. .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2006, 103 (51) :19272-19277
[10]   A miniaturized high-voltage integrated power supply for portable microfluidic applications [J].
Erickson, D ;
Sinton, D ;
Li, DQ .
LAB ON A CHIP, 2004, 4 (02) :87-90