Ultrasensitive Photoelectrochemical Immunoassay for Matrix Metalloproteinase-2 Detection Based on CdS:Mn/CdTe Cosensitized TiO2 Nanotubes and Signal Amplification of SiO2@Ab2 Conjugates

被引:145
作者
Fan, Gao-Chao [1 ]
Han, Li [1 ]
Zhu, Hua [1 ]
Zhang, Jian-Rong [1 ,2 ]
Zhu, Jun-Jie [1 ]
机构
[1] Nanjing Univ, Sch Chem & Chem Engn, State Key Lab Analyt Chem Life Sci, Nanjing 210093, Jiangsu, Peoples R China
[2] Nanjing Univ, Sch Chem & Life Sci, Jinling Coll, Nanjing 210089, Jiangsu, Peoples R China
关键词
CDS QUANTUM DOTS; MESOPOROUS SILICA NANOPARTICLES; DRUG-DELIVERY; PROTEASE ACTIVITY; IMMUNOSENSOR; CARCINOMA; STRATEGY; NANOCRYSTALS; CELL; ANGIOGENESIS;
D O I
10.1021/ac504027d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An ultrasensitive photoelectrochemical sandwich immunoassay was developed to detect matrix metalloproteinase-2 (MMP-2, antigen, Ag) based on CdS:Mn/CdTe cosensitized TiO2 nanotubes (TiO2-NTs) and signal amplification of SiO2@Ab(2) conjugates. Specifically, the TiO2-NTs electrode was first deposited with CdS:Mn by successive ionic layer adsorption and reaction technique and then further coated with CdTe quantum dots (QDs) via the layer-by-layer method, forming TiO2-NTs/CdS:Mn/CdTe cosensitized structure, which was employed as a matrix to immobilize capture MMP-2 antibodies (Ab(1)); whereas, SiO2 nanoparticles were coated with signal MMP-2 antibodies (Ab(2)) to form SiO2@Ab(2) conjugates, which were used as signal amplification elements via the specific antibody-antigen immunoreaction between Ag and Ab(2). The ultrahigh sensitivity of this immunoassay derived from the two major reasons as below. First, the TiO2-NTs/CdS:Mn/CdTe cosensitized structure could adequately absorb the light energy, dramatically promote electron transfer, and effectively inhibit the electron-hole recombination, resulting in significantly enhanced photocurrent intensity of the sensing electrode. However, in the presence of target Ag, the immobilized SiO2@Ab(2) conjugates could evidently increase the steric hindrance of the sensing electrode and effectively depress the electron transfer, leading to obviously decreased photocurrent intensity. Accordingly, the well-designed photoelectrochemical immunoassay exhibited a low detection limit of 3.6 fg/mL and a wide linear range from 10 fg/mL to 500 pg/mL for target Ag detection. Meanwhile, it also presented good reproducibility, specificity, and stability and might open a new promising platform for the detection of other important biomarkers.
引用
收藏
页码:12398 / 12405
页数:8
相关论文
共 45 条
  • [1] A Photoelectrochemical Immunosensor Based on Au-Doped TiO2 Nanotube Arrays for the Detection of α-Synuclein
    An, Yarui
    Tang, Linlin
    Jiang, Xiaoli
    Chen, Hua
    Yang, Meicheng
    Jin, Litong
    Zhang, Shengping
    Wang, Chuangui
    Zhang, Wen
    [J]. CHEMISTRY-A EUROPEAN JOURNAL, 2010, 16 (48) : 14439 - 14446
  • [2] [Anonymous], 2008, ANGEW CHEM INT EDIT, DOI DOI 10.1002/ANGE.200802469
  • [3] Cancer therapy - Matrix metalloproteinase inhibitors and cancer: Trials and tribulations
    Coussens, LM
    Fingleton, B
    Matrisian, LM
    [J]. SCIENCE, 2002, 295 (5564) : 2387 - 2392
  • [4] A new signal amplification strategy of photoelectrochemical immunoassay for highly sensitive interleukin-6 detection based on TiO2/CdS/CdSe dual co-sensitized structure
    Fan, Gao-Chao
    Ren, Xiao-Lin
    Zhu, Cheng
    Zhang, Jian-Rong
    Zhu, Jun-Jie
    [J]. BIOSENSORS & BIOELECTRONICS, 2014, 59 : 45 - 53
  • [5] Induction of cell migration by matrix metalloprotease-2 cleavage of laminin-5
    Giannelli, G
    FalkMarzillier, J
    Schiraldi, O
    StetlerStevenson, WG
    Quaranta, V
    [J]. SCIENCE, 1997, 277 (5323) : 225 - 228
  • [6] Giannopoulos G, 2008, ANTICANCER RES, V28, P1875
  • [7] Photoelectrochemical immunosensor for label-free detection and quantification of anti-cholera toxin antibody
    Haddour, Naoufel
    Chauvin, Jerome
    Gondran, Chantal
    Cosnier, Serge
    [J]. JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2006, 128 (30) : 9693 - 9698
  • [8] Tissue MMP-2/TIMP-2-complex are better prognostic factors than serum MMP-2, MMP-9 or TIMP-1 in Stage I-III lung carcinoma
    Hoikkala, S.
    Paakko, P.
    Soini, Y.
    Makitaro, R.
    Kinnula, V.
    Turpeenniemi-Hujanen, T.
    [J]. CANCER LETTERS, 2006, 236 (01) : 125 - 132
  • [9] Photoelectrochemical detection of pentachlorophenol with a Multiple Hybrid CdSexTe1-x/TiO2 Nanotube Structure-Based Label-Free Immunosensor
    Kang, Qing
    Yang, Lixia
    Chen, Yufang
    Luo, Shenglian
    Wen, Lingfei
    Cai, Qingyun
    Yao, Shouzhuo
    [J]. ANALYTICAL CHEMISTRY, 2010, 82 (23) : 9749 - 9754
  • [10] Matrix Metalloproteinases: Regulators of the Tumor Microenvironment
    Kessenbrock, Kai
    Plaks, Vicki
    Werb, Zena
    [J]. CELL, 2010, 141 (01) : 52 - 67