Serial Doppler echocardiographic evaluation of small-sized Sorin Bicarbon prostheses

被引:13
作者
De Carlo, M [1 ]
Milano, AD [1 ]
Nardi, C [1 ]
Mecozzi, G [1 ]
Bortolotti, U [1 ]
机构
[1] Univ Pisa, Sch Med, Cardiothorac Dept, I-56100 Pisa, Italy
关键词
D O I
10.1016/S0022-5223(02)73395-3
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The Sorin Bicarbon prosthesis (Sorin Biomedica, Saluggia, Italy) is a bileaflet valve with curved-profile leaflets, a rolling hinge mechanism, and a pyrolytic carbon-coated titanium alloy housing and sewing ring. Although the Sorin Bicarbon prosthesis has been implanted in greater than 80,000 patients, and reference values on the hemodynamic performance of valve prostheses are needed to avoid patient-prosthesis mismatch, few Doppler echocardiographic data are available on the prosthesis in the aortic position. The aim of this study is to provide a detailed echocardiographic evaluation of the hemodynamic performance and regression of left ventricular hypertrophy after aortic valve replacement with the Sorin Bicarbon prosthesis. Methods: The study included 182 patients who received a 21-mm (n = 61) or 23-mm (n = 121) Sorin Bicarbon prosthesis for pure or prevalent aortic stenosis who underwent serial echocardiograms at 3, 6, and 12 months after aortic. valve replacement. Results: Mean and peak gradients significantly decreased (P < .001) during follow-up to values of 12 +/- 3 and 22 +/- 6 mm Hg for the 21-mm prosthesis and values of 11 +/- 4 and 19 +/- 6 mm Hg for the 23-mm prosthesis at I year. Left ventricular mass index showed a 17% decrease to 120 +/- 27 g/m(2) in recipients of the 21-mm prosthesis (P < .001),and a 21% decrease to 123 +/- 29 g/m(2) in recipients of the 23-mm prosthesis (P < .001). A larger prosthesis size was the only predictor of a higher left ventricular mass index regression. Among recipients of the 21-mm prosthesis, body surface area of greater than 1.85 m(2) was,associated with a lower regression of left ventricular mass index. The effective orifice area index was 1.00 +/- 0.11 and 1.08 +/- 0.14 cm(2)/m(2) in recipients of the 21-mm and 23-mm prostheses, respectively. Conclusions: Size 21 mm and 23 mm Sorin Bicarbon prostheses show low transprosthetic gradients, with significant reduction of left ventricular mass index during the first postoperative year. The reported effective orifice areas might be useful for aortic valve replacement in patients with a small aortic annulus to avoid patient-prosthesis mismatch.
引用
收藏
页码:337 / 343
页数:7
相关论文
共 27 条
  • [1] Normal echocardiographic characteristics of the Sorin Bicarbon bileaflet prosthetic heart valve in the mitral and aortic positions
    Badano, L
    Mocchegiani, R
    Bertoli, D
    DeGaetano, G
    Carratino, L
    Pasetti, L
    Caudullo, M
    Budini, A
    Mannello, B
    Passerone, G
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1997, 10 (06) : 632 - 643
  • [2] Influence of aortic valve replacement, prosthesis type, and size on functional outcome and ventricular mass in patients with aortic stenosis
    Bech-Hanssen, O
    Caidahl, K
    Wall, B
    Mykén, P
    Larsson, S
    Wallentin, I
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1999, 118 (01) : 57 - 65
  • [3] Borman J B, 1996, J Cardiovasc Surg (Torino), V37, P43
  • [4] Bicarbon valve - European multicenter clinical evaluation
    Borman, JB
    Brands, WGB
    Camilleri, L
    Cotrufo, M
    Daenen, W
    Gandjbakhch, I
    Infantes, C
    Khayat, A
    Laborde, F
    Pellegrini, A
    Piwnica, A
    Reichart, B
    Sharony, R
    Walesby, R
    Warembourg, H
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1998, 13 (06) : 685 - 693
  • [5] Borman JB, 1998, J CARDIOVASC SURG, V39, P99
  • [6] Casselman F, 1997, J HEART VALVE DIS, V6, P410
  • [7] ECHOCARDIOGRAPHIC DESCRIPTION OF THE CARBOMEDICS BILEAFLET PROSTHETIC HEART-VALVE
    CHAMBERS, J
    CROSS, J
    DEVERALL, P
    SOWTON, E
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (02) : 398 - 405
  • [8] De Carlo M, 1999, J HEART VALVE DIS, V8, P522
  • [9] ECHOCARDIOGRAPHIC ASSESSMENT OF LEFT-VENTRICULAR HYPERTROPHY - COMPARISON TO NECROPSY FINDINGS
    DEVEREUX, RB
    ALONSO, DR
    LUTAS, EM
    GOTTLIEB, GJ
    CAMPO, E
    SACHS, I
    REICHEK, N
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) : 450 - 458
  • [10] Di Salvo C, 1996, J Cardiovasc Surg (Torino), V37, P13