Evaluation of inter- and intra-fraction 6D motion for stereotactic body radiation therapy of spinal metastases: influence of treatment time

被引:13
作者
Henni, Ahmed Hadj [1 ]
Gensanne, David [1 ]
Roge, Maximilien [1 ]
Hanzen, Chantal [1 ]
Bulot, Guillaume [1 ]
Colard, Elyse [1 ]
Thureau, Sebastien [1 ]
机构
[1] Ctr Henri Becquerel, 1 Rue Amiens, F-76000 Rouen, France
关键词
SBRT; CBCT; Spinal; Inter-fraction; Intra-fraction; Treatment time; CONE-BEAM CT; MODULATED ARC THERAPY; INTRAFRACTION MOTION; CONSENSUS GUIDELINES; RADIOTHERAPY; IMMOBILIZATION; ACCURACY; GUIDANCE;
D O I
10.1186/s13014-021-01892-5
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background The objective of this study was to analyze the amplitude of translational and rotational movements occurring during stereotactic body radiotherapy (SBRT) of spinal metastases in two different positioning devices. The relevance of intra-fractional imaging and the influence of treatment time were evaluated. Methods Twenty patients were treated in the supine position either (1) on a body vacuum cushion with arms raised and resting on a clegecel or (2) on an integrated SBRT solution consisting of a SBRT table top, an Orfit (TM) AIO system, and a vacuum cushion. Alignments between the cone beam computed tomography (CBCT) and the planning computed tomography allowed corrections of inter- and intra-fraction positional shifts using a 6D table. The absolute values of the translational and rotational setup errors obtained for 329 CBCT were recorded. The translational 3D vector, the maximum angle, and the characteristic times of the treatment fractions were calculated. Results An improvement in the mean (SD) inter-fraction 3D vector (mm) from 7.8 (5.9) to 5.9 (3.8) was obtained by changing the fixation devices from (1) to (2) (p < 0.038). The maximum angles were less than 2 degrees for a total of 87% for (1) and 96% for (2). The mean (SD) of the intra-fraction 3D vectors (mm) was lower for the new 1.1 (0.8) positioning fixation (2) compared to the old one (1) 1.7 (1.7) (p = 0.004). The angular corrections applied in the intra-fraction were on average very low (0.4 degrees) and similar between the two systems. A strong correlation was found between the 3D displacement vector and the fraction time for (1) and (2) with regression coefficients of 0.408 (0.262-0.555, 95% CI) and 0.069 (0.010-0.128, 95% CI), respectively. An accuracy of 1 mm would require intra-fraction imaging every 5 min for both systems. If the expected accuracy was 2 mm, then only system (2) could avoid intra-fractional imaging. Conclusions This study allowed us to evaluate setup errors of two immobilization devices for spine SBRT. The association of inter- and intra-fraction imaging with 6D repositioning of a patient is inevitable. The correlation between treatment time and corrections to be applied encourages us to move toward imaging modalities which allow a reduction in fraction time.
引用
收藏
页数:9
相关论文
共 28 条
[1]   Automatic image registration performance for two different CBCT systems; variation with imaging dose [J].
Barber, J. ;
Sykes, J. R. ;
Holloway, L. ;
Thwaites, D. I. .
XVII INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY (ICCR 2013), 2014, 489
[2]   Stereotactic body radiation therapy: The report of AAPM Task Group 101 [J].
Benedict, Stanley H. ;
Yenice, Kamil M. ;
Followill, David ;
Galvin, James M. ;
Hinson, William ;
Kavanagh, Brian ;
Keall, Paul ;
Lovelock, Michael ;
Meeks, Sanford ;
Papiez, Lech ;
Purdie, Thomas ;
Sadagopan, Ramaswamy ;
Schell, Michael C. ;
Salter, Bill ;
Schlesinger, David J. ;
Shiu, Almon S. ;
Solberg, Timothy ;
Song, Danny Y. ;
Stieber, Volker ;
Timmerman, Robert ;
Tome, Wolfgang A. ;
Verellen, Dirk ;
Wang, Lu ;
Yin, Fang-Fang .
MEDICAL PHYSICS, 2010, 37 (08) :4078-4101
[3]   6D image guidance for spinal non-invasive stereotactic body radiation therapy: Comparison between ExacTrac X-ray 6D with kilo-voltage cone-beam CT [J].
Chang, Zheng ;
Wang, Zhiheng ;
Ma, Jinli ;
O'Daniel, Jennifer C. ;
Kirkpatrick, John ;
Yin, Fang-Fang .
RADIOTHERAPY AND ONCOLOGY, 2010, 95 (01) :116-121
[4]   International Spine Radiosurgery Consortium Consensus Guidelines for Target Volume Definition in Spinal Stereotactic Radiosurgery [J].
Cox, Brett W. ;
Spratt, Daniel E. ;
Lovelock, Michael ;
Bilsky, Mark H. ;
Lis, Eric ;
Ryu, Samuel ;
Sheehan, Jason ;
Gerszten, Peter C. ;
Chang, Eric ;
Gibbs, Iris ;
Soltys, Scott ;
Sahgal, Arjun ;
Deasy, Joe ;
Flickinger, John ;
Quader, Mubina ;
Mindea, Stefan ;
Yamada, Yoshiya .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (05) :E597-E605
[5]   Stereotactic body radiotherapy for spine and bony pelvis using flattening filter free volumetric modulated arc therapy, 6D cone-beam CT and simple positioning techniques: Treatment time and patient stability [J].
Dahele, Max ;
Slotman, Ben ;
Verbakel, Wilko .
ACTA ONCOLOGICA, 2016, 55 (06) :795-798
[6]   International consensus recommendations for target volume delineation specific to sacral metastases and spinal stereotactic body radiation therapy (SBRT) [J].
Dunne, Emma M. ;
Sahgal, Arjun ;
Lo, Simon S. ;
Bergman, Alanah ;
Kosztyla, Robert ;
Dea, Nicolas ;
Chang, Eric L. ;
Chang, Ung-Kyu ;
Chao, Samuel T. ;
Faruqi, Salman ;
Ghia, Amol J. ;
Redmond, Kristin J. ;
Soltys, Scott G. ;
Liu, Mitchell C. .
RADIOTHERAPY AND ONCOLOGY, 2020, 145 :21-29
[7]   Inter- and intra-fraction motion in stereotactic body radiotherapy for spinal and paraspinal tumours using cone-beam CT and positional correction in six degrees of freedom [J].
Finnigan, Renee ;
Lamprecht, Brock ;
Barry, Tamara ;
Jones, Kimberley ;
Boyd, Joshua ;
Pullar, Andrew ;
Burmeister, Bryan ;
Foote, Matthew .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2016, 60 (01) :112-118
[8]   Localization Accuracy and Immobilization Effectiveness of a Stereotactic Body Frame for a Variety of Treatment Sites [J].
Foster, Ryan ;
Meyer, Jeffrey ;
Iyengar, Puneeth ;
Pistenmaa, David ;
Timmerman, Robert ;
Choy, Hak ;
Solberg, Timothy .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (05) :911-916
[9]   Loss of biological effect in prolonged fraction delivery [J].
Fowler, JF ;
Welsh, JS ;
Howard, SP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 59 (01) :242-249
[10]   Performance evaluation of an automated image registration algorithm using an integrated kilovoltage imaging and guidance system [J].
Fox, Timothy ;
Huntzinger, Calvin ;
Johnstone, Peter ;
Ogunleye, Tomi ;
Elder, Eric .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2006, 7 (01) :97-104