How does c-view image quality compare with conventional 2D FFDM?

被引:67
作者
Nelson, Jeffrey S. [1 ]
Wells, Jered R. [1 ]
Baker, Jay A. [2 ]
Samei, Ehsan [1 ,2 ,3 ,4 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Clin Imaging Phys Grp, Durham, NC 27705 USA
[2] Duke Univ, Med Ctr, Dept Radiol, Carl E Ravin Adv Imaging Labs, Durham, NC 27705 USA
[3] Duke Univ, Pratt Sch Engn, Dept Biomed Engn, Durham, NC 27705 USA
[4] Duke Univ, Pratt Sch Engn, Dept Elect & Comp Engn, Durham, NC 27705 USA
关键词
digital mammography; digital breast tomosynthesis; synthesized 2D mammogram; C-VIEW; mammography image quality; DIGITAL BREAST TOMOSYNTHESIS; NOISE; MAMMOGRAMS; RESOLUTION; IMPACT;
D O I
10.1118/1.4947293
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The FDA approved the use of digital breast tomosynthesis (DBT) in 2011 as an adjunct to 2D full field digital mammography (FFDM) with the constraint that all DBT acquisitions must be paired with a 2D image to assure adequate interpretative information is provided. Recently manufacturers have developed methods to provide a synthesized 2D image generated from the DBT data with the hope of sparing patients the radiation exposure from the FFDM acquisition. While this much needed alternative effectively reduces the total radiation burden, differences in image quality must also be considered. The goal of this study was to compare the intrinsic image quality of synthesized 2D C-VIEW and 2D FFDM images in terms of resolution, contrast, and noise. Methods: Two phantoms were utilized in this study: the American College of Radiology mammography accreditation phantom (ACR phantom) and a novel 3D printed anthropomorphic breast phantom. Both phantoms were imaged using a Hologic Selenia Dimensions 3D system. Analysis of the ACR phantom includes both visual inspection and objective automated analysis using in-house software. Analysis of the 3D anthropomorphic phantom includes visual assessment of resolution and Fourier analysis of the noise. Results: Using ACR-defined scoring criteria for the ACR phantom, the FFDM images scored statistically higher than C-VIEW according to both the average observer and automated scores. In addition, between 50% and 70% of C-VIEW images failed to meet the nominal minimum ACR accreditation requirements-primarily due to fiber breaks. Software analysis demonstrated that C-VIEW provided enhanced visualization of medium and large microcalcification objects; however, the benefits diminished for smaller high contrast objects and all low contrast objects. Visual analysis of the anthropomorphic phantom showed a measureable loss of resolution in the C-VIEW image (11 lp/mm FFDM, 5 lp/mm C-VIEW) and loss in detection of small microcalcification objects. Spectral analysis of the anthropomorphic phantom showed higher total noise magnitude in the FFDM image compared with C-VIEW Whereas the FFDM image contained approximately white noise texture, the C-VIEW image exhibited marked noise reduction at midfrequency and high frequency with far less noise suppression at low frequencies resulting in a mottled noise appearance. Conclusions: Their analysis demonstrates many instances where the C-VIEW image quality differs from FFDM. Compared to FFDM, C-VIEW offers a better depiction of objects of certain size and contrast, but provides poorer overall resolution and noise properties. Based on these findings, the utilization of C-VIEW images in the clinical setting requires careful consideration, especially if considering the discontinuation of FFDM imaging. Not explicitly explored in this study is how the combination of DBT + C-VIEW performs relative to DBT + FFDM or FFDM alone. (C) 2016 American Association of Physicists in Medicine.
引用
收藏
页码:2538 / 2547
页数:10
相关论文
共 16 条
[1]   Automated analysis of the American College of Radiology mammographic accreditation phantom images [J].
Brooks, KW ;
Trueblood, JH ;
Kearfott, KJ ;
Lawton, DT .
MEDICAL PHYSICS, 1997, 24 (05) :709-723
[2]   Human observer detection experiments with mammograms and power-law noise [J].
Burgess, AE ;
Jacobson, FL ;
Judy, PF .
MEDICAL PHYSICS, 2001, 28 (04) :419-437
[3]   Tomosynthesis imaging: At a translational crossroads [J].
Dobbins, James T., III .
MEDICAL PHYSICS, 2009, 36 (06) :1956-1967
[4]  
Hendrick RE., 1999, MAMMOGRAPHY QUALITY
[5]   Overview of the evidence on digital breast tomosynthesis in breast cancer detection [J].
Houssami, Nehmat ;
Skaane, Per .
BREAST, 2013, 22 (02) :101-108
[6]   Development of realistic physical breast phantoms matched to virtual breast phantoms based on human subject data [J].
Kiarashi, Nooshin ;
Nolte, Adam C. ;
Sturgeon, Gregory M. ;
Segars, William P. ;
Ghate, Sujata V. ;
Nolte, Loren W. ;
Samei, Ehsan ;
Lo, Joseph Y. .
MEDICAL PHYSICS, 2015, 42 (07) :4116-4126
[7]   Breast tomosynthesis: Present considerations and future applications [J].
Park, Jeong Mi ;
Franken, Edmund A. J. R. ;
Garg, Megha ;
Fajardo, Laurie L. ;
Niklason, Loren T. .
RADIOGRAPHICS, 2007, 27 :S231-S240
[8]  
Rose A, 1974, Vision: Human and Electronic
[9]   Does image quality matter? Impact of resolution and noise on mammographic task performance [J].
Saunders, Robert S., Jr. ;
Baker, Jay A. ;
Delong, David M. ;
Johnson, Jeff P. ;
Samei, Ehsan .
MEDICAL PHYSICS, 2007, 34 (10) :3971-3981
[10]   Impact of resolution and noise characteristics of digital radiographic detectors on the detectability of lung nodules [J].
Saunders, RS ;
Samei, E ;
Hoeschen, C .
MEDICAL PHYSICS, 2004, 31 (06) :1603-1613