Tellurium Film Electrodes Deposited on Carbon and Mesoporous Carbon Screen-printed Substrates for Anodic Stripping Voltammetric Determination of Copper

被引:6
作者
Bobrowski, Andrzej [1 ]
Krolicka, Agnieszka [1 ]
Sliwa, Julia [1 ]
Zarebski, Jerzy [1 ]
Economou, Anastasios [2 ]
Kalcher, Kurt [3 ]
机构
[1] AGH Univ Sci & Technol, Fac Mat Sci & Ceram, Dept Bldg Mat Technol, Mickiewicza 30, PL-30059 Krakow, Poland
[2] Univ Athens, Dept Chem, Analyt Chem Lab, Athens 15771, Greece
[3] Karl Franzens Univ Graz, Inst Chem, Analyt Chem, Univ Pl 1, A-8010 Graz, Austria
关键词
Tellurium film electrode; screen-printed carbon electrode; stripping voltammetry; copper; MERCURY-FILM; LEAD FILM; ANTIMONY; PB(II); TARTRATE; SENSORS; CD(II); COBALT; FIELD;
D O I
10.1002/elan.201800165
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The performance of screen-printed electrodes modified insitu with tellurium film for the anodic stripping voltammetric (ASV) determination of Cu(II) is reported. It was found that two types of screen-printed substrates, namely carbon and mesoporous carbon, were optimal for this application. The selected insitu tellurium film modified electrodes were applied for the square wave ASV determination of copper at gL(-1) concentration levels. Well-defined and reproducible Cu oxidation stripping peaks were produced at a potential more negative than the anodic dissolution of tellurium. The highest sensitivity of Cu determination was achieved in 0.05M HCl containing 50gL(-1) Te(IV) after 300s of accumulation at -0.5V. Using the optimized procedure, a linear range from 2 to 35gL(-1) of Cu(II) was obtained with a detection limit of 0.5gL(-1) Cu(II) (S/N=3) for 300s of deposition time. Both sensors, carbon TeF-SPE and mesoporous carbon TeF-SPE, were successfully applied for the quantification of Cu in a certified reference surface water sample.
引用
收藏
页码:2004 / 2010
页数:7
相关论文
共 34 条
[1]  
Ashrafi A. M., 2012, SENSING ELECTROANALY, V7, P189
[2]   New procedures for voltammetric determination of copper (II) using antimony film-coated carbon paste electrodes [J].
Ashrafi, Amir M. ;
Vytras, Karel .
ELECTROCHIMICA ACTA, 2012, 73 :112-117
[3]   Anodic stripping voltammetry at bismuth-coated and uncoated carbon microdisk electrodes: Application to trace metals analysis in food samples [J].
Baldo, MA ;
Daniele, S .
ANALYTICAL LETTERS, 2004, 37 (05) :995-1011
[4]   Antimony Film Electrode Prepared In Situ in Hydrogen Potassium Tartrate in Anodic Stripping Voltammetric Trace Detection of Cd(II), Pb(II), Zn(II), Tl(I), In(III) and Cu(II) [J].
Bobrowski, A. ;
Putek, M. ;
Zarebski, J. .
ELECTROANALYSIS, 2012, 24 (05) :1071-1078
[5]  
Bobrowski A., 2016, INSIGHTS ANAL ELECTR, V1, P1
[6]   Determination of Copper(II) Through Anodic Stripping Voltammetry in Tartrate Buffer Using an Antimony Film Screen-printed Carbon Electrode [J].
Bobrowski, Andrzej ;
Maczuga, Mariola ;
Krolicka, Agnieszka ;
Konstanteli, Evangelia ;
Sakellaropoulou, Chrysavgi ;
Economou, Anastasios .
ANALYTICAL LETTERS, 2017, 50 (18) :2920-2936
[7]   Electrochemical sensing of copper employing tellurium film electrode [J].
Bobrowski, Andrzej ;
Krolicka, Agnieszka ;
Sliwa, Julia ;
Zarebski, Jerzy .
ELECTROCHIMICA ACTA, 2017, 252 :453-460
[8]   Renewable silver amalgam film electrodes in electrochemical stripping analysis-a review [J].
Bobrowski, Andrzej ;
Krolicka, Agnieszka ;
Bobrowski, Rafal .
JOURNAL OF SOLID STATE ELECTROCHEMISTRY, 2016, 20 (12) :3217-3228
[9]   Highly Sensitive and Selective Adsorptive Stripping Voltammetric Method Employing a Lead Film Screen-printed Electrode for Determination of Cobalt as its Nioximate Complex [J].
Bobrowski, Andrzej ;
Krolicka, Agnieszka ;
Maczuga, Mariola ;
Zarebski, Jerzy .
ELECTROANALYSIS, 2016, 28 (02) :343-349
[10]   A novel screen-printed electrode modified with lead film for adsorptive stripping voltammetric determination of cobalt and nickel [J].
Bobrowski, Andrzej ;
Krolicka, Agnieszka ;
Maczuga, Mariola ;
Zarebski, Jerzy .
SENSORS AND ACTUATORS B-CHEMICAL, 2014, 191 :291-297