'Pre-launch' finite element analysis of a short-stem total hip arthroplasty system consisting of two implant types

被引:11
作者
Lerch, Matthias [1 ]
Windhagen, Henning [1 ]
Kurtz, Agnes-Elisabeth [1 ]
Budde, Stefan [1 ]
Behrens, Bernd-Arno [2 ]
Bouguecha, Anas [2 ,3 ]
Almohallami, Amer [2 ,4 ]
机构
[1] Hannover Med Sch, Dept Orthopaed Surg, Anna von Borries Str 1-7, D-30625 Hannover, Germany
[2] Leibniz Univ Hannover, Inst Forming Technol & Machines, Univ 2, D-30823 Hannover, Germany
[3] Natl Sch Engn Gafsa, ENIS, Lab La2MP, Sfax, Tunisia
[4] PROFIL Verbindungstech GmbH & Co KG, Otto Hahn Str 22-24, D-61381 Friedrichsdorf, Germany
关键词
Total hip arthroplasty; Short-stemmed implant; Finite element analysis; Bone remodeling; BONE-MINERAL DENSITY; FEMORAL STEM; FOLLOW-UP; PROSTHESIS; FORCES; SIMULATION; PATTERNS; FEMUR;
D O I
10.1016/j.clinbiomech.2018.11.002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background We applied a previously established and validated numerical model to a novel short-stemmed implant for a 'pre-launch' investigation. Methods: The implant system consists of two different implant geometries for valgus/varus-positioned proximal femurs with differences in volume distribution, head/neck angle, and calcar alignment. The aim of the design was to achieve a better adaption to the anatomic conditions, resulting in a favourable load transfer. The implant type G showed the best fit to our model, but both stem geometries were implanted; the implant type B was used to compute an 'imperfection scenario'. Findings: Apparent bone density decreased by 4.3% in the entire femur with the implant type G, and by 12.3% with the implant type B. Bone mass loss was pronounced in the proximal calcar region. Apparent bone density increased at the lateral cortical ring and in the minor trochanter. The apparent bone density in the imperfection scenario was very similar to that of a straight stem, indicating a distal load transfer. Interpretation: No adverse effects of the A2 short-stemmed implant system on bone remodeling could be detected. The overall bone density reduction was acceptable, and wedge fixation was not observed, indicating that there was no distal load transfer. The simulation of an incongruous implant indicates the sensitivity of our model in response to modifications of implant positioning. Correct implant selection and positioning is crucial when using the A2 system.
引用
收藏
页码:31 / 37
页数:7
相关论文
共 41 条
  • [1] Pattern of periprosthetic bone remodeling around stable uncemented tapered hip stems: A prospective 84-month follow-up study and a median 156-month cross-sectional study with DXA
    Aldinger, PR
    Sabo, D
    Pritsch, M
    Thomsen, M
    Mau, H
    Ewerbeck, V
    Breusch, SJ
    [J]. CALCIFIED TISSUE INTERNATIONAL, 2003, 73 (02) : 115 - 121
  • [2] Fully Porous 3D Printed Titanium Femoral Stem to Reduce Stress-Shielding Following Total Hip Arthroplasty
    Arabnejad, Sajad
    Johnston, Burnett
    Tanzer, Michael
    Pasini, Damiano
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (08) : 1774 - 1783
  • [3] Influence of Hip Prosthesis Size and Its Coating Area on Bone Remodeling
    Behrens, B. -A.
    Bouguecha, A.
    Lerch, M.
    Windhagen, H.
    Almohallami, A.
    [J]. IEEE TRANSACTIONS ON NANOBIOSCIENCE, 2017, 16 (08) : 703 - 707
  • [4] Numerical investigations on the strain-adaptive bone remodelling in the periprosthetic femur: Influence of the boundary conditions
    Behrens, Bernd-Arno
    Nolte, Ingo
    Wefstaedt, Patrick
    Stukenborg-Colsman, Christina
    Bouguecha, Anas
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2009, 8
  • [5] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [6] Frictional and bone ingrowth properties of engineered surface topographies produced by electron beam technology
    Biemond, J. Elizabeth
    Aquarius, Rene
    Verdonschot, Nico
    Buma, Pieter
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2011, 131 (05) : 711 - 718
  • [7] The effect of muscle loading on the simulation of bone remodelling in the proximal femur
    Bitsakos, C
    Kerner, J
    Fisher, I
    Amis, AA
    [J]. JOURNAL OF BIOMECHANICS, 2005, 38 (01) : 133 - 139
  • [8] Continuous bone loss around a tapered uncemented femoral stem -: A long-term evaluation with DEXA
    Boden, Henrik S. G.
    Skoldenberg, Olof G.
    Salemyr, Mats F.
    Lundberg, Hans-Jerker
    Adolphson, Per Y.
    [J]. ACTA ORTHOPAEDICA, 2006, 77 (06) : 877 - 885
  • [9] Changes in bone mineral density in the proximal femur after cementless total hip arthroplasty - A five-year longitudinal study
    Brodner, W
    Bitzan, P
    Lomoschitz, F
    Krepler, P
    Jankovsky, R
    Lehr, S
    Kainberger, F
    Gottsauner-Wolf, F
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (01): : 20 - 26
  • [10] Bone preserving level of osteotomy in short-stem total hip arthroplasty does not influence stress shielding dimensions - a comparing finite elements analysis
    Burchard, Rene
    Braas, Sabrina
    Soost, Christian
    Graw, Jan Adriaan
    Schmitt, Jan
    [J]. BMC MUSCULOSKELETAL DISORDERS, 2017, 18