Corneal Astigmatism Measurements Comparison among Ray-Tracing Aberrometry, Partial Coherence Interferometry, and Scheimpflug Imaging System

被引:0
作者
Zhang, Yaqin [1 ]
Dong, Jing [2 ]
Zhang, Suhua [1 ]
Sun, Bin [1 ]
Wang, Xiaoliang [3 ]
Tang, Maolong [4 ]
Wang, Xiaogang [1 ]
机构
[1] Shanxi Eye Hosp, Taiyuan, Shanxi, Peoples R China
[2] Shanxi Med Univ, Hosp 1, Taiyuan, Shanxi, Peoples R China
[3] Shanghai Jiao Tong Univ, Sch Aeronaut & Astronaut, Shanghai, Peoples R China
[4] Oregon Hlth & Sci Univ, Casey Eye Inst, Portland, OR 97201 USA
基金
中国国家自然科学基金;
关键词
POWER MEASUREMENTS; AUTOMATED KERATOMETRY; AGREEMENT; REPEATABILITY; LOCATION; PENTACAM;
D O I
10.1155/2020/3012748
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To investigate interdevice agreement among corneal topography/ray-tracing aberrometry (iTrace), partial coherence interferometry (IOLMaster), and Scheimpflug imaging (Pentacam) for the measurement of corneal astigmatism. Methods. The analysis included 90 eyes of 90 subjects without ocular disease. The main outcome measures were corneal cylinder power and axis of astigmatism. All corneal astigmatism measurements were converted to two perpendicular components by using vector analysis. Interdevice agreement was assessed using Bland-Altman analysis, paired sample t-test, and one-way analysis of variance. Results. No significant interdevice difference existed in the astigmatism magnitude, cardinal component, and oblique component (all P>0.05). On comparing iTrace wavefront and simulated keratometry (SimK) astigmatism, significant differences were observed in the astigmatism magnitude and oblique component (both P<0.01), but not in the cardinal component (P=0.687). On comparing Pentacam pupil 3 mm and corneal vertex 3 mm axial astigmatism, significant difference was observed in the astigmatism magnitude (P<0.001), but not in the cardinal and oblique components (both P>0.05). Conclusions. The iTrace, IOLMaster, and Pentacam devices could be used interchangeably for corneal astigmatism measurement. However, the measurement difference in iTrace wavefront and SimK astigmatism and Pentacam pupil 3 mm and vertex 3 mm axial astigmatism should be considered in clinic practice.
引用
收藏
页数:7
相关论文
共 23 条
  • [11] Repeatability, reproducibility, and agreement of corneal power measurements obtained with a new corneal topographer
    Mao, Xinjie
    Savini, Giacomo
    Zhuo, Zuopao
    Feng, Yifan
    Zhang, Jian
    Wang, Qinmei
    Chen, Hao
    Huang, Jinhai
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (10) : 1561 - 1569
  • [12] Molebny VV, 2000, J REFRACT SURG, V16, pS572
  • [13] Predicting Postoperative Astigmatism Using Scheimpflug Keratometry (Pentacam) and Automated Keratometry (IOLMaster)
    Park, Choul Yong
    Do, Jae Rock
    Chuck, Roy S.
    [J]. CURRENT EYE RESEARCH, 2012, 37 (12) : 1091 - 1098
  • [14] Comparison of the Astigmatic Power of Toric Intraocular Lenses Using Three Toric Calculators
    Park, Hyun Ju
    Lee, Hun
    Woo, Young Jae
    Kim, Eung Kweon
    Soo, Kyoung Yul
    Kim, Ha Yan
    Kim, Tae-im
    [J]. YONSEI MEDICAL JOURNAL, 2015, 56 (04) : 1097 - 1105
  • [15] Scheimpflug Corneal Power Measurements for Intraocular Lens Power Calculation in Cataract Surgery
    Saad, Elie
    Shammas, Maya C.
    Shammas, H. John
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 156 (03) : 460 - 467
  • [16] Accuracy of Scheimpflug corneal power measurements for intraocular lens power calculation
    Savini, Giacomo
    Barboni, Piero
    Carbonelli, Michele
    Hoffer, Kenneth J.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (07) : 1193 - 1197
  • [17] Comparison of Corneal Powers Obtained from 4 Different Devices
    Shirayama, Mariko
    Wang, Li
    Weikert, Mitchell P.
    Koch, Douglas D.
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 148 (04) : 528 - 535
  • [18] Comparison of corneal astigmatism measurements of 2 optical biometer models for tonic intraocular lens selection
    Srivannaboon, Sabong
    Chirapapaisan, Chareenun
    Chonpimai, Pratuangsri
    Koodkaew, Sunisa
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2015, 41 (02) : 364 - 371
  • [19] Automated keratometry in routine cataract surgery: Comparison of Scheimpflug and conventional values
    Symes, Richard J.
    Ursell, Paul G.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) : 295 - 301
  • [20] Minimum amount of astigmatism that should be corrected
    Villegas, Eloy A.
    Alcon, Encarna
    Artal, Pablo
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2014, 40 (01) : 13 - 19