The biomechanics of plate fixation of periprosthetic femoral fractures near the tip of a total hip implant: cables, screws, or both?

被引:54
作者
Shah, S. [1 ,2 ]
Kim, S. Y. R. [2 ]
Dubov, A. [1 ]
Schemitsch, E. H. [1 ,3 ]
Bougherara, H. [2 ]
Zdero, R. [1 ,2 ]
机构
[1] St Michaels Hosp, Martin Orthopaed Biomech Lab, Toronto, ON M5B 1W8, Canada
[2] Ryerson Univ, Dept Mech & Ind Engn, Toronto, ON, Canada
[3] Univ Toronto, Fac Med, Toronto, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
biomechanics; FE analysis; synthetic femur; hip implant; plate; periprosthetic fracture; FINITE-ELEMENT-ANALYSIS; SYNTHETIC FEMURS; SHAFT FRACTURES; STRAIN DISTRIBUTION; CANCELLOUS BONE; ARTHROPLASTY; REPLACEMENT; COMPRESSION; STRESS; SYSTEM;
D O I
10.1177/0954411911413060
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Femoral shaft fractures after total hip arthroplasty (THA) remain a serious problem, since there is no optimal surgical repair method. Virtually all studies that examined surgical repair methods have done so clinically or experimentally. The present study assessed injury patterns computationally by developing three-dimensional (3D) finite element (FE) models that were validated experimentally. The investigation evaluated three different constructs for the fixation of Vancouver B1 periprosthetic femoral shaft fractures following THA. Experimentally, three bone plate repair methods were applied to a synthetic femur with a 5 mm fracture gap near the tip of a total hip implant. Repair methods were identical distal to the fracture gap, but used cables only (construct A), screws only (construct B), or cables plus screws (construct C) proximal to the fracture gap. Specimens were oriented in 15 degrees adduction to simulate the single-legged stance phase of walking, subjected to 1000 N of axial force, and instrumented with strain gauges. Computationally, a linearly elastic and isotropic 3D FE model was developed to mimic experiments. Results showed excellent agreement between experimental and FE strains, yielding a Pearson linearity coefficient, R(2), of 0.92 and a slope for the line of best data fit of 1.06. FE-computed axial stiffnesses were 768 N/mm (construct A), 1023 N/mm (construct B), and 1102 N/mm (construct C). FE surfaces stress maps for cortical bone showed Von Mises stresses, excluding peaks, of 0-8 MPa (construct A), 0-15 MPa (construct B), and 0-20 MPa (construct C). Cables absorbed the majority of load, followed by the plates and then the screws. Construct A yielded peak stress at one of the empty holes in the plate. Constructs B and C had similar bone stress patterns, and can achieve optimal fixation.
引用
收藏
页码:845 / 856
页数:12
相关论文
共 47 条
[1]  
[Anonymous], ADULT RECONSTRUCTION
[2]  
Beals RK, 1996, CLIN ORTHOP RELAT R, P238
[3]  
BETHEA JS, 1982, CLIN ORTHOP RELAT R, P95
[4]   The biomechanics of a validated finite element model of stress shielding in a novel hybrid total knee replacement [J].
Bougherara, H. ;
Zdero, R. ;
Mahboob, Z. ;
Dubov, A. ;
Shah, S. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2010, 224 (H10) :1209-1219
[5]   The biomechanics of the T2 femoral nailing system: a comparison of synthetic femurs with finite element analysis [J].
Bougherara, H. ;
Zdero, R. ;
Miric, M. ;
Shah, S. ;
Hardisty, M. ;
Zalzal, P. ;
Schemitsch, E. H. .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2009, 223 (H3) :303-314
[6]   A preliminary biomechanical study of a novel carbon-fibre hip implant versus standard metallic hip implants [J].
Bougherara, Habiba ;
Zdero, Rad ;
Dubov, Anton ;
Shah, Suraj ;
Khurshid, Shaheen ;
Schemitsch, Emil H. .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (01) :121-128
[7]   A biomechanical assessment of modular and monoblock revision hip implants using FE analysis and strain gage measurements [J].
Bougherara, Habiba ;
Zdero, Rad ;
Shah, Suraj ;
Miric, Milan ;
Papini, Marcello ;
Zalzal, Paul ;
Schemitsch, Emil H. .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2010, 5
[8]   The reliability and validity of the Vancouver classification of femoral fractures after hip replacement [J].
Brady, OH ;
Garbuz, DS ;
Masri, BA ;
Duncan, CP .
JOURNAL OF ARTHROPLASTY, 2000, 15 (01) :59-62
[9]  
Brady OH, 1999, Tech Orthop, V14, P107
[10]  
Chapman M.W., 2001, CHAPMANS ORTHOPAEDIC, P671