Ultrasonic frequency analysis of antibody-linked hydrogel biosensors for rapid point of care testing

被引:6
作者
Dion, Jonathan R. [1 ]
Burns, David H. [1 ]
机构
[1] McGill Univ, Dept Chem, Montreal, PQ H3A 2K6, Canada
基金
加拿大健康研究院;
关键词
Ultrasound; Hydrogel; Point-of-care; Biosensor; Antibody; Analyte recognition; CONFORMATIONAL-CHANGES; CONTRAST AGENTS; ACETAMINOPHEN; PARACETAMOL; SENSORS; PHASE; PHARMACOKINETICS;
D O I
10.1016/j.talanta.2010.11.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Analyte quantification in highly scattering media such as tissue, blood, and other biological fluids is challenging using conventional spectroscopic methods. Ultrasound easily penetrates these opaque samples, yet currently provides little chemical information. We have developed a general approach for creating hydrogel biosensors based on antibody-linked cellulose polymers. Target recognition induces changes to the sensor stiffness and size, which is accompanied by characteristic changes to a measured ultrasonic frequency profile. Using this technique, nM sensitivity for acetaminophen is demonstrated in a series of biofluids including whole blood, blood plasma, saliva, and urine. Likewise, this methodology is attractive for point of care diagnostics due to the short measurement time, simple methodology which excludes pretreatment of samples, and has minimal chemical or buffer requirements. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:1364 / 1370
页数:7
相关论文
共 35 条
[1]  
[Anonymous], 1966, Applied regression analysis
[2]   Point-of-care i-STAT cardiac troponin I for assessment of patients with symptoms suggestive of acute coronary syndrome [J].
Apple, FS ;
Ler, R ;
Chung, AY ;
Berger, MJ ;
Murakami, MM .
CLINICAL CHEMISTRY, 2006, 52 (02) :322-325
[3]   MULTISPECTRAL ANALYSIS FOR QUANTITATIVE MEASUREMENTS OF MYOGLOBIN OXYGEN FRACTIONAL SATURATION IN THE PRESENCE OF HEMOGLOBIN INTERFERENCE [J].
ARAKAKI, LSL ;
BURNS, DH .
APPLIED SPECTROSCOPY, 1992, 46 (12) :1919-1928
[4]   Oscillating bubble concentration and its size distribution using acoustic emission spectra [J].
Avvaru, Balasubrahmanyam ;
Pandit, Aniruddha B. .
ULTRASONICS SONOCHEMISTRY, 2009, 16 (01) :105-115
[5]   Comparison of urine and serum testing for early detection of acetaminophen ingestion [J].
Bebarta, Vikhyat S. ;
MacDaniel, Joe ;
Schwertner, Harvey A. ;
Martin, James F. .
MILITARY MEDICINE, 2007, 172 (04) :399-401
[6]   A survey of the 2001 to 2005 quartz crystal microbalance biosensor literature: applications of acoustic physics to the analysis of biomolecular interactions [J].
Cooper, Matthew A. ;
Singleton, Victoria T. .
JOURNAL OF MOLECULAR RECOGNITION, 2007, 20 (03) :154-184
[7]   Electronic nose and electronic tongue integration for improved classification of clinical and food samples [J].
Di Natale, C ;
Paolesse, R ;
Macagnano, A ;
Mantini, A ;
D'Amico, A ;
Legin, A ;
Lvova, L ;
Rudnitskaya, A ;
Vlasov, Y .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 64 (1-3) :15-21
[8]   Pharmacokinetics of rectal paracetamol after repeated dosing in children [J].
Hahn, TW ;
Henneberg, SW ;
Holm-Knudsen, RJ ;
Eriksen, K ;
Rasmussen, SN ;
Rasmussen, M .
BRITISH JOURNAL OF ANAESTHESIA, 2000, 85 (04) :512-519
[9]   Oscillations of polymeric microbubbles: Effect of the encapsulating shell [J].
Hoff, L ;
Sontum, PC ;
Hovem, JM .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2000, 107 (04) :2272-2280
[10]  
Hoff L, 1996, ULTRASON, P1441, DOI 10.1109/ULTSYM.1996.584337