Biomechanical Robustness of a Contemporary Cementless Stem to Surgical Variation in Stem Size and Position

被引:15
作者
Al-Dirini, Rami M. A. [1 ]
O'Rourke, Dermot [1 ]
Huff, Daniel [2 ]
Martelli, Saulo [1 ]
Taylor, Mark [1 ]
机构
[1] Flinders Univ S Australia, Coll Sci & Engn, Med Device Res Inst, Adelaide, SA 5043, Australia
[2] DePuy Synthes, Johnson & Johnson, Warsaw, IN 46581 USA
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2018年 / 140卷 / 09期
基金
澳大利亚研究理事会;
关键词
surgical variability; primary stability; implant evaluation; total hip replacement; TOTAL HIP-ARTHROPLASTY; FINITE-ELEMENT-ANALYSIS; FEMORAL COMPONENT ANTEVERSION; PRIMARY STABILITY; BONE-RESORPTION; PATIENT VARIABILITY; REPLACEMENT; ACCURACY; FEMUR; THA;
D O I
10.1115/1.4039824
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Successful designs of total hip replacement (THR) need to be robust to surgical variation in sizing and positioning of the femoral stem. This study presents an automated method for comprehensive evaluation of the potential impact of surgical variability in sizing and positioning on the primary stability of a contemporary cementless femoral stein (Corail (R) DePuy Synthes). A patient-specific finite element (FE) model of a femur was generated from computed tomography (CT) images from a female donor. An automated algorithm was developed to span the plausible surgical envelope of implant positions constrained by the inner cortical boundary. The analysis was performed on four stem sizes: oversized, ideal (nominal) sized, and undersized by up to two stem sizes. For each size, Latin hypercube sampling was used to generate models for 100 unique alignment scenarios. For each scenario, peak hip contact and muscle forces published for stair climbing were scaled to the donor's body weight and applied to the model. The risk of implant loosening was assessed by comparing the bone-implant micromotion/strains to thresholds (150 mu m and 7000 mu epsilon) above which fibrous tissue is expected to prevail and the periprosthetic bone to yield, respectively. The risk of long-term loosening due to adverse bone resorption was assessed using bone adaptation theory. The range of implant positions generated effectively spanned the available intracortical space. The Corail stein was found stable and robust to changes in size and position, with the majority of the bone-implant interface undergoing micromotion and interfacial strains that are well below 150 pm and 7000 pa, respectively. Nevertheless, the range of implant positions generated caused an increase of up to 50% in peak micromotion and up to 25% in interfacial strains, particularly for retroverted stems placed in a medial position.
引用
收藏
页数:12
相关论文
共 57 条
[1]   Finite element modelling of primary hip stem stability: The effect of interference fit [J].
Abdul-Kadir, Mohammed Rafiq ;
Hansen, Ulrich ;
Klabunde, Ralf ;
Lucas, Duncan ;
Amis, Andrew .
JOURNAL OF BIOMECHANICS, 2008, 41 (03) :587-594
[2]  
Al-Dirini R. M. A., 2016, ORTH RES SOC ORS ANN, P1185
[3]  
AOANJRR, 2015, HIP KNEE ARTHR ANN R
[4]  
Bader RJ., 2001, Hip Int, V11, P80, DOI [10.1177/112070000101100203, DOI 10.1177/112070000101100203]
[5]   Efficient computational method for assessing the effects of implant positioning in cementless total hip replacements [J].
Bah, Mamadou T. ;
Nair, Prasanth B. ;
Taylor, Mark ;
Browne, Martin .
JOURNAL OF BIOMECHANICS, 2011, 44 (07) :1417-1422
[6]  
Bah MT, 2010, 17 C EUR SOC BIOM ED
[7]   The pathology of total joint arthroplasty - II. Mechanisms of implant failure [J].
Bauer, TW ;
Schils, J .
SKELETAL RADIOLOGY, 1999, 28 (09) :483-497
[8]   DETERMINATION OF ACCURACY OF PREOPERATIVE TEMPLATING OF NONCEMENTED FEMORAL PROSTHESES [J].
CARTER, LW ;
STOVALL, DO ;
YOUNG, TR .
JOURNAL OF ARTHROPLASTY, 1995, 10 (04) :507-513
[9]   Preoperative planning for primary total hip arthroplasty [J].
Della Valle, AG ;
Padgett, DE ;
Salvati, EA .
JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2005, 13 (07) :455-462
[10]  
DePuy Synthes, 2016, CORAIL HIP SYST SURG