The Movesense Medical Sensor Chest Belt Device as Single Channel ECG for RR Interval Detection and HRV Analysis during Resting State and Incremental Exercise: A Cross-Sectional Validation Study

被引:27
|
作者
Rogers, Bruce [1 ]
Schaffarczyk, Marcelle [2 ]
Clauss, Martina [3 ]
Mourot, Laurent [4 ,5 ]
Gronwald, Thomas [6 ]
机构
[1] Univ Cent Florida, Coll Med, Dept Internal Med, Orlando, FL 32827 USA
[2] Univ Hamburg, Inst Human Movement Sci, Dept Sports & Exercise Med, D-20148 Hamburg, Germany
[3] Univ Leipzig, Fac Sport Sci, Inst Movement & Trainings Sci Sport, D-04109 Leipzig, Germany
[4] Univ Bourgogne Franche Comte, Exercise Performance Hlth Innovat EPHI Plaptform, EA3920 Pronost Factors & Regulatory Factors Cardi, F-25000 Besancon, France
[5] Natl Res Tomsk Polytech Univ, Div Phys Educ, Tomsk 634040, Russia
[6] MSH Med Sch Hamburg, Inst Interdisciplinary Exercise Sci & Sports Med, D-20457 Hamburg, Germany
关键词
heart rate variability; wearable; medical device; health care; exercise; HEART-RATE-VARIABILITY; NONLINEAR DYNAMICS; ASSOCIATION; AGE; MORTALITY; ADULTS; RISK;
D O I
10.3390/s22052032
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The value of heart rate variability (HRV) in the fields of health, disease, and exercise science has been established through numerous investigations. The typical mobile-based HRV device simply records interbeat intervals, without differentiation between noise or arrythmia as can be done with an electrocardiogram (ECG). The intent of this report is to validate a new single channel ECG device, the Movesense Medical sensor, against a conventional 12 channel ECG. A heterogeneous group of 21 participants performed an incremental cycling ramp to failure with measurements of HRV, before (PRE), during (EX), and after (POST). Results showed excellent correlations between devices for linear indexes with Pearson's r between 0.98 to 1.0 for meanRR, SDNN, RMSSD, and 0.95 to 0.97 for the non-linear index DFA a1 during PRE, EX, and POST. There was no significant difference in device specific meanRR during PRE and POST. Bland-Altman analysis showed high agreement between devices (PRE and POST: meanRR bias of 0.0 and 0.4 ms, LOA of 1.9 to -1.8 ms and 2.3 to -1.5; EX: meanRR bias of 11.2 to 6.0 ms; LOA of 29.8 to -7.4 ms during low intensity exercise and 8.5 to 3.5 ms during high intensity exercise). The Movesense Medical device can be used in lieu of a reference ECG for the calculation of HRV with the potential to differentiate noise from atrial fibrillation and represents a significant advance in both a HR and HRV recording device in a chest belt form factor for lab-based or remote field-application.