Analysis of image quality and dose calculation accuracy in cone beam CT acquisitions with limited projection data (half scan, half fan) with regard to usability for adaptive radiation therapy treatment planning

被引:7
作者
Kowatsch, Matthias [1 ,2 ]
Winkler, Peter [1 ]
Zurl, Brigitte [1 ]
Konrad, Thomas [1 ]
Schroettner, Joerg [2 ]
机构
[1] Med Univ Graz, Univ Klin Strahlentherapie Radioonkol, A-8036 Graz, Austria
[2] Graz Univ Technol, Inst Hlth Care Engn, Graz, Austria
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2011年 / 21卷 / 01期
关键词
cone beam CT; image quality; treatment planning; dose calculation; COMPUTED-TOMOGRAPHY; GUIDED RADIOTHERAPY; PRACTICAL METHOD; SYSTEM;
D O I
10.1016/j.zemedi.2010.08.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Subject of this study is the question of whether cone beam CT (CBCT) images with reduced projection data are suitable for use in adaptive radiation therapy (ART) treatment planning. For this purpose image quality and dose calculation accuracy depending on imaging modality were analysed. In this context, two CBCT-methods will be indicated having reduced projection data sets: Scans acquired with 200 degrees rotation angle in order to accelerate the CBCT process (half scan), or scans with an asymmetric cone beam and detector offset, used to enlarge the field-of-view (half fan). Methods: For three different CBCT-modes (On-Board-Imaging, Varian Medical Systems), two of them based on reduced projection data, and a conventional multidetector CT system, the main image quality parameters were studied. Treatment plans for two phantoms were transferred to all datasets and re-computed to analyse dose calculation accuracy. Furthermore imaging dose was measured for all modalities. Results: All three CBCT-modes showed similar results with regard to image quality. It was found, that a reduction in projection data does not necessarily involve deterioration in image quality parameters. For dose calculation based on CBCT images, a good agreement with the reference plan was found, with a maximum deviation for the mean dose in regions of interest of 1.1%. Imaging dose was found to be 2.5 cGy and 2.9 cGy for the large-FOV mode and the partial rotation mode, respectively, and 5.4 cGy for the 360 degrees-full fan mode.
引用
收藏
页码:11 / 18
页数:8
相关论文
共 19 条
[1]   A quality assurance program for image quality of cone-beam CT guidance in radiation therapy [J].
Bissonnette, Jean-Pierre ;
Moseley, Douglas J. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2008, 35 (05) :1807-1815
[2]   Advances in image-guided radiation therapy [J].
Dawson, Laura A. ;
Jaffray, David A. .
JOURNAL OF CLINICAL ONCOLOGY, 2007, 25 (08) :938-946
[3]   Image-guided radiotherapy: rationale, benefits, and limitations [J].
Dawson, Laura A. ;
Sharpe, Michael B. .
LANCET ONCOLOGY, 2006, 7 (10) :848-858
[4]   A study on adaptive IMRT treatment planning using kV cone-beam CT [J].
Ding, George X. ;
Duggan, Dennis M. ;
Coffey, Charles W. ;
Deeley, Matthew ;
Hallahan, Dennis E. ;
Cmelak, Anthony ;
Malcolm, Arnold .
RADIOTHERAPY AND ONCOLOGY, 2007, 85 (01) :116-125
[5]   A PRACTICAL METHOD TO MEASURE THE MTF OF CT SCANNERS [J].
DROEGE, RT ;
MORIN, RL .
MEDICAL PHYSICS, 1982, 9 (05) :758-759
[6]   A PRACTICAL METHOD TO ROUTINELY MONITOR RESOLUTION IN DIGITAL IMAGES [J].
DROEGE, RT .
MEDICAL PHYSICS, 1983, 10 (03) :337-343
[7]   Dose and image quality for a cone-beam C-arm CT system [J].
Fahrig, Rebecca ;
Dixon, Robert ;
Payne, Thomas ;
Morin, Richard L. ;
Ganguly, Arundhuti ;
Strobel, Norbert .
MEDICAL PHYSICS, 2006, 33 (12) :4541-4550
[8]   Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapy [J].
Hatton, Joan ;
McCurdy, Boyd ;
Greer, Peter B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (15) :N329-N346
[9]   Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy [J].
Islam, Mohammad K. ;
Purdie, Thomas G. ;
Norrlinger, Bernhard D. ;
Alasti, Hamideh ;
Moseley, Douglas J. ;
Sharpe, Michael B. ;
Siewerdsen, Jeffrey H. ;
Jaffray, David A. .
MEDICAL PHYSICS, 2006, 33 (06) :1573-1582
[10]   A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets [J].
Jaffray, DA ;
Drake, DG ;
Moreau, M ;
Martinez, AA ;
Wong, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03) :773-789