X-ray vision: the accuracy and repeatability of a technology that allows clinicians to see spinal X-rays superimposed on a person's back

被引:5
作者
Aaskov, Jacob [1 ]
Kawchuk, Gregory N. [1 ,2 ,3 ]
Hamaluik, Kenton D. [2 ]
Boulanger, Pierre [4 ]
Hartvigsen, Jan [1 ,3 ]
机构
[1] Univ Southern Denmark, Sports Sci & Clin Biomech, Odense, Denmark
[2] Univ Alberta, Phys Therapy, Edmonton, AB, Canada
[3] Univ Southern Denmark, Nord Inst Chiropract & Clin Biomech, Odense, Denmark
[4] Univ Alberta, Comp Sci, Edmonton, AB, Canada
关键词
X-ray; Heads up display; Mixed reality; Spine; AUGMENTED REALITY;
D O I
10.7717/peerj.6333
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objective: Since the discovery of ionizing radiation, clinicians have evaluated X-ray images separately from the patient. The objective of this study was to investigate the accuracy and repeatability of a new technology which seeks to resolve this historic limitation by projecting anatomically correct X-ray images on to a person's skin. Methods: A total of 13 participants enrolled in the study, each having a pre-existin g anteroposterior lumbar X-ray. Each participant's image was uploaded into the Hololens Mixed reality system which when worn, allowed a single examiner to view a participant's own X-ray superimposed on the participant's back. The projected image was topographically corrected using depth information obtained by the Hololens system then aligned via existing anatomic landmarks. Using this superimposed image, vertebral levels were identified and validated against spinous process locations obtained by ultrasound. This process was repeated 1-5 days later. The projection of each vertebra was deemed to be "on-target" if it fell within the known morphological dimensions of the spinous process for that specific vertebral level. Results: The projection system created on-target projections with respect to individual vertebral levels 73% of the time with no significant difference seen between testing sessions. The average repeatability for all vertebral levels between testing sessions was 77%. Conclusion: These accuracy and repeatability data suggest that the accuracy and repeatability of projecting X-rays directly on to the skin is feasible for identifying underlying anatomy and as such, has potential to place radiological evaluation within the patient context. Future opportunities to improve this procedure will focus on mitigating potential sources of error.
引用
收藏
页数:11
相关论文
共 17 条
[1]   Inaccuracies in the Use of Magnification Markers in Digital Hip Radiographs [J].
Archibeck, Michael J. ;
Cummins, Tamara ;
Tripuraneni, Krishna R. ;
Carothers, Joshua T. ;
Murray-Krezan, Cristina ;
Hattab, Mohammad ;
White, Richard E., Jr. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (08) :1812-1817
[2]  
Bourdel N, 2017, FERTIL STERIL, V107, P737, DOI [10.1016/j.fertnstert.2016.12.016, 10.1016/.fertnstert.2016.12.016]
[3]   Ultrasonography of the Adult Thoracic and Lumbar Spine for Central Neuraxial Blockade [J].
Chin, Ki Jinn ;
Karmakar, Manoj Kumar ;
Peng, Philip .
ANESTHESIOLOGY, 2011, 114 (06) :1459-1485
[4]   The use of pMRI to validate the identification of palpated bony landmarks [J].
Cooper, Kay ;
Alexander, Lyndsay ;
Hancock, Elizabeth ;
Smith, Francis W. .
MANUAL THERAPY, 2013, 18 (04) :289-293
[5]   Head-mounted display augmented reality to guide pedicle screw placement utilizing computed tomography [J].
Gibby, Jacob T. ;
Swenson, Samuel A. ;
Cvetko, Steve ;
Rao, Raj ;
Javan, Ramin .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2019, 14 (03) :525-535
[6]   A wearable headset for monitoring electromyography responses within spinal surgery [J].
Golab, M. R. ;
Breedon, P. J. ;
Vloeberghs, M. .
EUROPEAN SPINE JOURNAL, 2016, 25 (10) :3214-3219
[7]   A simple and convenient method guide to determine the magnification of digital Xrays for preoperative planning in total hip arthroplasty [J].
Heep, Hansjoerg ;
Xu, Jie ;
Loechteken, Christian ;
Wedemeyer, Christian .
ORTHOPEDIC REVIEWS, 2012, 4 (01) :55-58
[8]   Clinical Feasibility of a Wearable Mixed-Reality Device in Neurosurgery [J].
Incekara, Fatih ;
Smits, Marion ;
Dirven, Clemens ;
Vincent, Arnaud .
WORLD NEUROSURGERY, 2018, 118 :E422-E427
[9]   Ultrasound-Guided Fine-Needle Aspiration in the Neck Region Using an Optical See-Through Head-Mounted Display: A Randomized Controlled Trial [J].
Kaneko, Naoki ;
Tsunoda, Mayumi ;
Mitsuhashi, Masatsugu ;
Okubo, Keisuke ;
Takeshima, Taro ;
Sehara, Yoshihide ;
Nagai, Mutsumi ;
Kawai, Kensuke .
JOURNAL OF ULTRASOUND IN MEDICINE, 2017, 36 (10) :2071-2077
[10]   MEASUREMENT OF OBSERVER AGREEMENT FOR CATEGORICAL DATA [J].
LANDIS, JR ;
KOCH, GG .
BIOMETRICS, 1977, 33 (01) :159-174