Preparation and characterization of monolithic porous capillary columns loaded with chromatographic particles

被引:168
作者
Dulay, MT [1 ]
Kulkarni, RP [1 ]
Zare, RN [1 ]
机构
[1] Stanford Univ, Dept Chem, Stanford, CA 94305 USA
关键词
D O I
10.1021/ac9806456
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Using sol-gel technology, a porous glass matrix (xerogel) is formed in a capillary column and acts as a support for a stationary phase of chromatographic particles used in capillary electrochromatography. Preparation of the sol-gel matrix and immobilization of the octadecylsilica (ODS) stationary phase occur in a single step. The presence of the particles in the column greatly reduces matrix cracking caused by internal pressure differentials within the pores of the sol-gel matrix. Good electroosmotic now is achieved in part because of the inherent negative charge of both the particles and the sol-gel matrix. The performance of these sol-gel/ODS capillary columns was evaluated with a mixture of aromatic and nonaromatic organic compounds. Efficiencies of up to 80 000 plates/m were observed in columns with immobilized 3-mu m ODS particles. The efficiency and resolution are enhanced when 3-mu m ODS particles are used in place of the 5-mu m particles.
引用
收藏
页码:5103 / 5107
页数:5
相关论文
共 44 条
[1]   Sintered octadecylsilica as monolithic column packing in capillary electrochromatography and micro high-performance liquid chromatography [J].
Asiaie, R ;
Huang, X ;
Farnan, D ;
Horváth, C .
JOURNAL OF CHROMATOGRAPHY A, 1998, 806 (02) :251-263
[2]  
BANDINI GE, 1995, ANALYST, V120, P1025
[3]  
BOHLIN K, 1993, J CHROMATOGR, V645, P41
[4]  
BRINKER CJ, 1990, SOL GEL SCI PHYS CHE, pCH1
[5]   ELECTRICALLY DRIVEN OPEN-TUBULAR LIQUID-CHROMATOGRAPHY [J].
BRUIN, GJM ;
TOCK, PPH ;
KRAAK, JC ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1990, 517 :557-572
[6]   Sol-gel coating technology for the preparation of solid-phase microextraction fibers of enhanced thermal stability [J].
Chong, SL ;
Wang, DX ;
Hayes, JD ;
Wilhite, BW ;
Malik, A .
ANALYTICAL CHEMISTRY, 1997, 69 (19) :3889-3898
[7]  
COLON LA, 1997, ANAL CHEM, V15, pA4461
[8]   PREPARATION OF OPEN-TUBULAR COLUMNS FOR REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CREGO, AL ;
DIEZMASA, JC ;
DABRIO, MV .
ANALYTICAL CHEMISTRY, 1993, 65 (11) :1615-1621
[9]  
DADOO R, UNPUB ANAL CHEM
[10]   Automated capillary electrochromatography: Reliability and reproducibility studies [J].
Dulay, MT ;
Yan, C ;
Rakestraw, DJ ;
Zare, RN .
JOURNAL OF CHROMATOGRAPHY A, 1996, 725 (02) :361-366