FMEA of MR-Only Treatment Planning in the Pelvis

被引:28
作者
Kim, Joshua [1 ]
Miller, Brett [1 ]
Siddiqui, M. Salim [1 ]
Movsas, Benjamin [1 ]
Glide-Hurst, Carri [1 ]
机构
[1] Henry Ford Hlth Syst, Dept Radiat Oncol, Detroit, MI 48202 USA
基金
美国国家卫生研究院;
关键词
FAILURE MODE; CT; DISTORTION; SEQUENCES; PHANTOM; IMAGES;
D O I
10.1016/j.adro.2018.08.024
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the implementation of a magnetic resonance (MR)-only workflow (ie, implementing MR simulation as the primary planning modality) using failure mode and effects analysis (FMEA) in comparison with a conventional multimodality (MR simulation in conjunction with computed tomography simulation) workflow for pelvis external beam planning. Methods and Materials: To perform the FMEA, a multidisciplinary 9-member team was assembled and developed process maps, identified potential failure modes (FMs), and assigned numerical values to the severity (S), frequency of occurrence (O), and detectability (D) of those FMs. Risk priority numbers (RPNs) were calculated via the product of S, O, and D as a metric for evaluating relative patient risk. An alternative 3-digit composite number (SOD) was computed to emphasize high-severity FMs. Fault tree analysis identified the causality chain leading to the highest-severity FM. Results: Seven processes were identified, 3 of which were shared between workflows. Image fusion and target delineation subprocesses using the conventional workflow added 9 and 10 FMs, respectively, with 6 RPNs >100. By contrast, synthetic computed tomography generation introduced 3 major subprocesses and propagated 46 unique FMs, 15 with RPNs >100. For the conventional workflow, the largest RPN scores were introduced by image fusion (RPN range, 120-192). For the MR-only workflow, the highest RPN scores were from inaccuracies in target delineation resulting from misinterpretation of MR images (RPN = 240) and insufficient management of patient- and system-level distortions (RPN = 210 and 168, respectively). Underestimation (RPN = 140) or overestimation (RPN = 192) of bone volume produced higher RPN scores. The highest SODs for both workflows were related to changes in target location because of internal anatomy changes (conventional = 961, MR-only = 822). Conclusions: FMEA identified areas for mitigating risk in MR-only pelvis RTP, and SODs identified high-severity process modes. Efforts to develop a quality management program to mitigate high FMs are underway. (C) 2018 The Author(s). Published by Elsevier Inc. on behalf of American Society for Radiation Oncology.
引用
收藏
页码:168 / 176
页数:9
相关论文
共 30 条
[1]  
[Anonymous], 2008, POTENTIAL FAILURE MO, V4th
[2]  
Bajaj A, 2016, ONCOL TIMES, V38, P14
[3]  
Barentsz J, 2016, RAD SOC N AM 2016 SC
[4]  
Carbone TA, 2008, ENG MANAGE J, V16, P28
[5]  
Chen S, 2016, J APPL CLIN MED PHYS, V17, P6065
[6]   The contribution of magnetic resonance imaging to the three-dimensional treatment planning of localized prostate cancer [J].
Debois, M ;
Oyen, R ;
Maes, F ;
Verswijvel, G ;
Gatti, G ;
Bosmans, H ;
Feron, M ;
Bellon, E ;
Kutcher, G ;
Van Poppel, H ;
Vanuytsel, L .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (04) :857-865
[7]   An Atlas-Based Electron Density Mapping Method for Magnetic Resonance Imaging (MRI)-Alone Treatment Planning and Adaptive MRI-Based Prostate Radiation Therapy [J].
Dowling, Jason A. ;
Lambert, Jonathan ;
Parker, Joel ;
Salvado, Olivier ;
Fripp, Jurgen ;
Capp, Anne ;
Wratten, Chris ;
Denham, James W. ;
Greer, Peter B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (01) :E5-E11
[8]   CT and MRI derived source localization error in a custom prostate phantom using automated image coregistration [J].
Dubois, DF ;
Bice, WS ;
Prestige, BR .
MEDICAL PHYSICS, 2001, 28 (11) :2280-2284
[9]   EVALUATION OF SAFETY IN A RADIATION ONCOLOGY SETTING USING FAILURE MODE AND EFFECTS ANALYSIS [J].
Ford, Eric C. ;
Gaudette, Ray ;
Myers, Lee ;
Vanderver, Bruce ;
Engineer, Lilly ;
Zellars, Richard ;
Song, Danny Y. ;
Wong, John ;
DeWeese, Theodore L. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03) :852-858
[10]   Investigation of a method for generating synthetic CT models from MRI scans of the head and neck for radiation therapy [J].
Hsu, Shu-Hui ;
Cao, Yue ;
Huang, Ke ;
Feng, Mary ;
Balter, James M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (23) :8419-8435