Antibody fragment recognition layers for surface plasmon resonance biosensing: a parametric study

被引:0
|
作者
Magalhaes, Andre [1 ,2 ]
Bordeira, Sandro [1 ]
Almeida, Ana Cristina [1 ]
Fontes, Vanessa [1 ]
Leite Costa, Maria Joao [1 ]
Fonseca, Luis P. [2 ]
da Fonseca, Joao Garcia [1 ]
机构
[1] Biosurfit SA, Ed ICAT, Campus Fac Ciencias, P-1749016 Lisbon, Portugal
[2] Inst Super Tecn, P-1049001 Lisbon, Portugal
来源
ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS VII | 2009年 / 7169卷
关键词
Immunosensor; Surface plasmon resonance; Antibody fragments; Non-specific binding; C-reactive protein;
D O I
10.1117/12.808820
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A comparative study is reported regarding the use of two different surface plasmon resonance (SPR) biosensors, a homemade SPR grating biosensor and a reference prism coupled biosensor, to perform quantification of C-reactive protein (CRP) in human blood serum. Surface functionalization was conducted using anti-CRP fragments immobilized directly on gold. Adsorption time optimization for the antibody fragments monolayer, non-specific binding (NSB) resistance evaluation and CRP detection were conducted, with better results achieved by the grating biosensor on all topics, namely less functionalization time, higher resistance to NSB and wider CRP dynamic concentration range. A study regarding comparison between continuous flow and surface coating immobilization is also reported in this work. We have shown that surface coating immobilization achieves similar NSB resistance and CRP detection results, allowing a 75% assay cost reduction by lower solution volume requirement. Results suggest that the coating immobilization technique is the best suited to be used in further studies in order to obtain a viable immunosensor for CRP and other biomarkers detection in complex biological fluids.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Estrogen conjugation and antibody binding interactions in surface plasmon resonance biosensing
    Mitchell, John S.
    Wu, Yinqiu
    Cook, Christian J.
    Main, Lyndsay
    STEROIDS, 2006, 71 (07) : 618 - 631
  • [2] Comparative Assessment of Oriented Antibody Immobilization on Surface Plasmon Resonance Biosensing
    Chang, Chia-Chen
    Chuang, Tsung-Liang
    Wang, Da-Shin
    Wang, Ching-Ho
    Lin, Chii-Wann
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 2013, 60 (12) : 1449 - 1456
  • [3] Airborne surface plasmon resonance biosensing
    Chinowsky, T
    Naimushin, A
    Soelberg, S
    Spinelli, C
    Kauffman, P
    Yee, S
    Furlong, C
    ENVIRONMENTAL MONITORING AND REMEDIATION III, 2004, 5270 : 182 - 188
  • [4] Immobilization of antibody fragment for immunosensor application based on surface plasmon resonance
    Lee, W
    Oh, BK
    Lee, WH
    Choi, JW
    COLLOIDS AND SURFACES B-BIOINTERFACES, 2005, 40 (3-4) : 143 - 148
  • [5] Surface plasmon resonance polarizator for biosensing and imaging
    Patskovsky, S.
    Meunier, M.
    Kabashin, A. V.
    OPTICS COMMUNICATIONS, 2008, 281 (21) : 5492 - 5496
  • [6] REVIEW: ADVANCES AND APPLICATIONS OF SURFACE PLASMON RESONANCE BIOSENSING INSTRUMENTATION
    Wang, Xiaoping
    Zhan, Shuyue
    Huang, Zihao
    Hong, Xiayun
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2013, 41 (06) : 574 - 607
  • [7] Fragment Screening by Surface Plasmon Resonance
    Navratilova, Iva
    Hopkins, Andrew L.
    ACS MEDICINAL CHEMISTRY LETTERS, 2010, 1 (01): : 44 - 48
  • [8] Antibody networks for surface plasmon resonance immunosensors
    Brynda, E
    Homola, J
    Houska, M
    Pfeifer, P
    Skvor, J
    SENSORS AND ACTUATORS B-CHEMICAL, 1999, 54 (1-2) : 132 - 136
  • [9] Optical biosensing using surface plasmon resonance spectroscopy
    Homola, J
    Pfeifer, P
    Brynda, E
    Skvor, J
    Houska, M
    Schwotzer, G
    Latka, I
    Willsch, R
    CHEMICAL, BIOCHEMICAL, AND ENVIRONMENTAL FIBER SENSORS IX, 1997, 3105 : 318 - 324
  • [10] Multi-analyte surface plasmon resonance biosensing
    Homola, J
    Vaisocherová, H
    Dostálek, J
    Piliarik, M
    METHODS, 2005, 37 (01) : 26 - 36