Mechanical behavior of in vivo degraded second generation resorbable magnesium scaffolds (RMS)

被引:13
作者
Brandt-Wunderlich, Christoph [1 ]
Ruppelt, Philipp [2 ]
Zumstein, Philine [2 ]
Schmidt, Wolfram [3 ]
Arbeiter, Daniela [3 ]
Schmitz, Klaus-Peter [1 ,3 ]
Grabow, Niels [3 ]
机构
[1] Inst ImplantTechnol & Biomat eV, Friedrich Barnewitz Str 4, D-18119 Rostock, Germany
[2] BIOTRONIK AG, Bulach, Switzerland
[3] Rostock Univ, Med Ctr, Inst Biomed Engn, Rostock, Germany
关键词
Resorbable scaffold; Magnesium; in vivo degradation; Mechanical properties; Vascular restoration; Magmaris; ABSORBABLE METAL SCAFFOLD; BIORESORBABLE SCAFFOLDS; BIOSOLVE-II; SUSTAINED SAFETY; PERFORMANCE; VITRO; CORROSION; MULTICENTER; PARADIGM; ALLOYS;
D O I
10.1016/j.jmbbm.2018.12.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Resorbable magnesium scaffolds are used for the treatment of atherosclerotic coronary vascular disease and furthermore, for vascular restoration therapy. Recently, the first-in-man clinical studies with Magmaris showed promising results regarding the target lesion failure as well as vasomotion properties after 12 and 24 month. The consistency of in vivo degraded magnesium alloys in a cardiovascular environment is qualitatively described in literature, but only little has been disclosed about the actual change in mechanical properties and the behavior of the magnesium alloy degradation products. In the present study, uncoated magnesium scaffolds 3.0 x 20 mm were implanted in coronary arteries of two healthy Goetinriger mini-swine. The scaffolds were explanted to evaluate the mechanical properties of the degraded magnesium scaffolds after 180 days in vivo. Ex vivo sample preparation and test conditions were adapted to a customized compression test setup which was developed to investigate the micro-scale scaffold fragments (width 225 +/- 75 mu m, thickness 150 mu m). As reference bare undegraded magnesium scaffold fragments were tested. Mechanical parameters relating to force as a function of displacement were determined for both sample groups. The undegraded samples showed no fracturing at the maximum applied force of 8 N, whereas the in vivo degraded test samples showed forces of 0.411 +/- 0.197 N at the first fracturing and a maximum force of 0.956 +/- 0.525 N. The deformation work, calculated as area beneath the force-displacement curve, of the in vivo degraded test samples was reduced by approximately 87-88% compared to the undegraded samples (5.20 mN mm and 40.79 mN mm, both at 7.5% deformation). The indication for a complete loss of structural integrity through a reduction of mechanical properties after a certain degradation time increases the chance to restore vascular function and physiological vasomotion in the stented vessel compartment.
引用
收藏
页码:174 / 181
页数:8
相关论文
共 25 条
[21]   Bioresorbable scaffolds: a new paradigm in percutaneous coronary intervention [J].
Tenekecioglu, Erhan ;
Farooq, Vasim ;
Bourantas, Christos V. ;
Silva, Rafael Cavalcante ;
Onuma, Yoshinobu ;
Yilmaz, Mustafa ;
Serruys, Patrick W. .
BMC CARDIOVASCULAR DISORDERS, 2016, 16
[22]   Serial Observation of Drug-Eluting Absorbable Metal Scaffold Multi-Imaging Modality Assessment [J].
Waksman, Ron ;
Prati, Francesco ;
Bruining, Nico ;
Haude, Michael ;
Boese, Dirk ;
Kitabata, Hironori ;
Erne, Paul ;
Verheye, Stefan ;
Degen, Hubertus ;
Vermeersch, Paul ;
Di Vito, Luca ;
Koolen, Jacques ;
Erbel, Raimund .
CIRCULATION-CARDIOVASCULAR INTERVENTIONS, 2013, 6 (06) :644-653
[23]   Bioresorbable drug-eluting magnesium-alloy scaffold: design and feasibility in a porcine coronary model [J].
Wittchow, Eric ;
Adden, Nina ;
Riedmueller, Johannes ;
Savard, Claudine ;
Waksman, Ron ;
Braune, Marlen .
EUROINTERVENTION, 2013, 8 (12) :1441-1450
[24]   In vitro degradation of biodegradable polymer-coated magnesium under cell culture condition [J].
Xu, Liping ;
Yamamoto, Akiko .
APPLIED SURFACE SCIENCE, 2012, 258 (17) :6353-6358
[25]   First biodegradable metal stent in a child with congenital heart disease: Evaluation of macro and histopathology [J].
Zartner, Peter ;
Buettner, Maike ;
Singer, Helmut ;
Sigler, Matthias .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (03) :443-446